Vital Signs™ B1714XXXRecalls & reported problems
Airlife · Vital Signs™ Pediatric Anethesia Breathing Circuit · Class 1
All models of Vital Signs (48)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10190752124004), received between Jan 24, 2023 and Jul 6, 2023 — classified as 4 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Leak/Splash6
- Material Puncture/Hole3
- Crack1
- Defective Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryJul 6, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED TO VYAIRE MEDICAL THAT WHEN THE PATIENT WAS SEDATED FOR INTUBATION THE ENDOTRACHEAL (ET) TUBE WAS PLACED BY ANESTHESIA AND HOOKED UP TO THE CIRCUIT.WHEN THE PATIENT WAS MANUALLY VENTILATED, IT SOUNDED LIKE THERE WAS A LEAK IN THE B1714XXX ANES CIRCUIT, PED, 108 IN EXP, 1L BAG. THE ANESTHESIOLOGIST NOTED THERE WAS NOT GOOD VENTILATION. THE PATIENT WAS EXTUBATED AND VENTILATED USING A BAG VALVE MASK WHILE THE CIRCUIT TUBING AND BAG WERE REPLACED. AFTERWARDS, THE SOUND OF THE LEAK WAS GONE AND BETTER VENTILATION WAS ACHIEVED. MORE MEDICATION WAS PROVIDED TO SEDATE FOR INTUBATION WAS DONE AND THE ET WAS REPLACED WITH GOOD SEAL IN THE TUBE AND THE CIRCUIT WAS VENTILATING AS IT SHOULD BE. AT THIS TIME, THE CUSTOMER HAS NOT PROVIDED ANY INFORMATION REGARDING PATIENT HARM OR INJURY ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17263966 ↗InjuryMay 24, 2023Leak/Splash; Material Puncture/Hole; Leak/Splash; Material Puncture/Hole
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE B1714XXX -ANES CIRCUIT, PED, 108 IN EXP, 1L BAG HAS HOLES IN THE CIRCUIT CAUSING AN AIR LEAK. THE PRODUCT WAS IN THE PROCESS OF BEING ATTACHED TO THE PATIENT WHEN THE DAMAGE WAS FOUND. IT WAS NOTICED WHEN THE TUBING WAS EXTENDED AND THE ANESTHESIA GAS BEGAN TO LEAK. A NEW CIRCUIT WAS OBTAINED AND USED IN PLACE OF THE FAULTY PRODUCT. AT THIS TIME, THE CUSTOMER HAS NOT PROVIDED ANY INFORMATION REGARDING PATIENT HARM OR INJURY ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16990310 ↗InjuryMay 24, 2023Leak/Splash; Material Puncture/Hole; Leak/Splash; Material Puncture/Hole
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE B1714XXX -ANES CIRCUIT, PED, 108 IN EXP, 1L BAG HAS HOLES IN THE CIRCUIT CAUSING AN AIR LEAK. THE PRODUCT WAS IN THE PROCESS OF BEING ATTACHED TO THE PATIENT WHEN THE DAMAGE WAS FOUND. IT WAS NOTICED WHEN THE TUBING WAS EXTENDED AND THE ANESTHESIA GAS BEGAN TO LEAK. A NEW CIRCUIT WAS OBTAINED AND USED IN PLACE OF THE FAULTY PRODUCT. AT THIS TIME, THE CUSTOMER HAS NOT PROVIDED ANY INFORMATION REGARDING PATIENT HARM OR INJURY ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16990309 ↗MalfunctionMay 24, 2023Crack; Material Puncture/Hole; Crack; Material Puncture/Hole
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE B1714XXX - ANES CIRCUIT, PED, 108 IN EXP, 1L BAG WERE NOT PASSING THE LEAK TESTS ON MACHINES DUE TO HOLES AND CRACKS IN THE CIRCUIT HOSE.THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16990298 ↗MalfunctionMar 21, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE B1714XXX ANES CIRCUIT, PED, 108 IN EXP, 1L BAG WAS LEAKING. THE PRODUCT WAS IN THE PROCESS OF BEING ATTACHED TO THE PATIENT WHEN THE DAMAGE WAS FOUND. IT WAS NOTICED WHEN THE TUBING WAS EXTENDED AND THE ANESTHESIA GAS BEGAN TO LEAK. A NEW CIRCUIT WAS OBTAINED AND USED IN PLACE OF THE FAULTY PRODUCT. THERE WAS NO HARM BUT THIS ISSUE CAUSE A DELAY OF TREATMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16583707 ↗MalfunctionMar 16, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE B1714XXX ANES CIRCUIT, PED, 108 IN EXP, 1L BAG WAS LEAKING. THE PRODUCT WAS IN THE PROCESS OF BEING ATTACHED TO THE PATIENT WHEN THE DAMAGE WAS FOUND. IT WAS NOTICED WHEN THE TUBING WAS EXTENDED AND THE ANESTHESIA GAS BEGAN TO LEAK. A NEW CIRCUIT WAS OBTAINED AND USED IN PLACE OF THE FAULTY PRODUCT. THERE WAS NO HARM BUT THIS ISSUE CAUSE A DELAY OF TREATMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16554710 ↗MalfunctionMar 16, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE B1714XXX ANES CIRCUIT, PED, 108 IN EXP, 1L BAG WAS LEAKING. THE PRODUCT WAS IN THE PROCESS OF BEING ATTACHED TO THE PATIENT WHEN THE DAMAGE WAS FOUND. IT WAS NOTICED WHEN THE TUBING WAS EXTENDED AND THE ANESTHESIA GAS BEGAN TO LEAK. A NEW CIRCUIT WAS OBTAINED AND USED IN PLACE OF THE FAULTY PRODUCT. THERE WAS NO HARM BUT THIS ISSUE CAUSE A DELAY OF TREATMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16554284 ↗InjuryJan 24, 2023Defective Component; Defective Component
THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THE ANESTHESIOLOGIST WAS UNABLE TO GIVE POSITIVE PRESSURE VENTILATION WITH THE ANESTHESIA MACHINE WHILE VITAL SIGNS¿ PEDIATRIC ANESTHESIA BREATHING CIRCUIT WAS IN USE. THE PATIENT WAS SUCTIONED, RE-INTUBATED, AND WAS CONTINUED TO BE VENTILATED BY BAG VALVE MASK. THE CUSTOMER CONFIRMED THAT NO HARM WAS DONE TO THE PATIENT BECAUSE THE ANESTHESIOLOGIST WAS QUICK TO GIVE MEDICAL INTERVENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16236151 ↗
Other Circuit, breathing (w connector, adaptor, y piece) devices
- Medline Industries, Inc. DYNJ905503FMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)3 recalls0 reports
- Medline Industries, Inc. PHS972096014BMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)3 recalls0 reports
- Medline Industries, Inc. DYNJ0185290GMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)3 recalls0 reports
- Medline Industries, Inc. DYNJ905503GMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)2 recalls0 reports
- Medline Industries, Inc. DYNJAA262AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)2 recalls0 reports
- Medline Industries, Inc. DYNJ905567AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)2 recalls0 reports
- MEDLINE INDUSTRIES, INC. DYNJ86657AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL199AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page