Airlife™ Sterile Water for Inhalation, USP, 500 ml 002620 (UDI 10190752141643)Recalls & reported problems
Vyaire Medical, Inc. · Sterile Water for Inhalation, USP, 500 ml · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10190752141643), received between Nov 21, 2022 and Dec 19, 2023 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Fitting Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fitting Problem3
- Material Integrity Problem2
- Difficult to Open or Close1
- Gas/Air Leak1
- Obstruction of Flow1
- Product Quality Problem1
- Smoking1
- Sparking1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Low Oxygen Saturation1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 19, 2023Fitting Problem; Fitting Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT DURING USE, THE HOSPITAL STAFF NOTICED THAT THE 002620 - HUMIDIFIER KIT 500ML 12/CS WAS EASILY DISENGAGED FROM THE OXYGEN GAUGE. THEY ALSO OBSERVED THAT IT CONSTANTLY ALARMS, AND THE PRESSURE IS BUILDING UP INSIDE THE BOTTLE, WHICH CAUSES IT TO EXPAND OR BLOAT. UPON INSPECTION, THEY NOTICED THAT THE OXYGEN PORT OPENING WAS TOO SMALL.THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18361414 ↗MalfunctionDec 19, 2023Fitting Problem; Fitting Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT DURING USE, THE HOSPITAL STAFF NOTICED THAT THE (B)(6) - HUMIDIFIER KIT 500ML 12/CS WAS EASILY DISENGAGED FROM THE OXYGEN GAUGE. THEY ALSO OBSERVED THAT IT CONSTANTLY ALARMS, AND THE PRESSURE IS BUILDING UP INSIDE THE BOTTLE, WHICH CAUSES IT TO EXPAND OR BLOAT. UPON INSPECTION, THEY NOTICED THAT THE OXYGEN PORT OPENING WAS TOO SMALL. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18361408 ↗MalfunctionDec 19, 2023Fitting Problem; Fitting Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT DURING USE, THE HOSPITAL STAFF NOTICED THAT THE 002620 - HUMIDIFIER KIT 500ML 12/CS WAS EASILY DISENGAGED FROM THE OXYGEN GAUGE. THEY ALSO OBSERVED THAT IT CONSTANTLY ALARMS, AND THE PRESSURE IS BUILDING UP INSIDE THE BOTTLE, WHICH CAUSES IT TO EXPAND OR BLOAT. UPON INSPECTION, THEY NOTICED THAT THE OXYGEN PORT OPENING WAS TOO SMALL. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18361355 ↗InjuryDec 19, 2023Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED TO VYAIRE MEDICAL THAT 2620 HUMIDIFIER KIT 500ML 12/CS IS PRODUCING 0 AIR SOUND FROM THE CANNULA EVEN WHEN IT IS ON 4 LITERS DURING PATIENT USE. THE PATIENTS SATURATION DIPPED BELOW 60. THE NURSE DISCONNECTED THE OXYGEN TUBING FROM THE HUMIDIFIER AND REPLACED THE HUMIDIFIER WITH A NEW ONE AND THE PATIENTS SATURATION INCREASED TO 78. THEY WERE ABLE TO WEAN 02 BACK DOWN TO 2L. THEY OBSERVED A LARGE CRACK IN THE PLASTIC WHERE OXYGEN MUST HAVE BEEN LEAKING OUT. FURTHERMORE, THERE WAS NO HARM ON THE PATIENT.
Patient problems: Low Oxygen Saturation
MAUDE report 18360734 ↗MalfunctionNov 15, 2023Product Quality Problem; Product Quality Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE 002620 - HUMIDIFIER KIT 500ML 12/CS WAS SO FULL THAT BUBBLES BACKED UP TO THE TUBING AND WATER BACKED UP IN THE CANNULA. ALSO, THE TUBING FOR OXYGEN WAS BEING FILLED WITH WATER, AND WATER DRIPS INTO THE NOSE OF THE PATIENT AND INTO THE CANNULA. THE PATIENT'S STATUS IS CURRENTLY UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18145419 ↗MalfunctionOct 5, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE NOZZLE WHERE THE NASAL CANNULA IS PLACED DOES NOT HAVE A HOLE TO ALLOW OXYGEN TO COME OUT, SO PRESSURE BUILDS UP IN THE 002620 - HUMIDIFIER KIT 500ML 12/CS AND WATER COMES OUT OF THE CONNECTOR ON TOP. THE SAID ISSUE OCCURRED DURING USE ON A PATIENT. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM/INJURY ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17880115 ↗MalfunctionJun 14, 2023Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED TO VYAIRE MEDICAL THAT WHEN THE BOTTLE OF HUMIDIFIER KIT 500ML 12/CS IS OPENED ON THE FRONT WITH A TWIST TAB, IT DOESN¿T ALWAYS COMPLETELY OPEN, AND SOMETIMES ISN¿T OBVIOUS WHEN IT'S HOOKED UP TO FLOW ON THE WALL. WHEN IT'S HOOKED UP AND TURNED ON, FLOW DOESN¿T COME OUT THE FRONT VERY WELL. THE CUSTOMER TRIED TO OPEN THE HOLE WITH A SCISSOR AND REMOVE IT FROM THE WALL, CAUSING A WATER FOUNTAIN OF BACK PRESSURE TO SPRAY ALL OVER THE STAFF MEMBER AND THE ROOM. THIS HAPPENS OCCASIONALLY AND NOT WITH ALL BOTTLES. ALSO, THE CUSTOMER IS UNABLE TO PUT MORE THAN 3 LITERS OF FLOW THROUGH THE SYSTEM BECAUSE IT POPS OFF AND WHEN DISCONNECTED, CAUSES THE FOUNTAIN TO OCCUR AND SPRAYS ALL OVER THE ROOM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17126507 ↗MalfunctionFeb 8, 2023Smoking; Obstruction of Flow; Sparking; Smoking; Obstruction of Flow; Sparking
IT WAS REPORTED TO VYAIRE MEDICAL THAT VYAIRE HUMIDIFIER WAS FOUND ON A PATIENT WITH WATER CASCADING FROM THE FLOWMETER SCREW TOP. THE NOZZLE THAT THE CANNULA TUBING ATTACHES TO IS OCCLUDED. AFTER THE REMOVAL OF THE "SNAP OFF" PLASTIC TUBE THE WATER CAUSED AN ELECTRICAL CORD TO SINGE, SPARK AND SMOKE. UPON EXAMINING THE "NOZZLE" IT IS OCCLUDED AND AIR CANNOT PASS THROUGH IT. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16328975 ↗MalfunctionNov 21, 2022Material Integrity Problem; Material Integrity Problem
THE CUSTOMER REPORTED TO VYAIRE MEDICAL THE HUMIDIFIER KIT 500ML 12/CS STERILE WATER INHALATION WAS NOT BUBBLING PROPERLY DURING PATIENT USE, AND STARTED TO SPOUT STERILE WATER FROM THE TOP DUE TO INCREASE PRESSURE. THE CUSTOMER CONFIRMED NO HARM WAS DONE TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15838067 ↗
Other HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE) devices
- Fisher & Paykel Healthcare PT101USFisher & Paykel Healthcare Limited · Humidifier, Respiratory Gas, (Direct Patient Interface)2 recalls5 deaths1,068 reports
- HUDSON RCI IPN913776Teleflex Incorporated · HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE)436 reports
- F&P Optiflow+ OPT944Fisher & Paykel Healthcare Limited · Humidifier, Respiratory Gas, (Direct Patient Interface)5 deaths392 reports
- Fisher & Paykel Healthcare PT100USFisher & Paykel Healthcare Limited · Humidifier, Respiratory Gas, (Direct Patient Interface)2 recalls59 reports
- MR850 MR850JHUFisher & Paykel Healthcare Limited · Humidifier, Respiratory Gas, (Direct Patient Interface)93 reports
- HUDSON RCI HUD42500Medline Industries, Inc. · HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE)87 reports
- F&P Optiflow Junior 2 OJR412Fisher & Paykel Healthcare Limited · Humidifier, Respiratory Gas, (Direct Patient Interface)80 reports
- F&P Airvo 900PT561Fisher & Paykel Healthcare Limited · Humidifier, Respiratory Gas, (Direct Patient Interface)1 death56 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page