Skip to content

AirLife™ 4893TRecalls & reported problems

Airlife · Tri-Flo™ Suction Cath-N-Glove™ Kit · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10190752148956), received between Feb 6, 2025 and Feb 6, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Packaging Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Packaging Problem1

Patient problems

  • Bacterial Infection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 6, 2025Packaging Problem; Packaging Problem

    SUNMED HOLDINGS LLC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE SECOND OF TWO REPORTS. REFER TO 8030673-2025-00007 FOR THE FIRST REPORT IT WAS REPORTED, THE SUCTION CATHETER WAS CONTAMINATED WITH A [UNKNOWN] BLACK MATERIAL. UPON DISCOVERY, A NEW CLOSED SUCTION SYSTEM WAS PLACED. [THE DISCOVERY MAY HAVE LEAD TO] A DELAY IN SUCTIONING. IT WAS ADDITIONALLY REPORTED, THE PATIENT WAS POSITIVE FOR SERRATIA (WATER BACTERIA) IN THEIR ENDOTRACHEAL TUBE. AT THE TIME OF THE REPORT, THE PATIENT WAS VENTED, AND IN ISOLATION FOR POSITIVE CULTURES FOR SERRATIA.

    Patient problems: Bacterial Infection

    MAUDE report 21329455 ↗

Other Airway suction kit devices

All Airway suction kit devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page