- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10192253007855), received between Oct 16, 2024 and May 18, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Gastrointestinal Regurgitation1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 18, 2026Break
THE CUSTOMER REPORTED A TEAR IN THE INSERTED PORTION OF THE NG TUBE WHICH WAS NOTED WHEN THE TUBE WAS REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25209134 ↗MalfunctionMay 1, 2026Material Separation
THE CUSTOMER REPORTED THE PATIENT ATTEMPTED TO PULL OUT THE IRIS TUBE AND THE NURSE INTERVENED BY GRASPING THE TUBE NEAR THE NARE TO PREVENT REMOVAL OF BOTH THE TUBE AND THE BRIDLE. DURING THIS EVENT, THE TUBE SPLIT AT THE 84CM MARK BETWEEN THE POINT WHERE THE NURSE WAS HOLDING IT AND WHERE THE PATIENT WAS PULLING. NO HARM TO THE PATIENT REPORTED. PER ADDITIONAL INFORMATION PROVIDED, THE REMAINDER OF THE TUBE WAS REMOVED BY HAND, NO ADDITIONAL PROCEDURES WERE PERFORMED. A NEW TUBE WAS NOT PLACED AFTER THE REMOVAL OF THIS TUBE AS THE PATIENT WAS COMBATIVE AND THE MEDICAL TEAM OPTED NOT TO REINSERT ANOTHER TUBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25057542 ↗MalfunctionApr 17, 2026Break
THE CUSTOMER REPORTED THE IRIS TUBE WAS BEING REMOVED AND THE NURSE NOTICED A BREAK IN THE TUBING. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24916319 ↗MalfunctionOct 16, 2024Break; Break
THE CUSTOMER REPORTED THAT DOBHOFF TUBE RUPTURED LONGITUDINALLY APPROXIMATE 2 CM IN LENGTH. ENTERAL FEEDS WERE REFLUXING OUT OF THE PATIENT'S LEFT NARE. PER CUSTOMER, THE TUBE WAS REMOVED AND REPLACED. ADDITIONAL INFORMATION WAS RECEIVED AND STATED THAT THERE WAS NO MEDICAL INTERVENTION OR TREATMENT PROVIDED. THE PATIENT IS DOING OK.
Patient problems: Gastrointestinal Regurgitation
MAUDE report 20464464 ↗
Other Tubes, gastrointestinal (and accessories) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-05055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page