CARDINAL HEALTH 8884711006ERecalls & reported problems
Cardinal Health 200, LLC · DOBBHOFF 10FR 43IN W/STYLET With ENFit Connectors · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10192253011258), received between Jul 27, 2026 and Sep 24, 2026 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Improper or Incorrect Procedure or Method1
Patient problems
- Dehydration1
- Hyponatremia1
- Needle Stick/Puncture1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2026Detachment of Device or Device Component
Patient problems: Needle Stick/Puncture
MAUDE report 50890125 ↗InjuryJul 27, 2026Improper or Incorrect Procedure or Method
The customer reported the patient was found to have a Dobhoff feeding tube with the stylet still in place several days after placement which led to dehydration and hyponatremia. Per the customer, while the metal stylet is essential for placement of the tube, removal of the stylet is a challenge. The stylet plastic tip is the very same color as the ports on the feeding tube and can be difficult to differentiate easily for users. There are two ports, so inexperienced users may think the stylet is supposed to be there and use the smaller port, restricting the flow rate of feeds and fluids. Per additional information provided on July 21, 2026, the treatment for dehydration was IV fluids x 24hours. There was no treatment provided for the hyponatremia which resolved prior to discharge, monitoring only. The tube was in place for 5 days before the stylet was found to still be in the tube. The stylet was removed and the tube remained in place with placement confirmed by x-ray.
Patient problems: Dehydration; Hyponatremia
MAUDE report 50000546 ↗
Other Tubes, gastrointestinal (and accessories) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-05090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- Kangaroo 8884711253ECardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)100 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page