Skip to content

CARDINAL HEALTH 8884720825ERecalls & reported problems

Cardinal Health 200, LLC · ENTRIFLEX 8FR 36 INCH WITH STYLET and With ENFit Connectors · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10192253011654), received between Nov 24, 2025 and Feb 10, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Difficult to Remove1

Patient problems

  • Foreign Body In Patient1
  • Sore Throat1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 10, 2026Difficult to Remove

    THE CUSTOMER REPORTED THEY HAVE SIGNIFICANT PROBLEMS REMOVING THE MANDREL DURING EACH INSTALLATION. THIS HAS RESULTED IN TREATMENT FAILURES AND COMPLAINTS FROM FAMILIES. THEY EVEN TRIED REMOVING THE MANDREL BEFORE INSTALLATION, BUT THIS WAS NOT SUCCESSFUL. ACCORDING TO THE CUSTOMER, THE PACKAGING DOES NOT INCLUDE ANY INFORMATION THAT WOULD ENABLE THEM TO PROVIDE SAFE AND EFFECTIVE CARE. THE CLINICIAN NURSE STATED THEY PUT THIS KANGAROO TUBING IN 2-3 TIMES A YEAR AND IT IS USUALLY APPLIED BY A GASTROENTEROLOGIST BUT ACCORDING TO THEIR DECISION, IT IS NOW APPLIED BY THE NURSE. THE LAST TIME WAS LAST WEEK, ON A CHILD, TWO YEARS OLD, AROUND 10 KG. WHEN THE KANGAROO TUBING WAS INSERTED, IT WAS IMPOSSIBLE TO REMOVE THE MANDREL. DURING 48 HOURS, THE CHILD HAD NOT BEEN FED THROUGH ANY TUBING. THE CHILD HAS AN IV DRIP TO SUPPLY WHAT WAS NEEDED. AFTER A FEW DAYS, THEY PUT IN A NORMAL NASOGASTRIC TUBE. NURSES HAVE ALSO MADE A TEST WITH THREE SAMPLES, OUT OF THE PATIENT, AND IT WAS NOT POSSIBLE TO REMOVE THE MANDREL, EXCEPT FOR FROM ONE TUBE ONLY. PER ADDITIONAL INFORMATION PROVIDED, THE CUSTOMER DID NOT ACTIVATE THE HYDROMER PRIOR TO OR AFTER PLACEMENT. THE PATIENT'S DIGESTIVE TRACT WAS IRRIT

    Patient problems: Sore Throat

    MAUDE report 24310512 ↗
  • MalfunctionNov 24, 2025Break; Break

    THE CUSTOMER REPORTED THE 8FR NG TUBE TIP BROKE OFF WHILE IN PLACE. THE MOM NOTED RESISTANCE 1.5 WEEKS AGO FOLLOWING ADMINISTRATION OF MEDICATION BUT WAS EVENTUALLY ABLE TO FLUSH THE TUBE WITH WATER AGAIN. NO FURTHER ISSUES WERE NOTED BY THE FAMILY BUT UPON CT SCAN REVIEW BY RADIOLOGIST ON (B)(6), THE TIP OF NG TUBE WAS SEEN IN THE SMALL INTESTINE. THE PATIENT HAS NOT YET PASSED THE TIP OF NG TUBE BUT THE REMAINDER OF THE TUBE WAS REMOVED IN THE CLINIC. PER ADDITIONAL INFORMATION PROVIDED ON 24NOV2025, THE CT SCAN WAS PERFORMED ON (B)(6), NOT (B)(6) WHICH WAS INITIALLY REPORTED IN ERROR. THIS WAS A ROUTINE SCAN FOR MONITORING PURPOSES WITH REGARDS TO THE PATIENT'S MEDICAL HISTORY. AS OF LAST THURSDAY, THE TIP OF THE NG TUBE HAD NOT PASSED YET. PER FURTHER INFORMATION PROVIDED ON 02DEC2025, AS OF (B)(6), THE TIP OF THE NG TUBE HAD STILL NOT PASSED. PER ADDITIONAL INFORMATION PROVIDED ON 11DEC2025, THE TIP HAS NOT PASSED YET. THE CURRENT PLAN AS PER SURGERY IS TO WAIT AND LET IT PASS ON ITS OWN.

    Patient problems: Foreign Body In Patient

    MAUDE report 23638820 ↗

Other Tubes, gastrointestinal (and accessories) devices

All Tubes, gastrointestinal (and accessories) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page