CARDINAL HEALTH 8884720858ERecalls & reported problems
Cardinal Health 200, LLC · ENTRFLX 8FR 43IN W/STYLT&Y SITE With ENFit Connectors · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10192253011715), received between Dec 8, 2025 and Sep 25, 2026 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Material Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation2
- Break1
- Difficult to Remove1
Patient problems
- Foreign Body In Patient3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjurySep 25, 2026Material Separation
The customer reported the end remained in the patient post removal and needed to be removed via endoscopy.
Patient problems: Foreign Body In Patient
MAUDE report 50900378 ↗InjuryMay 8, 2026Material Separation
THE CUSTOMER REPORTED THAT THE DOBHOFF TIP NG TUBE BROKE INTO TWO PIECES WHILE INDWELLING IN THE PATIENT. A PORTION OF THE FEEDING TUBE WAS RETAINED WITHIN THE PATIENT. THE BREAK OF THE TUBE OCCURRED AT THE AREA OF 5.6-8CM ON THE TUBE. THE TUBE WAS INDWELLING FOR 2 DAYS BEFORE BREAK. THE REMAINING PORTION OF TUBE WAS REMOVED ENDOSCOPICALLY. FEED HISTORY: JEVITY 1.5 CAL. WATER HISTORY: 11/13 30 ML Q4H: 1522, 1742, 2059, 11/14 50 ML Q4H: 0128, 0550, 0930, 1302, 1706, 2108, 11/15 100 ML Q4H: 0130, 0520, 0913, 1316, 1811, 2200. MEDICATION HISTORY: AMIODARONE, ASPIRIN, LIPITOR, LISINOPRIL, LOPRESSOR, MULTIVITAMIN, OXYCODONE, GYLCOLAX, SENAKOT-S, BRILINTA, TRAZADONE, POTASSIUM CHLORIDE. SUPPLEMENT HISTORY: PROSOURCE 60 ML TID, BANATROL 60 ML BID.
Patient problems: Foreign Body In Patient
MAUDE report 25126773 ↗MalfunctionApr 30, 2026Difficult to Remove
THE CUSTOMER REPORTED AN ISSUE WITH THE NASOGASTRIC TUBES. SEVERAL NURSES HAVE REPORTED A PROBLEM WHEN REMOVING THE GUIDE WIRE AFTER THE TUBE HAS BEEN INSERTED. THEY STATED THAT THEY PROPERLY MOISTENED THE TUBE BEFOREHAND. THE REPORTER STATED THEY TESTED A TUBE THEMSELF, AND THE GUIDE SLIDES CORRECTLY AFTER HUMIDIFICATION. PER ADDITIONAL INFORMATION PROVIDED ON (B)(6) 2026, THE ISSUE THE NURSES WERE EXPERIENCING WAS THAT THEY WERE COMPLETELY UNABLE TO REMOVE THE STYLET. IT WAS STATED THAT THEY PUT WATER IN THE TUBE BUT THE PROCESS USED AND QUANTITY OF WATER INJECTED IS NOT KNOWN. TO RESOLVE THE ISSUES, THE FEEDING TUBE HAS BEEN REMOVED AND REPLACED. NO FURTHER INFORMATION IS AVAILABLE REGARDING SPECIFIC INCIDENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25053389 ↗MalfunctionDec 8, 2025Break; Break
THE CUSTOMER REPORTED THE SECTION OF THE TUBE WITH THE WEIGHT DETACHED FROM THE TUBE AND REMAINED IN THE PATIENT'S STOMACH. A CONSULTATION HAD TO BE CARRIED OUT WITH ANESTHESIA AND GASTROENTEROLOGY FOR FOLLOW UP. THEY ARE AWAITING EVALUATION FOR SEDATION WITH GASTROSCOPY OR SURGERY UNDER GENERAL ANESTHESIA. PER ADDITIONAL INFORMATION PROVIDED ON NOVEMBER 25, 2025, THEY ARE WAITING FOR THE END OF THE TUBE TO COME OUT THROUGH THE NATURAL CHANNELS. THEY TAKE X-RAYS REGULARLY TO FOLLOW THE PATIENT'S JOURNEY. THE PATIENT IS STABLE AND DOING WELL. HE SHOULD BE TRANSFERRED TO PEDIATRICS SHORTLY.
Patient problems: Foreign Body In Patient
MAUDE report 23742947 ↗
Other Tubes, gastrointestinal (and accessories) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-05090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- Kangaroo 8884711253ECardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)100 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page