Kendall SCD 74021 (UDI 10192253016147)Recalls & reported problems
Cardinal Health 200, LLC · Kendall SCD Sequential Compression Comfort Sleeves, Knee Length, Small · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10192253016147), received between Nov 13, 2024 and Jul 25, 2025 — classified as 2 injuries.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
Patient problems
- Hematoma1
- Hypoesthesia1
- Nerve Damage1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
CUSTOMER REPORTS: A HEMATOMA WAS DISCOVERED DURING THE USE OF SCD700, AND AN INCISION WAS REQUIRED FOR RETRIEVAL. I HAVE BEEN USING IT SINCE (B)(6) AND NOTICED THAT A HEMATOMA HAD FORMED ON THE MORNING OF (B)(6). THE PATIENT HAD EDEMA EVEN BEFORE USING THE SCD700 SYSTEM. IT IS UNKNOWN HOW MUCH EDEMA IT WAS. CURRENTLY, THE TREATMENT IS OVER AND THE PATIENT'S CONDITION IS STABLE, AND SCD IS NOT BEING USED. IT SEEMS THAT THE ALARM DOES NOT GO OFF DURING USE, BUT I ASKED FOR AN INVESTIGATION OF THE MAIN UNIT AND CONSUMABLES TO SEE IF THERE IS A POSSIBILITY THAT THE PRESSURE HAS TEMPORARILY INCREASED. THE USAGE PERIOD WAS 9 DAYS.
Patient problems: Hematoma
MAUDE report 22617878 ↗InjuryNov 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
CUSTOMER REPORTS THAT THE USE OF THE SCD COMFORT SLEEVE CAUSED PERONEAL NERVE PALSY. ON TUESDAY, (B)(6), SHE UNDERWENT SPINE SURGERY AT THE ORTHOPEDIC DEPARTMENT AND RETURNED TO THE WARD. DECREASED MANUAL MUSCLE STRENGTH MEASUREMENT (MMT) AND HYPOESTHESIA UPON RETURNING TO THE ROOM. PERHAPS BECAUSE THE WOUND IS PAINFUL, PAINKILLERS ARE USED IN THE OPERATING ROOM. AFTER 3 HOURS, STRONG COMPLAINTS OF HYPESTHESIA. THE DOCTOR'S REPORT WAS THAT THE PATIENT WAS DIAGNOSED WITH PERONEAL NERVE PALSY. A PLANTAR SLEEVE IS USED DURING SURGERY, AND A SLEEVE BELOW THE KNEE IS USED AFTER SURGERY. THE FOOT PUMP WAS TURNED ON AND STARTED AFTER OBSERVATION OF VITAL SIGNS AND THE LIKE. AT THE 3 HOUR TIME OF OBSERVATION, THERE WAS MMT, SENSORY CONFIRMATION, AND PAIN, AND SINCE BLOCK INJECTIONS WERE NOT GIVEN, PAIN RELIEF WAS PERFORMED. FOUR HOURS LATER, IT WAS REPORTED TO THE DOCTOR, AND DURING THE EXAMINATION, HE POINTED OUT THAT THE LOWER LIMB SLEEVE WAS COMPRESSING THE FIBULA HEAD AND HE NOTICED THAT COMPRESSION IT WAS FROM THE LOWER LIMB SLEEVE OF THE FOOT PUMP. THE CAUSE WAS DETERMINED TO BE "INAPPROPRIATE USE OF A FOOT PUMP TO PREVENT VENOUS EMBOLISM". AS A COUNTERMEASURE, THE PUBLIC WAS INFORME
Patient problems: Hypoesthesia; Nerve Damage
MAUDE report 20678414 ↗
Other SLEEVE, LIMB, COMPRESSIBLE devices
- Kendall 84021Cardinal Health 200, LLC · SLEEVE, LIMB, COMPRESSIBLE2 recalls0 reports
- Kendall SCD 29525Cardinal Health, Inc. · Sleeve, Limb, Compressible56 reports
- RAW MATERIAL 30MDjo, LLC · SLEEVE, LIMB, COMPRESSIBLE52 reports
- Kendall 84021BCardinal Health 200, LLC · SLEEVE, LIMB, COMPRESSIBLE1 recall0 reports
- AIRCAST 3011-PLDjo, LLC · SLEEVE, LIMB, COMPRESSIBLE1 recall0 reports
- Kendall 84021PCardinal Health 200, LLC · SLEEVE, LIMB, COMPRESSIBLE1 recall0 reports
- N/AStryker Sustainability Solutions, Inc. · SLEEVE, LIMB, COMPRESSIBLE1 recall0 reports
- N/AStryker Sustainability Solutions, Inc. · SLEEVE, LIMB, COMPRESSIBLE1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.