Skip to content

Argyle 8888505115 (UDI 10192253038972)Recalls & reported problems

Cardinal Health 200, LLC · Yankauer Open Tip, No Vent · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10192253038972), received between Aug 20, 2025 and Sep 11, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED THAT THE DEVICE BROKE OFF WHILE BEING USED DURING A TKA (TOTAL KNEE ARTHROPLASTY). ADDITIONAL INFORMATION RECEIVED ON 22AUG2025 STATED THAT IT WAS THE TIP OF THE YANKAUER THAT BROKE OFF. IT WAS VISIBLE IN THE WOUND AND REMOVED WITHOUT COMPLICATION. NO INTERVENTION OR TREATMENT REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23025887 ↗
  • MalfunctionAug 20, 2025Break

    BROKE OFF WHILE BEING USED DURING A TKA [TOTAL KNEE ARTHROPLASTY].

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22850257 ↗

Other APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED devices

All APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.