Argyle 8888160333 (UDI 10192253040302)Recalls & reported problems
Cardinal Health 200, LLC · Argyle Polyurethane Umbilical Vessel Catheter Single Lumen 3.5 Fr/Ch (1.2 mm) x 9-8/10"" (25 cm) 0.23 mL Prime Volume" · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10192253040302), received between Mar 24, 2025 and Sep 11, 2026 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Connection Problem” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- May 2, 2025Open, Classified
Packaging defect may compromise sterility of the product. Use of a non-sterile product on the patient could lead to infection in neonates.
Z-1872-2025 · Root cause: Package design/selection · FDA record ↗
Most reported problems
Device problems
- Connection Problem4
- Break2
- Fluid/Blood Leak2
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Foreign Body In Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 11, 2026Connection Problem
The customer reported that the umbilical catheter end did not secure properly to the stopcock. One retained sample reportedly appeared to be missing part of the lip at the catheter end. The clinical staff reported difficulty achieving a secure connection with the stopcock. Additional information was received from the customer and stated that the issue seemed to occur while the line was on. It appeared the lip at the end of the catheter may have been stripped away. There was no harm reported. Unfortunately, the customer did not have any more additional information.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50731551 ↗MalfunctionDec 10, 2025Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED AIR/OXYGEN IN THE LINE. ADDITIONAL INFORMATION WAS RECEIVED FROM THE CUSTOMER AND STATED THAT THEY DID NOT OBSERVE ANY CRACKS, HOLES OR BREAKS IN THE LINE. THERE WAS NO HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23771353 ↗MalfunctionSep 11, 2025Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THAT THE CATHETER NOTED TO BE LEAKING IV FLUIDS AT THE CONNECTION BETWEEN THE CATHETER AND THE BUTTERFLY PLASTIC CLIP PORTION. THERE WAS NO HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23025884 ↗MalfunctionJun 12, 2025Break; Connection Problem; Break; Connection Problem
THE CUSTOMER REPORTED THAT THREADING ON PORTION OF THE CATHETER WHERE THE SYRINGE OR CONNECTORS INSERT WAS BROKEN OFF, THE CONNECTOR COULD NOT BE SECURED TO CATHETER. IT LOOKS AS IF THE UMBILICAL LINE APPARENTLY FAILED DUE TO THE THREADING ON THE CATHETER BEING CHANGED, AS THEY ARE NOW TABS INSTEAD OF THREADS. THERE IS NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22195333 ↗MalfunctionApr 25, 2025Connection Problem; Connection Problem
THE CUSTOMER REPORTED DEFECTIVE CONNECTOR. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21909355 ↗MalfunctionMar 28, 2025Connection Problem; Connection Problem
THE CUSTOMER REPORTED THAT THE LUR-LOCK END OF UMBILICAL ARTERIAL CATHETER (UAC) WOULD NOT SCREW TIGHT TO TRANSDUCER END (SPINNING IN CONTINUOUS CIRCLE WITHOUT LOCKING). THE CUSTOMER TRIED ANOTHER TRANSDUCER AND SAME PROBLEM OCCURRED. THE CUSTOMER FURTHER STATED THAT THEY ATTACHED ANESTHESIA EXTENSION TUBING TO UAC ON ONE END AND THE OTHER TO TRANSDUCER. THERE WAS NO HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21724827 ↗InjuryMar 24, 2025Break; Break
THE CUSTOMER REPORTED THAT THE NURSE PRACTITIONER (NP) CHANGED THE MALPOSITIONED UMBILICAL VESSEL CATHETER (UVC). THE OLD CATHETER WAS REMOVED, AND A NEW CATHETER WOULD NOT PASS THROUGH LIVER TO IVC/ ATRIAL JUNCTION. THE CATHETER WAS RETRACTED TO LOW LYING AND SUTURED AS APPROXIMATELY 3CM. THE NP WAS FLUSHING/DRAWING FROM LINE AND NOTICED AIR BUBBLES. NP ATTEMPTED TO REMOVE THE LINE, AND THE CATHETER ¿SNAPPED¿ LEAVING APPROXIMATELY 3 CM OF CATHETER IN BABY WITH THE END NOT VISIBLE AT CORD STUMP. THERE WAS NO BLOOD LOSS REPORTED, STAT BABYGRAM WAS ORDERED AND SURGERY WAS CONSULTED. THE RADIOGRAPH DEMONSTRATED THAT THE CATHETER SEGMENT WAS BELOW THE LIVER, BUT NOT IN CORD STUMP. THE BABY MADE NIL PER OS (NPO) AND PERIPHERAL INTRAVENOUS LINE (PIV) X 2 WAS PLACED. THE BABY WAS AIR TRANSPORTED TO LEVINE CHILDREN¿S HOSPITAL IN CHARLOTTE, NORTH CAROLINA FOR PEDIATRIC INTERVENTIONAL SURGERY TO RETRIEVE RETAINED PIECE OF CATHETER. ADDITIONAL INFORMATION WAS RECEIVED FROM THE CUSTOMER AND STATED THAT THE INITIALLY MALPOSITIONED UMBILICAL VESSEL CATHETER WAS ALSO (B)(4) UMBILICAL CATHETER SNGL LUMEN 3.5FR AND THE SAME BATCH NUMBER 2431800139. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN. THE
Patient problems: Foreign Body In Patient
MAUDE report 21680349 ↗
Other CATHETER, UMBILICAL ARTERY devices
- Argyle 8888160333Cardinal Health, Inc. · CATHETER, UMBILICAL ARTERY2 deaths66 reports
- Argyle 8888160556Cardinal Health 200, LLC · CATHETER, UMBILICAL ARTERY1 recall1 report
- Argyle 8888160341Cardinal Health 200, LLC · CATHETER, UMBILICAL ARTERY1 recall1 report
- Argyle 8888160119Cardinal Health 200, LLC · CATHETER, UMBILICAL ARTERY1 recall0 reports
- Argyle 8888160648Cardinal Health 200, LLC · CATHETER, UMBILICAL ARTERY1 recall0 reports
- Argyle 8888160325Cardinal Health 200, LLC · CATHETER, UMBILICAL ARTERY1 recall0 reports
- Argyle 8888160341Cardinal Health, Inc. · CATHETER, UMBILICAL ARTERY36 reports
- Argyle 8888160531Cardinal Health, Inc. · CATHETER, UMBILICAL ARTERY23 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.