Skip to content

NAMIC 60131554Recalls & reported problems

Medline Industries, Inc. · MTS LEFT HEART KIT-MERITER HOSP - PG · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10193489046939), received between May 16, 2024 and Sep 25, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (4 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 4 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 24, 2026Open, Classified

    Medline has identified the presence of particulate within the fluid path of the Manifolds.

    Z-2145-2026 · Root cause: Under Investigation by firm · FDA record ↗

  2. Feb 27, 2026Open, Classified

    Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

    Z-1718-2026 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Connection Problem4
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 25, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50904258 ↗
  • MalfunctionMay 16, 2024Connection Problem; Connection Problem

    SYRINGE WOULD NOT STAY ATTACHED TO THE MANIFOLD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19336796 ↗
  • MalfunctionMay 16, 2024Connection Problem; Connection Problem

    SYRINGE WOULD NOT STAY ATTACHED TO THE MANIFOLD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19336702 ↗
  • MalfunctionMay 16, 2024Connection Problem; Connection Problem

    SYRINGE WOULD NOT STAY ATTACHED TO THE MANIFOLD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19333551 ↗
  • MalfunctionMay 16, 2024Connection Problem; Connection Problem

    SYRINGE WOULD NOT STAY ATTACHED TO THE MANIFOLD

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19332916 ↗

Other Cardiovascular procedure kit devices

All Cardiovascular procedure kit devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page