Skip to content

Medline ENFIT331820Recalls & reported problems

Medline Industries, Inc. · BALLOON BUTTON,MEDLINE,ENFIT,18FR,2.0CM · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10193489143676), received between Mar 31, 2026 and Mar 31, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Leak/Splash” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 31, 2026Leak/Splash

    PATIENT CALLED STATING, "THAT THE BALLOON HAS A LEAKAGE AND IS CAUSING HER TO GET IRRITATION CAN WE REPLACE HER THIS ITEM."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24746729 ↗

FDA clearance and classification

The FDA regulates product code PSH (percutaneous replacement gastrostomy tube kit) under 21 CFR 876.5980 as a Class II device. 21 CFR 876.5980 ↗

Common questions about the Medline ENFIT331820

Has the Medline ENFIT331820 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10193489143676), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Medline ENFIT331820?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Mar 31, 2026. That report lists device problem “Leak/Splash” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Medline ENFIT331820?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Medline ENFIT331820 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Medline ENFIT331820, and is it still sold?

The FDA's device database lists Medline Industries, Inc. as the company responsible for it (brand Medline, model ENFIT331820). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page