Medline Industries, Inc. DYNJ907168ARecalls & reported problems
Medline Industries, Inc. · LITHOTOMY · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- May 4, 2026Class IIOpen, Classified
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Z-2497-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Other General surgery tray devices
- Halyard MSBM036-03_0012Avid Medical, Inc. · General surgery tray5 recalls0 reports
- Medline Industries, Inc. DYNJ58534Medline Industries, Inc. · General surgery tray5 recalls0 reports
- Medline Industries, Inc. DYNJ82007AMedline Industries, Inc. · General surgery tray4 recalls0 reports
- Medline Industries, Inc. DYNJ42382BMedline Industries, Inc. · General surgery tray4 recalls0 reports
- Centurion ART690Centurion Medical Products Corporation · General Surgery Tray4 recalls0 reports
- Medline Industries, Inc. DYNJ44866PMedline Industries, Inc. · General surgery tray4 recalls0 reports
- Medline Industries, Inc. DYNDL1350AMedline Industries, Inc. · General surgery tray4 recalls0 reports
- Medline Industries, Inc. DYNJ69692Medline Industries, Inc. · General surgery tray4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.