Skip to content

Medline Industries, Inc. DYNJ907880ARecalls & reported problems

Medline Industries, Inc. · HAND KIT · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jan 7, 2026Class IIOpen, Classified

    Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

    Z-2114-2026 · Root cause: Process control · Recall details · FDA record ↗

  2. Nov 17, 2022Class IIOpen, Classified

    Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.

    Z-1096-2023 · Root cause: Device Design · Recall details · FDA record ↗

Other Orthopedic tray devices

All Orthopedic tray devices →