Skip to content

Medline Industries, Inc. CDS985504GRecalls & reported problems

Medline Industries, Inc. · OPEN HEART · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10195327150822), received between Nov 8, 2022 and Nov 8, 2022 — classified as 1 injury.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. May 4, 2026Class IIOpen, Classified

    During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

    Z-2487-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 8, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    ACCORDING TO THE FACILITY ON (B)(6) 2022 DURING AN OPEN HEART SURGICAL CASE THE RAYTEC SPONGE FRACTURED DURING USE AND WAS LEAVING SPONGE FRAGMENTS IN THE PATIENT THAT NEEDED TO BE REMOVED BY THE SURGEON.

    Patient problems: Foreign Body In Patient

    MAUDE report 15754983 ↗

Other Cardiovascular procedure kit devices

All Cardiovascular procedure kit devices →

Sources and updates