Skip to content

Medline Industries, Inc. DYNJ68676BRecalls & reported problems

Medline Industries, Inc. · ACDF NEURO PACK · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Mar 19, 2026Class IIOpen, Classified

    During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

    Z-2038-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Nov 21, 2024Class IIOpen, Classified

    Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination.

    Z-0794-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  3. Nov 17, 2022Class IIOpen, Classified

    Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.

    Z-1096-2023 · Root cause: Device Design · Recall details · FDA record ↗

Other Orthopedic tray devices

All Orthopedic tray devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page