Medline Industries, Inc. DYNJ42903DRecalls & reported problems
Medline Industries, Inc. · ANGIO PACK · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10195327171117), received between Aug 11, 2026 and Aug 11, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Adhesive Too Strong” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 18, 2023Class IIOpen, Classified
Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
Z-2160-2023 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Adhesive Too Strong1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 11, 2026Adhesive Too Strong
It was reported that "Today we have had several instances of the femoral drape being very sticky and difficulty to unfold."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50213511 ↗
Other Cardiac catheterization kit devices
- Medline Industries, Inc. DYNJ44700JMedline Industries, Inc. · Cardiac catheterization kit3 recalls1 report
- Medline Industries, Inc. ECVC2555Centurion Medical Products Corporation · Cardiac catheterization kit3 recalls0 reports
- Medline Industries, Inc. DYNJ83089Medline Industries, Inc. · Cardiac catheterization kit3 recalls0 reports
- Medline Industries, Inc. DYNJ0578916XMedline Industries, Inc. · Cardiac catheterization kit3 recalls0 reports
- Medline Industries, Inc. MNS9935Medline Industries, Inc. · Cardiac catheterization kit3 recalls0 reports
- Medline Industries, Inc. ECVC5540ACenturion Medical Products Corporation · Cardiac catheterization kit3 recalls0 reports
- Medline Industries, Inc. ECVC2745Centurion Medical Products Corporation · Cardiac catheterization kit3 recalls0 reports
- Medline Industries, Inc. ECVC6505Centurion Medical Products Corporation · Cardiac catheterization kit3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page