Skip to content

Medline Industries, Inc. DYNJ63673DRecalls & reported problems

Medline Industries, Inc. · GLOVE PACK 7.0 (15PK) · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10195327304843), received between Apr 29, 2026 and Apr 29, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Puncture/Hole” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 27, 2026Class IOpen, Classified

    Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

    Z-1725-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Puncture/Hole1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2026Material Puncture/Hole

    IT WAS REPORTED THAT THE STERILE GLOVES TEAR EASILY DURING DONNING. ADDITIONALLY, THE GLOVES WERE REPORTED TO ROLL DOWN AT THE WRIST, ALLOWING FLUID TO BREACH THE BARRIER AND ENTER THE INTERIOR OF THE GLOVE, RESULTING IN A POTENTIAL RISK OF CONTAMINATION. THIS ISSUE WAS REPORTED TO OCCUR CONSISTENTLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25029025 ↗

Other Surgeon's gloves devices

All Surgeon's gloves devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page