Medline Industries, Inc. DYNJ64040DRecalls & reported problems
Medline Industries, Inc. · HEART CABG PK · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10195327391867), received between Sep 24, 2026 and Sep 24, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Device Slipped” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 4, 2026Class IIOpen, Classified
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Z-2487-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Slipped1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2026Device Slipped
It was reported that in the last two weeks the clamp has slipped off of the aorta 3 times for 2 different surgeons in the middle of the procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50885418 ↗
Other Cardiovascular procedure kit devices
- Medline Industries, Inc. CDS840396YMedline Industries, Inc. · Cardiovascular procedure kit6 recalls0 reports
- Medline Industries, Inc. CDS983513FMedline Industries, Inc. · Cardiovascular procedure kit6 recalls0 reports
- Medline Industries, Inc. CDS840396XMedline Industries, Inc. · Cardiovascular procedure kit5 recalls0 reports
- Medline Industries, Inc. CDS984355GMedline Industries, Inc. · Cardiovascular procedure kit5 recalls0 reports
- Medline Industries, Inc. DYNJ51038RMedline Industries, Inc. · Cardiovascular procedure kit4 recalls2 reports
- NAMIC 60033123Medline Industries, Inc. · Cardiovascular procedure kit4 recalls0 reports
- Medline Industries, Inc. DYNJ905079FMedline Industries, Inc. · Cardiovascular procedure kit4 recalls0 reports
- Medline Industries, Inc. DYNJ0376608OMedline Industries, Inc. · Cardiovascular procedure kit4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page