Skip to content

CARDINAL HEALTH SMA13CCEMZRecalls & reported problems

Cardinal Health 200, LLC · C SECTION PACK · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10197106788530), received between Apr 22, 2025 and Apr 22, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Twisted/Bent” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 22, 2025Material Twisted/Bent

    FROM STAFF: C/S PACK BULB SYRINGES FOR INFANT SUCTIONING AT DELIVERY-IS BENT AND DOES NOT WORK-BOTH IN PACK. 3RD C/S NOTED THIS WEEK WHERE BULB SYRINGES PROVIDED NONFUNCTIONAL AND NOT ABLE TO DETECT UNTIL TIME OF USE. FURTHER INSPECTION SHOWS STEM SLIGHTLY BENT THAT PREVENTS ADEQUATE USE. 3RD BULB SYRINGE PACKAGED SEPARATELY OPENED AND USED AND ADEQUATELY WORKS-COMPARED 2 AFTER DELIVERY AND DISCREPANCY NOTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21881306 ↗

Other Cesarean section tray devices

All Cesarean section tray devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page