Skip to content

MEDLINE 81472RHRecalls & reported problems

Medline Industries, Inc. · SJM INQUIRY CATH STRBL MD 4POLE 2/5/2 5F · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K242225); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Jul 31, 2025Class IOpen, Classified

    Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

    Z-2612-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

FDA clearance and classification

Cleared through 510(k) K242225 on Nov 29, 2024 (decision: Substantially Equivalent). Submitted by Surgical Instrument Service and Savings, Inc. as “Medline ReNewal Reprocessed Abbott Inquiry Steerable Diagnostic Catheter”. The FDA review took 123 days. FDA record ↗

The FDA regulates product code NLH (catheter, recording, electrode, reprocessed) under 21 CFR 870.1220 as a Class II device. 21 CFR 870.1220 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K242225. One submission often covers many models, so check the product and lot details in each record.

  • Z-1320-2026 · Dec 22, 2025 · Open, Classified
    Reprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81105RH 81172RH; Abbott Inquiry Steerable Diagnostic Cat

Common questions about the MEDLINE 81472RH

Has the MEDLINE 81472RH been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2612-2025 (Class I), began on Jul 31, 2025: Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the MEDLINE 81472RH FDA approved?

It was cleared through 510(k) K242225 on Nov 29, 2024. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the MEDLINE 81472RH, and is it still sold?

The FDA's device database lists Medline Industries, Inc. as the company responsible for it (brand MEDLINE, model 81472RH). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, recording, electrode, reprocessed devices

All Catheter, recording, electrode, reprocessed devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page