Medline Industries, Inc. DYNJ901046MRecalls & reported problems
Medline Industries, Inc. · GYN ROBOTIC · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Dec 16, 2025Class IIOpen, Classified
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Z-1088-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Oct 10, 2025Class IOpen, Classified
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Z-0450-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Other Gynecological laparoscopic kit devices
- Medline Industries, Inc. DYNJ44864SMedline Industries, Inc. · Gynecological laparoscopic kit4 recalls0 reports
- Medline Industries, Inc. DYNJ909398AMedline Industries, Inc. · Gynecological laparoscopic kit4 recalls0 reports
- Medline Industries, Inc. DYNJ38844NMedline Industries, Inc. · Gynecological laparoscopic kit4 recalls0 reports
- Medline Industries, Inc. DYNJ66298DMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls1 report
- Medline Industries, Inc. DYNJ906926BMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls0 reports
- Medline Industries, Inc. DYNJ49191BMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls0 reports
- Medline Industries, Inc. DYNJ52564DMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls0 reports
- Medline Industries, Inc. CDS920099SMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page