Recalls naming this device

  1. Dec 16, 2025Class IIOpen, Classified

    Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

    Z-1086-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗