Skip to content

Medline Industries, Inc. DYNJ0525321ORecalls & reported problems

Medline Industries, Inc. · DAVINCI ROBOTIC PACK-LF · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Dec 16, 2025Class IIOpen, Classified

    Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

    Z-1087-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Other General surgery tray devices

All General surgery tray devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page