Portex 526060Recalls & reported problems
Icu Medical, Inc. · Class 2
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10351688041128), received between Jan 25, 2023 and Jun 21, 2023 — classified as 7 injuries, 1 malfunction.
The most frequently reported specific device problem is “Crack” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Crack7
- Break1
Patient problems
- Decreased Respiratory Rate7
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 21, 2023Break; Break
IT WAS REPORTED, THAT THE CANNULA BROKE INSIDE OF THE PATIENT. PARENTS USED TWEEZERS TO GET THE CANNULA OUT. NO ADVERSE EFFECTS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17171047 ↗InjuryJan 25, 2023Crack; Crack
IT WAS REPORTED THAT THE CANNULA BEGINS TO CRACK AFTER 1-3 DAYS OF INSERTION AND NEEDS TO BE REPLACED. THE ISSUE TOOK PLACE IN THE PATIENT'S HOME. THERE HAS BEEN NO REPORT OF PATIENT INJURY OR NO OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT, AFTER TRACH REPLACEMENT TOOK PLACE.
Patient problems: Decreased Respiratory Rate
MAUDE report 16244145 ↗InjuryJan 25, 2023Crack; Crack
IT WAS REPORTED THAT THE CANNULA BEGINS TO CRACK AFTER 1-3 DAYS OF INSERTION AND NEEDS TO BE REPLACED. THERE HAS BEEN NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.
Patient problems: Decreased Respiratory Rate
MAUDE report 16244130 ↗InjuryJan 25, 2023Crack; Crack
IT WAS REPORTED THAT THE CANNULA BEGINS TO CRACK AFTER 1-3 DAYS OF INSERTION AND NEEDS TO BE REPLACED. THE ISSUE TOOK PLACE IN THE PATIENT'S HOME. THERE HAS BEEN NO REPORT OF PATIENT INJURY OR NO OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT, AFTER TRACH REPLACEMENT TOOK PLACE.
Patient problems: Decreased Respiratory Rate
MAUDE report 16244128 ↗InjuryJan 25, 2023Crack; Crack
IT WAS REPORTED THAT THE CANNULA BEGINS TO CRACK AFTER 1-3 DAYS OF INSERTION AND NEEDS TO BE REPLACED. THE ISSUE TOOK PLACE IN THE PATIENT'S HOME. THERE HAS BEEN NO REPORT OF PATIENT INJURY OR NO OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT, AFTER TRACH REPLACEMENT TOOK PLACE.
Patient problems: Decreased Respiratory Rate
MAUDE report 16244125 ↗InjuryJan 25, 2023Crack; Crack
IT WAS REPORTED THAT THE CANNULA BEGINS TO CRACK AFTER 1-3 DAYS OF INSERTION AND NEEDS TO BE REPLACED. THE ISSUE TOOK PLACE IN THE PATIENT'S HOME. THERE HAS BEEN NO REPORT OF PATIENT INJURY OR NO OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT, AFTER TRACH REPLACEMENT TOOK PLACE.
Patient problems: Decreased Respiratory Rate
MAUDE report 16244088 ↗InjuryJan 25, 2023Crack; Crack
IT WAS REPORTED THAT THE CANNULA BEGINS TO CRACK AFTER 1-3 DAYS OF INSERTION AND NEEDS TO BE REPLACED. THE ISSUE TOOK PLACE IN THE PATIENT'S HOME. THERE HAS BEEN NO REPORT OF PATIENT INJURY OR NO OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT, AFTER TRACH REPLACEMENT TOOK PLACE.
Patient problems: Decreased Respiratory Rate
MAUDE report 16244034 ↗InjuryJan 25, 2023Crack; Crack
IT WAS REPORTED THAT THE CANNULA BEGINS TO CRACK AFTER 1-3 DAYS OF INSERTION AND NEEDS TO BE REPLACED. THE ISSUE TOOK PLACE IN THE PATIENT'S HOME. THERE HAS BEEN NO REPORT OF PATIENT INJURY OR NO OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT, AFTER TRACH REPLACEMENT TOOK PLACE.
Patient problems: Decreased Respiratory Rate
MAUDE report 16244006 ↗
Other Tube, tracheostomy (w/wo connector) devices
- Portex 101/860/080CZIcu Medical, Inc. · Tube, tracheostomy (w/wo connector)188 reports
- Portex 101/860/075CZIcu Medical, Inc. · Tube, tracheostomy (w/wo connector)115 reports
- Bivona 60SP040Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall20 reports
- Bivona 60SP050Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall13 reports
- Bivona 60P040Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall12 reports
- Portex 101/860/070CZIcu Medical, Inc. · Tube, tracheostomy (w/wo connector)60 reports
- Bivona 60SP045Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall8 reports
- Bivona 60SN040Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall5 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.