Jelco 3073Recalls & reported problems

Icu Medical, Inc. · GTIN for PROTECTIVE W SAFETY IV CATHETER - RADIOPAQUE 24G 1/EA · Class 2

Quick summary

This ICU Medical Jelco 3073 protective IV catheter is a short-term safety needle used for IV therapy. Reports mostly describe the catheter tip fraying, bending or splitting as it is inserted, sometimes so the needle poked through the tip and blood would not flush; some catheters were bent or torn straight from the packaging. There are 33 reports, with no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
33
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 10351688071293), received between Aug 7, 2018 and Sep 23, 2024 — classified as 33 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (28 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent28
  • Material Frayed19
  • Material Perforation18
  • Material Split, Cut or Torn7
  • Contamination /Decontamination Problem3
  • Mechanical Problem3
  • Obstruction of Flow3
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions28
  • No Consequences Or Impact To Patient3
  • Insufficient Information2

Recent adverse event reports

The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2024Complete Blockage; Complete Blockage

    IT WAS REPORTED THAT THERE WERE TWO ATTEMPTS AT IV THERAPY, BOTH OF WHICH WERE SUCCESSFUL IN RECOVERING BLOOD, BUT IT WAS NOT FLUSHING. THE END OF CATHETER WAS SPREAD AFTER REMOVAL. THEY PULLED 2 MORE UNUSED FROM SUPPLY ROOM AND BOTH OF THOSE WERE DEFECTIVE ALSO. THERE WAS PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20280894 ↗
  • MalfunctionApr 19, 2022Material Frayed; Material Perforation; Material Twisted/Bent; Material Frayed; Material Perforation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CATHETER TIPS WERE FRAYING UPON INSERTION. THE NEEDLE WAS POKING THROUGH CATHETER TIP. THE NEEDLE TIP WAS BENT PRIOR TO PACKAGES BEING OPENED AND THE CATHETER WAS FLIMSY ENOUGH TO BE BENT WITH AIR BLOWING ON IT, CAUSING MULTIPLE STICKS UPON INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14147713 ↗
  • MalfunctionApr 19, 2022Material Frayed; Material Perforation; Material Twisted/Bent; Material Frayed; Material Perforation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CATHETER TIPS WERE FRAYING UPON INSERTION. THE NEEDLE WAS POKING THROUGH CATHETER TIP. THE NEEDLE TIP WAS BENT PRIOR TO PACKAGES BEING OPENED AND THE CATHETER WAS FLIMSY ENOUGH TO BE BENT WITH AIR BLOWING ON IT, CAUSING MULTIPLE STICKS UPON INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14147709 ↗
  • MalfunctionApr 19, 2022Material Frayed; Material Perforation; Material Twisted/Bent; Material Frayed; Material Perforation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CATHETER TIPS WERE FRAYING UPON INSERTION. THE NEEDLE WAS POKING THROUGH CATHETER TIP. THE NEEDLE TIP WAS BENT PRIOR TO PACKAGES BEING OPENED AND THE CATHETER WAS FLIMSY ENOUGH TO BE BENT WITH AIR BLOWING ON IT, CAUSING MULTIPLE STICKS UPON INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14147627 ↗
  • MalfunctionApr 19, 2022Material Frayed; Material Perforation; Material Twisted/Bent; Material Frayed; Material Perforation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CATHETER TIPS WERE FRAYING UPON INSERTION. THE NEEDLE WAS POKING THROUGH CATHETER TIP. THE NEEDLE TIP WAS BENT PRIOR TO PACKAGES BEING OPENED AND THE CATHETER WAS FLIMSY ENOUGH TO BE BENT WITH AIR BLOWING ON IT, CAUSING MULTIPLE STICKS UPON INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14147357 ↗
  • MalfunctionApr 19, 2022Material Frayed; Material Perforation; Material Twisted/Bent; Material Frayed; Material Perforation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CATHETER TIPS WERE FRAYING UPON INSERTION. THE NEEDLE WAS POKING THROUGH CATHETER TIP. THE NEEDLE TIP WAS BENT PRIOR TO PACKAGES BEING OPENED AND THE CATHETER WAS FLIMSY ENOUGH TO BE BENT WITH AIR BLOWING ON IT, CAUSING MULTIPLE STICKS UPON INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14147177 ↗
  • MalfunctionApr 19, 2022Material Frayed; Material Perforation; Material Twisted/Bent; Material Frayed; Material Perforation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CATHETER TIPS WERE FRAYING UPON INSERTION. THE NEEDLE WAS POKING THROUGH CATHETER TIP. THE NEEDLE TIP WAS BENT PRIOR TO PACKAGES BEING OPENED AND THE CATHETER WAS FLIMSY ENOUGH TO BE BENT WITH AIR BLOWING ON IT, CAUSING MULTIPLE STICKS UPON INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14146800 ↗
  • MalfunctionApr 19, 2022Material Frayed; Material Perforation; Material Twisted/Bent; Material Frayed; Material Perforation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CATHETER TIPS WERE FRAYING UPON INSERTION. THE NEEDLE WAS POKING THROUGH CATHETER TIP. THE NEEDLE TIP WAS BENT PRIOR TO PACKAGES BEING OPENED AND THE CATHETER WAS FLIMSY ENOUGH TO BE BENT WITH AIR BLOWING ON IT, CAUSING MULTIPLE STICKS UPON INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14146591 ↗
  • MalfunctionApr 19, 2022Material Frayed; Material Perforation; Material Twisted/Bent; Material Frayed; Material Perforation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CATHETER TIPS WERE FRAYING UPON INSERTION. THE NEEDLE WAS POKING THROUGH CATHETER TIP. THE NEEDLE TIP WAS BENT PRIOR TO PACKAGES BEING OPENED AND THE CATHETER WAS FLIMSY ENOUGH TO BE BENT WITH AIR BLOWING ON IT, CAUSING MULTIPLE STICKS UPON INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14146293 ↗
  • MalfunctionApr 19, 2022Material Frayed; Material Perforation; Material Twisted/Bent; Material Frayed; Material Perforation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CATHETER TIPS WERE FRAYING UPON INSERTION. THE NEEDLE WAS POKING THROUGH CATHETER TIP. THE NEEDLE TIP WAS BENT PRIOR TO PACKAGES BEING OPENED AND THE CATHETER WAS FLIMSY ENOUGH TO BE BENT WITH AIR BLOWING ON IT, CAUSING MULTIPLE STICKS UPON INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14145957 ↗
  • MalfunctionApr 19, 2022Material Frayed; Material Perforation; Material Twisted/Bent; Material Frayed; Material Perforation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CATHETER TIPS WERE FRAYING UPON INSERTION. THE NEEDLE WAS POKING THROUGH CATHETER TIP. THE NEEDLE TIP WAS BENT PRIOR TO PACKAGES BEING OPENED AND THE CATHETER WAS FLIMSY ENOUGH TO BE BENT WITH AIR BLOWING ON IT, CAUSING MULTIPLE STICKS UPON INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14145630 ↗
  • MalfunctionApr 19, 2022Material Frayed; Material Perforation; Material Twisted/Bent; Material Frayed; Material Perforation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CATHETER TIPS WERE FRAYING UPON INSERTION. THE NEEDLE WAS POKING THROUGH CATHETER TIP. THE NEEDLE TIP WAS BENT PRIOR TO PACKAGES BEING OPENED AND THE CATHETER WAS FLIMSY ENOUGH TO BE BENT WITH AIR BLOWING ON IT, CAUSING MULTIPLE STICKS UPON INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14144655 ↗
  • MalfunctionApr 19, 2022Material Frayed; Material Perforation; Material Twisted/Bent; Material Frayed; Material Perforation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CATHETER TIPS WERE FRAYING UPON INSERTION. THE NEEDLE WAS POKING THROUGH CATHETER TIP. THE NEEDLE TIP WAS BENT PRIOR TO PACKAGES BEING OPENED AND THE CATHETER WAS FLIMSY ENOUGH TO BE BENT WITH AIR BLOWING ON IT, CAUSING MULTIPLE STICKS UPON INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14144015 ↗
  • MalfunctionApr 19, 2022Material Frayed; Material Perforation; Material Twisted/Bent; Material Frayed; Material Perforation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CATHETER TIPS WERE FRAYING UPON INSERTION. THE NEEDLE WAS POKING THROUGH CATHETER TIP. THE NEEDLE TIP WAS BENT PRIOR TO PACKAGES BEING OPENED AND THE CATHETER WAS FLIMSY ENOUGH TO BE BENT WITH AIR BLOWING ON IT, CAUSING MULTIPLE STICKS UPON INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14143655 ↗
  • MalfunctionApr 19, 2022Material Frayed; Material Perforation; Material Twisted/Bent; Material Frayed; Material Perforation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CATHETER TIPS WERE FRAYING UPON INSERTION. THE NEEDLE WAS POKING THROUGH CATHETER TIP. THE NEEDLE TIP WAS BENT PRIOR TO PACKAGES BEING OPENED AND THE CATHETER WAS FLIMSY ENOUGH TO BE BENT WITH AIR BLOWING ON IT, CAUSING MULTIPLE STICKS UPON INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14143242 ↗
  • MalfunctionApr 19, 2022Material Frayed; Material Perforation; Material Twisted/Bent; Material Frayed; Material Perforation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CATHETER TIPS WERE FRAYING UPON INSERTION. THE NEEDLE WAS POKING THROUGH CATHETER TIP. THE NEEDLE TIP WAS BENT PRIOR TO PACKAGES BEING OPENED AND THE CATHETER WAS FLIMSY ENOUGH TO BE BENT WITH AIR BLOWING ON IT, CAUSING MULTIPLE STICKS UPON INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14143016 ↗
  • MalfunctionApr 19, 2022Material Frayed; Material Perforation; Material Twisted/Bent; Material Frayed; Material Perforation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CATHETER TIPS WERE FRAYING UPON INSERTION. THE NEEDLE WAS POKING THROUGH CATHETER TIP. THE NEEDLE TIP WAS BENT PRIOR TO PACKAGES BEING OPENED AND THE CATHETER WAS FLIMSY ENOUGH TO BE BENT WITH AIR BLOWING ON IT, CAUSING MULTIPLE STICKS UPON INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14142936 ↗
  • MalfunctionApr 19, 2022Material Frayed; Material Perforation; Material Twisted/Bent; Material Frayed; Material Perforation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CATHETER TIPS WERE FRAYING UPON INSERTION. THE NEEDLE WAS POKING THROUGH CATHETER TIP. THE NEEDLE TIP WAS BENT PRIOR TO PACKAGES BEING OPENED AND THE CATHETER WAS FLIMSY ENOUGH TO BE BENT WITH AIR BLOWING ON IT, CAUSING MULTIPLE STICKS UPON INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14142757 ↗
  • MalfunctionApr 19, 2022Material Frayed; Material Perforation; Material Twisted/Bent; Material Frayed; Material Perforation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CATHETER TIPS WERE FRAYING UPON INSERTION. THE NEEDLE WAS POKING THROUGH CATHETER TIP. THE NEEDLE TIP WAS BENT PRIOR TO PACKAGES BEING OPENED AND THE CATHETER WAS FLIMSY ENOUGH TO BE BENT WITH AIR BLOWING ON IT, CAUSING MULTIPLE STICKS UPON INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14142644 ↗
  • MalfunctionApr 15, 2022Material Puncture/Hole

    HOLE NOTED IN IV CATHETER TIP DURING IV INSERTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14117725 ↗
  • MalfunctionApr 12, 2022Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE SHEATH WAS BENT, THE CATHETER CREASED AND WOULD NOT ALLOW THE NEEDLE TO RETRACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14089080 ↗
  • MalfunctionMar 31, 2022Material Twisted/Bent; Material Split, Cut or Torn; Material Twisted/Bent; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE WHITE CATHETER PIECE HAS EITHER BEEN BENT OR TORN BEFORE USE OR HAS FAILED UPON USE. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13969074 ↗
  • MalfunctionMar 31, 2022Material Twisted/Bent; Material Split, Cut or Torn; Material Twisted/Bent; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE WHITE CATHETER PIECE HAS EITHER BEEN BENT OR TORN BEFORE USE OR HAS FAILED UPON USE. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13968058 ↗
  • MalfunctionMar 31, 2022Material Twisted/Bent; Material Split, Cut or Torn; Material Twisted/Bent; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE WHITE CATHETER PIECE HAS EITHER BEEN BENT OR TORN BEFORE USE OR HAS FAILED UPON USE. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13967801 ↗
  • MalfunctionMar 31, 2022Material Twisted/Bent; Material Split, Cut or Torn; Material Twisted/Bent; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE WHITE CATHETER PIECE HAS EITHER BEEN BENT OR TORN BEFORE USE OR HAS FAILED UPON USE. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13967670 ↗

Other Catheter,intravascular,therapeutic,short-term less than 30 days devices

All Catheter,intravascular,therapeutic,short-term less than 30 days devices →