Skip to content

Jelco 3080Recalls & reported problems

Icu Medical, Inc. · GTIN for PROTECTIV PLUS-W SAFETY CATHETER-RADIOPAQUE 22G 1/EA · Class 2

Quick summary

This ICU Medical Jelco 3080 is a short-term intravenous safety catheter with a needle guard, used for IV access. The 38 reports mostly describe parts of the catheter coming loose or detaching, sometimes with burrs, along with blood leaking from the hub and fluid leaking at the insertion site. Reports mention 2 injuries and no deaths, with 36 malfunctions. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
38
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 38 adverse event reports that list this exact device (DI 10351688071330), received between Aug 28, 2017 and Mar 30, 2026 — classified as 2 injuries, 36 malfunctions.

The most frequently reported specific device problem is “Dull, Blunt” (20 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 3 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Dull, Blunt20
  • Fluid/Blood Leak7
  • Detachment of Device or Device Component3
  • Device Operates Differently Than Expected2
  • Leak/Splash2
  • Material Fragmentation2
  • Positioning Problem2
  • Activation, Positioning or Separation Problem1

Patient problems

  • No Consequences Or Impact To Patient17
  • No Clinical Signs, Symptoms or Conditions13
  • Blood Loss3
  • Erythema2
  • Necrosis2
  • No Known Impact Or Consequence To Patient2
  • Swelling2
  • Unspecified Infection2

Recent adverse event reports

The 25 most recent of 38 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 30, 2026Fluid/Blood Leak; Leak/Splash

    IT WAS REPORTED THAT THE CATHETER DETACHED SPONTANEOUSLY AT THE CATHLON LEVEL PRIOR TO THE REMOVAL PROCEDURE. THIS RESULTS IN BLOOD LEAKAGE AND ENVIRONMENTAL CONTAMINATION, REQUIRING BOTH PATIENT REASSURANCE AND IMMEDIATE CLEANING. MULTIPLE BATCHES APPEAR TO BE AFFECTED. THESE INCIDENTS HAVE BEEN REPORTED BY THE IMAGING DEPARTMENTS, MRI, X-RAY, CT SCAN, AND NUCLEAR MEDICINE AND INVOLVED FOUR ITEMS, IN TOTAL, ONE FOR EACH ROOM AND ON EACH DIFFERENT PATIENT CONCERNED. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24734483 ↗
  • MalfunctionMar 30, 2026Fluid/Blood Leak; Leak/Splash

    IT WAS REPORTED THAT THE CATHETER DETACHED SPONTANEOUSLY AT THE CATHLON LEVEL PRIOR TO THE REMOVAL PROCEDURE. THIS RESULTS IN BLOOD LEAKAGE AND ENVIRONMENTAL CONTAMINATION, REQUIRING BOTH PATIENT REASSURANCE AND IMMEDIATE CLEANING. MULTIPLE BATCHES APPEAR TO BE AFFECTED. THESE INCIDENTS HAVE BEEN REPORTED BY THE IMAGING DEPARTMENTS, MRI, X-RAY, CT SCAN, AND NUCLEAR MEDICINE AND INVOLVED FOUR ITEMS, IN TOTAL, ONE FOR EACH ROOM AND ON EACH DIFFERENT PATIENT CONCERNED. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24734343 ↗
  • MalfunctionMar 19, 2026Environmental Particulates

    IT WAS FURTHER REPORTED THAT THE CATHETER HAD BURRS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24641070 ↗
  • MalfunctionNov 17, 2025Fluid/Blood Leak; Sharp Edges; Fluid/Blood Leak; Sharp Edges

    IT WAS REPORTED THAT THE PRODUCT DETACHED FROM THE SHARP HUB WHEN THE CATHETER WAS ADVANCED AND THE SHARP WAS RETRACTED. WHEN THIS HAPPENED, THE CATHETER FREE FLOWED WITH BLOOD RETURN AND CREATED A SAFETY RISK WITH LOTS OF BLOOD ON THE PATIENT/STRETCHER, ETC. THE FACILITY HAD MULTIPLE STAFF MEMBERS WHO ENCOUNTERED THE SAME SITUATION. SAMPLE IS NOT AVAILABLE FOR RETURN. THERE WERE PATIENT INVOLVEMENT AND NO HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23574138 ↗
  • MalfunctionJul 18, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AS SOON AS THE IV CATHETER WAS INSERTED THE BOTTOM ATTACHMENT TO IV CATHETER FELL OUT BY ITSELF, BLOOD LEAKED EVERYWHERE FROM ANTECUBITAL AREA TODAY, PATIENT CLOTHING AND BEDDING HAD TO BE CHANGED. PER REPORTER ANOTHER STAFF COMPLAINED ABOUT THIS PROBLEM TOO. THE REPORTER NOTED THAT THESE CATHETERS HAVE WINGS, VERSUS THE OLD IV NUMBER 22 THAT HAD NO WINGS AND HAD NO PROBLEMS. ON ANOTHER OCCASION, A RELATIVE OBSERVED THIS LEAKAGE OF BLOOD WHEN THE BOTTOM FELL OUT AS REPORTED EARLIER. THERE WAS A RISK TO HEALTH CARE STAFF AS THERE WAS AN EXPOSURE TO BLOOD. THERE WAS PATIENT INVOLVEMENT, BUT NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22545940 ↗
  • MalfunctionJul 18, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AS SOON AS THE IV CATHETER WAS INSERTED THE BOTTOM ATTACHMENT TO IV CATHETER FELL OUT BY ITSELF, BLOOD LEAKED EVERYWHERE FROM ANTECUBITAL AREA TODAY, PATIENT CLOTHING AND BEDDING HAD TO BE CHANGED. PER REPORTER ANOTHER STAFF COMPLAINED ABOUT THIS PROBLEM TOO. THE REPORTER NOTED THAT THESE CATHETERS HAVE WINGS, VERSUS THE OLD IV NUMBER 22 THAT HAD NO WINGS AND HAD NO PROBLEMS. ON ANOTHER OCCASION, A RELATIVE OBSERVED THIS LEAKAGE OF BLOOD WHEN THE BOTTOM FELL OUT AS REPORTED EARLIER. THERE WAS A RISK TO HEALTH CARE STAFF AS THERE WAS AN EXPOSURE TO BLOOD. THERE WAS PATIENT INVOLVEMENT, BUT NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22545939 ↗
  • MalfunctionApr 22, 2025Disconnection; Disconnection

    IT WAS REPORTED THAT THE CATHETER HUB WAS VERY LOOSE AND DETACHED EASILY. THE IV CATHETER WAS FALLING OFF THE NEEDLE JUST UPON REMOVING THE CAP. THE NEEDLE GUARD DID NOT SLIDE BACK WHEN PULLING BACK ON IT, RATHER IT DETACHED FROM THE CATHETER CAUSING THE BLOOD TO FLOW FREELY AND LEFT THE NEEDLE EXPOSED. SOMETIMES THIS HAPPENED SPONTANEOUSLY WITHOUT EVEN PULLING BACK ON THE NEEDLE GUARD. THIS HAD RESULTED IN ONE NEEDLE STICK INJURY. IV STARTED WITH A LOT OF DIFFICULTY AND WAS A SAFETY CONCERN. IT CAUSED AN INCREASED EXPOSURE TO BLOOD AND UNPROTECTED NEEDLES. THERE WAS NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: Needle Stick/Puncture

    MAUDE report 21885698 ↗
  • MalfunctionJan 13, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT WHEN THE NEEDLE ENTERED THE VEIN, BLOOD CAME OUT AT HUB PRIOR TO THREADING OF CATHETER. THE CLINICIAN REPORTED HEARING A "POP SOUND". THERE WAS PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21139121 ↗
  • MalfunctionSep 6, 2024Positioning Problem; Positioning Problem

    IT WAS REPORTED A DIFFICULTY IN PRICKING THE PATIENT AND GETTING INTO THE VEIN. THE CATHETER HANGS IN THE PATIENT'S VEIN AND IN SOME CASES, THEY FAILED TO INTRODUCE THE CATHETER TO THE PATIENT EVEN AFTER 4 ATTEMPTS. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20157193 ↗
  • MalfunctionSep 6, 2024Positioning Problem; Positioning Problem

    IT WAS REPORTED A DIFFICULTY IN PRICKING THE PATIENT AND GETTING INTO THE VEIN. THE CATHETER HANGS IN THE PATIENT'S VEIN AND IN SOME CASES, THEY FAILED TO INTRODUCE THE CATHETER TO THE PATIENT EVEN AFTER 4 ATTEMPTS. THERE WAS PATIENT INVOLVEMENT AND UNKNOWN PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20156907 ↗
  • MalfunctionSep 3, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THERE WAS A CATHETER LEAK WHEN THE NURSE STABBED PATIENT. THERE WAS PATIENT INVOLVEMENT, AND NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20126670 ↗
  • MalfunctionDec 3, 2021Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE IV CATHETER SHEARED OFF OR MISSING. ATTEMPTED TO FLUSH IV SALINE LOCK WITH NS 10ML. UPON FLUSHING IV WAS NOT PATENT AND LEAKING SO REMOVED IV AND NOTICED THERE WAS NO LONG CLEAR CATHETER PIECE, IT HAD BROKEN OFF COMPLETELY AT THE HUB OF THE CATHETER. UNKNOWN HOW LONG IT HAD BEEN LIKE THAT. THE IV WAS STARTED NEW YESTERDAY BY ANOTHER COMMUNITY PARAMEDIC WITH NO COMPLICATIONS. THE PT. HAD BEEN KNOW TO PICK AND PULL AT HER IVS OVER THE REFERRAL. IT WAS A 22G IV.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12929046 ↗
  • MalfunctionDec 2, 2021Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE IV CATHETER SHEARED OFF OR MISSING. ATTEMPTED TO FLUSH IV SALINE LOCK WITH NS 10ML. UPON FLUSHING IV WAS NOT PATENT AND LEAKING SO REMOVED IV AND NOTICED THERE WAS NO LONG CLEAR CATHETER PIECE, IT HAD BROKEN OFF COMPLETELY AT THE HUB OF THE CATHETER. UNKNOWN HOW LONG IT HAD BEEN LIKE THAT. THE IV WAS STARTED NEW YESTERDAY BY ANOTHER COMMUNITY PARAMEDIC WITH NO COMPLICATIONS. THE PT. HAD BEEN KNOW TO PICK AND PULL AT HER IVS OVER THE REFERRAL. IT WAS A 22G IV.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12920565 ↗
  • MalfunctionJun 23, 2021Dull, Blunt; Dull, Blunt

    DEVICE EVALUATION COMPETED AND SUMMARY IN H 10.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12049961 ↗
  • MalfunctionJul 24, 2018Dull, Blunt; Dull, Blunt

    INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL CATHETER IS DULL. NO FURTHER INFORMATION PROVIDED. NO PATIENT INJURY RESULTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7716005 ↗
  • MalfunctionJul 24, 2018Dull, Blunt; Dull, Blunt

    INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL CATHETER IS DULL. NO FURTHER INFORMATION PROVIDED. NO PATIENT INJURY RESULTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7715989 ↗
  • MalfunctionJul 2, 2018Dull, Blunt; Dull, Blunt

    INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL PERIPHERAL INTRAVENOUS CATHETER WAS DULL ON INSERTION. NO ADVERSE EFFECTS WERE EXPERIENCED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7656502 ↗
  • MalfunctionJun 29, 2018Dull, Blunt; Dull, Blunt

    INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL PERIPHERAL INTRAVENOUS CATHETER WAS DULL ON INSERTION. NO ADVERSE EFFECTS WERE EXPERIENCED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7652299 ↗
  • MalfunctionJun 29, 2018Dull, Blunt; Dull, Blunt

    INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL PERIPHERAL INTRAVENOUS CATHETER WAS DULL ON INSERTION. NO ADVERSE EFFECTS WERE EXPERIENCED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7652297 ↗
  • MalfunctionJun 29, 2018Dull, Blunt; Dull, Blunt

    INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL PERIPHERAL INTRAVENOUS CATHETER WAS DULL ON INSERTION. NO ADVERSE EFFECTS WERE EXPERIENCED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7652248 ↗
  • MalfunctionJun 29, 2018Dull, Blunt; Dull, Blunt

    INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL PERIPHERAL INTRAVENOUS CATHETER WAS DULL ON INSERTION. NO ADVERSE EFFECTS WERE EXPERIENCED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7652149 ↗
  • MalfunctionJun 29, 2018Dull, Blunt; Dull, Blunt

    INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL PERIPHERAL INTRAVENOUS CATHETER WAS DULL ON INSERTION. NO ADVERSE EFFECTS WERE EXPERIENCED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7652115 ↗
  • MalfunctionJun 29, 2018Dull, Blunt; Dull, Blunt

    INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL PERIPHERAL INTRAVENOUS CATHETER WAS DULL ON INSERTION. NO ADVERSE EFFECTS WERE EXPERIENCED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7652100 ↗
  • MalfunctionJun 29, 2018Dull, Blunt; Dull, Blunt

    INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL PERIPHERAL INTRAVENOUS CATHETER WAS DULL ON INSERTION. NO ADVERSE EFFECTS WERE EXPERIENCED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7652051 ↗
  • MalfunctionJun 29, 2018Dull, Blunt; Dull, Blunt

    INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL PERIPHERAL INTRAVENOUS CATHETER WAS DULL ON INSERTION. NO ADVERSE EFFECTS WERE EXPERIENCED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7652018 ↗

Other Catheter,intravascular,therapeutic,short-term less than 30 days devices

All Catheter,intravascular,therapeutic,short-term less than 30 days devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.