Jelco 335603Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of Jelco (101)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10351688071446), received between Jul 18, 2018 and Feb 18, 2021 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Difficult or Delayed Separation” (10 reports).
1 other recall cover devices cleared under the same FDA submission (K012128); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Difficult or Delayed Separation10
- Difficult to Remove3
Patient problems
- No Known Impact Or Consequence To Patient10
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 18, 2021Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT WHILE THE SMITHS MEDICAL CATHETER WAS INTO PATIENT'S VEIN, THE USER COULD NOT EASILY REMOVE THE MANDREL FROM THE CATHETER WHICH LED TO PAIN AND WOUND FOR PATIENT. NO FURTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11341524 ↗MalfunctionFeb 18, 2021Difficult to Remove; Difficult to Remove
WHILE CATHETER WAS INTO PATIENT'S VEIN, THE USER COULD NOT EASILY REMOVE THE MANDREL FROM THE CATHETER WHICH LED TO PAIN AND WOUND FOR PATIENT. ISSUE OCCURRED *FREQUENTLY DURING THE LAST FEW MONTHS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11341521 ↗MalfunctionFeb 18, 2021Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT WHILE THE SMITHS MEDICAL CATHETER WAS INTO PATIENT'S VEIN, THE USER COULD NOT EASILY REMOVE THE MANDREL FROM THE CATHETER WHICH LED TO PAIN AND WOUND FOR PATIENT. NO FURTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11341518 ↗MalfunctionJul 18, 2018Difficult or Delayed Separation; Difficult or Delayed Separation
INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL JELCO® ACUVANCE® SAFETY I.V. CATHETER WAS NOT SEPARATING EASILY AT THE HUB UPON THREADING THE PLASTIC CATHETER INTO THE PATIENT'S VEIN. REPORTED USE OF EXTRA FORCE WAS NEEDED TO SEPARATE WITH A CONCERN OF THIS POSSIBLY COLLAPSING THE VEIN FROM THE PRESSURE. THERE WERE NO REPORTED ADVERSE EFFECTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7699958 ↗MalfunctionJul 18, 2018Difficult or Delayed Separation; Difficult or Delayed Separation
INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL JELCO® ACUVANCE® SAFETY I.V. CATHETER WAS NOT SEPARATING EASILY AT THE HUB UPON THREADING THE PLASTIC CATHETER INTO THE PATIENT'S VEIN. REPORTED USE OF EXTRA FORCE WAS NEEDED TO SEPARATE WITH A CONCERN OF THIS POSSIBLY COLLAPSING THE VEIN FROM THE PRESSURE. THERE WERE NO REPORTED ADVERSE EFFECTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7699930 ↗MalfunctionJul 18, 2018Difficult or Delayed Separation; Difficult or Delayed Separation
INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL JELCO® ACUVANCE® SAFETY I.V. CATHETER WAS NOT SEPARATING EASILY AT THE HUB UPON THREADING THE PLASTIC CATHETER INTO THE PATIENT'S VEIN. REPORTED USE OF EXTRA FORCE WAS NEEDED TO SEPARATE WITH A CONCERN OF THIS POSSIBLY COLLAPSING THE VEIN FROM THE PRESSURE. THERE WERE NO REPORTED ADVERSE EFFECTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7699892 ↗MalfunctionJul 18, 2018Difficult or Delayed Separation; Difficult or Delayed Separation
INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL JELCO® ACUVANCE® SAFETY I.V. CATHETER WAS NOT SEPARATING EASILY AT THE HUB UPON THREADING THE PLASTIC CATHETER INTO THE PATIENT'S VEIN. REPORTED USE OF EXTRA FORCE WAS NEEDED TO SEPARATE WITH A CONCERN OF THIS POSSIBLY COLLAPSING THE VEIN FROM THE PRESSURE. THERE WERE NO REPORTED ADVERSE EFFECTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7699867 ↗MalfunctionJul 18, 2018Difficult or Delayed Separation; Difficult or Delayed Separation
INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL JELCO® ACUVANCE® SAFETY I.V. CATHETER WAS NOT SEPARATING EASILY AT THE HUB UPON THREADING THE PLASTIC CATHETER INTO THE PATIENT'S VEIN. REPORTED USE OF EXTRA FORCE WAS NEEDED TO SEPARATE WITH A CONCERN OF THIS POSSIBLY COLLAPSING THE VEIN FROM THE PRESSURE. THERE WERE NO REPORTED ADVERSE EFFECTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7699824 ↗MalfunctionJul 18, 2018Difficult or Delayed Separation; Difficult or Delayed Separation
INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL JELCO® ACUVANCE® SAFETY I.V. CATHETER WAS NOT SEPARATING EASILY AT THE HUB UPON THREADING THE PLASTIC CATHETER INTO THE PATIENT'S VEIN. REPORTED USE OF EXTRA FORCE WAS NEEDED TO SEPARATE WITH A CONCERN OF THIS POSSIBLY COLLAPSING THE VEIN FROM THE PRESSURE. THERE WERE NO REPORTED ADVERSE EFFECTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7699761 ↗MalfunctionJul 18, 2018Difficult or Delayed Separation; Difficult or Delayed Separation
INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL JELCO® ACUVANCE® SAFETY I.V. CATHETER WAS NOT SEPARATING EASILY AT THE HUB UPON THREADING THE PLASTIC CATHETER INTO THE PATIENT'S VEIN. REPORTED USE OF EXTRA FORCE WAS NEEDED TO SEPARATE WITH A CONCERN OF THIS POSSIBLY COLLAPSING THE VEIN FROM THE PRESSURE. THERE WERE NO REPORTED ADVERSE EFFECTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7699665 ↗MalfunctionJul 18, 2018Difficult or Delayed Separation; Difficult or Delayed Separation
INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL JELCO® ACUVANCE® SAFETY I.V. CATHETER WAS NOT SEPARATING EASILY AT THE HUB UPON THREADING THE PLASTIC CATHETER INTO THE PATIENT'S VEIN. REPORTED USE OF EXTRA FORCE WAS NEEDED TO SEPARATE WITH A CONCERN OF THIS POSSIBLY COLLAPSING THE VEIN FROM THE PRESSURE. THERE WERE NO REPORTED ADVERSE EFFECTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7699596 ↗MalfunctionJul 18, 2018Difficult or Delayed Separation; Difficult or Delayed Separation
INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL JELCO® ACUVANCE® SAFETY I.V. CATHETER WAS NOT SEPARATING EASILY AT THE HUB UPON THREADING THE PLASTIC CATHETER INTO THE PATIENT'S VEIN. REPORTED USE OF EXTRA FORCE WAS NEEDED TO SEPARATE WITH A CONCERN OF THIS POSSIBLY COLLAPSING THE VEIN FROM THE PRESSURE. THERE WERE NO REPORTED ADVERSE EFFECTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7699561 ↗MalfunctionJul 18, 2018Difficult or Delayed Separation; Difficult or Delayed Separation
INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL JELCO® ACUVANCE® SAFETY I.V. CATHETER WAS NOT SEPARATING EASILY AT THE HUB UPON THREADING THE PLASTIC CATHETER INTO THE PATIENT'S VEIN. REPORTED USE OF EXTRA FORCE WAS NEEDED TO SEPARATE WITH A CONCERN OF THIS POSSIBLY COLLAPSING THE VEIN FROM THE PRESSURE. THERE WERE NO REPORTED ADVERSE EFFECTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7699450 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K012128. One submission often covers many models, so check the product and lot details in each record.
- Z-0789-2017 · Nov 21, 2016 · TerminatedAcuvance2 Safety Catheter, Product No. 1720-AI Product Usage: The Acuvance 2 Safety Catheter is a Peripheral Intravenous Catheter (PIVC). PIVCs are designed for single use venipun
Other Catheter,intravascular,therapeutic,short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page