Skip to content

Jelco 328500Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of Jelco (101)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10351688081933), received between Jan 31, 2018 and Nov 10, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K160235); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Break1
  • Defective Device1
  • Delivery System Failure1
  • Fitting Problem1
  • Material Separation1
  • Material Twisted/Bent1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient1
  • Skin Tears1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 10, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE DEVICE LEAKED BLOOD IMMEDIATELY UPON THE REMOVAL OF THE SAFETY MECHANISM INSTEAD OF VALVING OFF THE BLOOD AS INTENDED. THEREFORE, BLOOD LEAKED ALL OVER THE PATIENT'S ARM AND BEDDING. THE ISSUE OCCURRED MULTIPLE TIMES. THERE WERE NO SERIOUS HARM AND NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23515950 ↗
  • MalfunctionOct 23, 2025Fluid/Blood Leak; Material Separation; Obstruction of Flow; Fluid/Blood Leak; Material Separation; Obstruction of Flow

    IT WAS REPORTED THAT, WHEN THE NEEDLE WAS REMOVED POST CATHETER ADVANCEMENT, A METAL PIECE WAS NOTED AT THE END OF IV NEEDLE AND THERE APPEARED TO BE THE BLOOD CONTROL VALVE THAT SHOULD BE INSIDE HUB OF THE CATHLON. PER REPORTER, THE IV CATH WAS FLUSHED SUCCESSFULLY. REPORTER ALLEGED THAT, WHEN NOT FLUSHING, BLOOD APPEARED TO BE LEAKING OUT OF IV CATH/HUB. REPORTER STATED THE IV WAS REMOVED AND IT WAS FOUND THE LONG CATHLON PORTION WAS NOT ATTACHED TO END OF HUB. PHYSICAL OCCLUSION OCCURRED WITH IV REMOVAL AND IN THE PROCESS THE LONG CATHLON PORTION OF IV REMAINED IN THE VEIN. REPORTER STATED THEY WERE ABLE TO PULL OUT THE CATHLON PORTION, BUT WERE UNABLE TO CONFIRM IF ALL PLASTIC WAS REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23366711 ↗
  • MalfunctionAug 16, 2023Fitting Problem; Material Twisted/Bent; Fitting Problem; Material Twisted/Bent

    IT WAS REPORTED THAT THE IV CATHETER TWISTED, DUE TO AN ILL FITTING HUB TO THE TUBING. THIS CAUSED THE IV TO DISLODGE AND THE SKIN SITE TO TEAR. ADVERSE PATIENT EFFECTS ARE UNKNOWN.

    Patient problems: Skin Tears

    MAUDE report 17560351 ↗
  • MalfunctionSep 6, 2022Defective Device; Defective Device

    IT WAS REPORTED THAT THE CATHETER WAS RIDGED AND DIFFICULT TO CHANNEL. THE DEVICE WAS UNABLE TO INSERT EASILY AND THE PATIENT SKIN CONTACT CAUSING PAIN AND BURNING IN THE VEIN: NO MEDICAL OR SURGICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15360385 ↗
  • MalfunctionFeb 4, 2022Break; Break

    IT WAS REPORTED THAT AFTER STARTING THE IV THE REPORTER WAS UNABLE TO HOOKUP THE IV LINE DUE TO PART OF THE PLASTIC CATHETER BREAKING OFF INSIDE THE GREEN CANNULA THAT WAS INSERTED INTO THE PATIENTS VEIN. THE ENTIRE GREEN CANNULA WAS PROMPTLY REMOVED FROM THE PATIENTS ARM AND THEN THE PATIENT NEEDED ANOTHER IV START. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13455873 ↗
  • MalfunctionJan 31, 2018Delivery System Failure; Delivery System Failure

    AN EMERGENCY DEPARTMENT NURSE REPORTED THAT A JELCO® VIAVALVE® SAFETY IV CATHETER WAS DIFFICULT TO THREAD FORWARD, REQUIRING TWO HANDS TO GET ENOUGH PRESSURE FOR FORWARD MOTION. THIS OFTEN RESULTS IN "BLOWING THE VEIN" AND THE "PATIENT HAS TO BE STUCK AGAIN". THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7232441 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K160235. One submission often covers many models, so check the product and lot details in each record.

  • Z-1218-2019 · Aug 22, 2018 · Terminated
    Jelco(R) ViaValve(TM) safety intravenous (I.V.) catheters, 18G x 1 1/4", Model Number REF 3265, MOD 10 The Jelco(R) ViaValve(TM) safety I.V. catheter is a sterile, non pyrogenic,

Other Catheter,intravascular,therapeutic,short-term less than 30 days devices

All Catheter,intravascular,therapeutic,short-term less than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page