Skip to content

Jelco 328300Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of Jelco (101)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10351688081971), received between May 14, 2022 and May 21, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).

1 other recall cover devices cleared under the same FDA submission (K160235); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak3
  • Audible Prompt/Feedback Problem1
  • Burst Container or Vessel1
  • Difficult to Insert1
  • Material Puncture/Hole1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Pain1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 21, 2026Fluid/Blood Leak

    IT WAS REPORTED BY NURSING STAFF THAT THE PATIENTS IV INITIATED WITH 24G VIAVALVE CATHETER AS PER SITE POLICY. PACKAGING INTACT. AFTER INSERTION, THE PATIENT NOTICED BLOOD WAS LEAKING FROM IV HUB WHERE IT MEETS THE INSERTION SITE. FLUSHED THE IV WITH NS SYRINGE AND THE SALINE LEAKED AROUND INSERTION SITE. IV CATHETER REMOVED AND INSPECTED. NOTED THAT THERE WAS A SMALL LEAK WHERE CATHETER MEETS THE HUB. THERE WAS PATIENT INVOLVEMENT WITH NO HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25241329 ↗
  • MalfunctionMay 13, 2026Obstruction of Flow

    AN EVENT INVOLVING A JELCO VIAVALVE IV CATHETER REPORTED THAT DURING THE SECOND INTRAVENOUS (IV) INSERTION ATTEMPT BY AN ADVANCED IV STARTER, THE IV DEVICE MALFUNCTIONED AFTER BLOOD RETURN WAS OBTAINED. THE NEEDLE AND A SMALL PLASTIC COMPONENT BECAME LODGED WITHIN THE IV HUB, PREVENTING USE OF THE IV FOR BLOODWORK OR FLUID ADMINISTRATION AND REQUIRING REMOVAL OF THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25161657 ↗
  • MalfunctionJun 4, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT IV CANNULA IS SUPPOSED TO BE BLOOD LESS. THE FILTER IS NOT WORKING AS IV CANNULAS HAVE BEEN LEAKING. THERE WAS PATIENT INVOLVEMENT, AND NO HARM/ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19463841 ↗
  • MalfunctionJun 4, 2024Burst Container or Vessel; Fluid/Blood Leak; Material Puncture/Hole; Burst Container or Vessel; Fluid/Blood Leak; Material Puncture/Hole

    IT WAS REPORTED THAT IV CATHETER 24G WAS INSERTED INTO PATIENT AND CAUSED SIGNIFICANT PAIN. PATIENT STARTED BLEEDING FROM THE IV INSERTION SITE (POKE) WHILE IV CATHETER WAS IN. WHEN FLUSHED WITH NORMAL SALINE, IV BLEW. WHEN RN TOOK THE IV OUT AND LOOKED CLOSELY - THE IV HAD "RIDGES" LIKE EXTRA PLASTIC ON CATHETER AND A HOLE. PATIENT GOT A NEW IV INSERTED AND EVERYTHING WAS FINE. THERE WAS PATIENT INVOLVEMENT AND MINIMAL PATIENT HARM.

    Patient problems: Pain

    MAUDE report 19461211 ↗
  • MalfunctionSep 6, 2022Difficult to Insert; Difficult to Insert

    IT WAS REPORTED WHEN STARTING IV'S IT WAS HARD TO BREAK THROUGH THE PATIENTS SKIN WITH THE NEEDLE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15362048 ↗
  • MalfunctionMay 14, 2022Audible Prompt/Feedback Problem; Audible Prompt/Feedback Problem

    IT WAS REPORTED THAT THE SAFETY DID NOT ENGAGE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14403076 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K160235. One submission often covers many models, so check the product and lot details in each record.

  • Z-1218-2019 · Aug 22, 2018 · Terminated
    Jelco(R) ViaValve(TM) safety intravenous (I.V.) catheters, 18G x 1 1/4", Model Number REF 3265, MOD 10 The Jelco(R) ViaValve(TM) safety I.V. catheter is a sterile, non pyrogenic,

Other Catheter,intravascular,therapeutic,short-term less than 30 days devices

All Catheter,intravascular,therapeutic,short-term less than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page