Jelco 328300Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of Jelco (101)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10351688081971), received between May 14, 2022 and May 21, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K160235); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak3
- Audible Prompt/Feedback Problem1
- Burst Container or Vessel1
- Difficult to Insert1
- Material Puncture/Hole1
- Obstruction of Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Pain1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 21, 2026Fluid/Blood Leak
IT WAS REPORTED BY NURSING STAFF THAT THE PATIENTS IV INITIATED WITH 24G VIAVALVE CATHETER AS PER SITE POLICY. PACKAGING INTACT. AFTER INSERTION, THE PATIENT NOTICED BLOOD WAS LEAKING FROM IV HUB WHERE IT MEETS THE INSERTION SITE. FLUSHED THE IV WITH NS SYRINGE AND THE SALINE LEAKED AROUND INSERTION SITE. IV CATHETER REMOVED AND INSPECTED. NOTED THAT THERE WAS A SMALL LEAK WHERE CATHETER MEETS THE HUB. THERE WAS PATIENT INVOLVEMENT WITH NO HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25241329 ↗MalfunctionMay 13, 2026Obstruction of Flow
AN EVENT INVOLVING A JELCO VIAVALVE IV CATHETER REPORTED THAT DURING THE SECOND INTRAVENOUS (IV) INSERTION ATTEMPT BY AN ADVANCED IV STARTER, THE IV DEVICE MALFUNCTIONED AFTER BLOOD RETURN WAS OBTAINED. THE NEEDLE AND A SMALL PLASTIC COMPONENT BECAME LODGED WITHIN THE IV HUB, PREVENTING USE OF THE IV FOR BLOODWORK OR FLUID ADMINISTRATION AND REQUIRING REMOVAL OF THE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25161657 ↗MalfunctionJun 4, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT IV CANNULA IS SUPPOSED TO BE BLOOD LESS. THE FILTER IS NOT WORKING AS IV CANNULAS HAVE BEEN LEAKING. THERE WAS PATIENT INVOLVEMENT, AND NO HARM/ADVERSE EVENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19463841 ↗MalfunctionJun 4, 2024Burst Container or Vessel; Fluid/Blood Leak; Material Puncture/Hole; Burst Container or Vessel; Fluid/Blood Leak; Material Puncture/Hole
IT WAS REPORTED THAT IV CATHETER 24G WAS INSERTED INTO PATIENT AND CAUSED SIGNIFICANT PAIN. PATIENT STARTED BLEEDING FROM THE IV INSERTION SITE (POKE) WHILE IV CATHETER WAS IN. WHEN FLUSHED WITH NORMAL SALINE, IV BLEW. WHEN RN TOOK THE IV OUT AND LOOKED CLOSELY - THE IV HAD "RIDGES" LIKE EXTRA PLASTIC ON CATHETER AND A HOLE. PATIENT GOT A NEW IV INSERTED AND EVERYTHING WAS FINE. THERE WAS PATIENT INVOLVEMENT AND MINIMAL PATIENT HARM.
Patient problems: Pain
MAUDE report 19461211 ↗MalfunctionSep 6, 2022Difficult to Insert; Difficult to Insert
IT WAS REPORTED WHEN STARTING IV'S IT WAS HARD TO BREAK THROUGH THE PATIENTS SKIN WITH THE NEEDLE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15362048 ↗MalfunctionMay 14, 2022Audible Prompt/Feedback Problem; Audible Prompt/Feedback Problem
IT WAS REPORTED THAT THE SAFETY DID NOT ENGAGE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14403076 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K160235. One submission often covers many models, so check the product and lot details in each record.
- Z-1218-2019 · Aug 22, 2018 · TerminatedJelco(R) ViaValve(TM) safety intravenous (I.V.) catheters, 18G x 1 1/4", Model Number REF 3265, MOD 10 The Jelco(R) ViaValve(TM) safety I.V. catheter is a sterile, non pyrogenic,
Other Catheter,intravascular,therapeutic,short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page