AccuDrain® INS8401Recalls & reported problems
Integra Lifesciences Corporation · AccuDrain® External CSF Drainage System with Anti-Reflux Valve · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 10381780023883), received between Jun 29, 2022 and Aug 6, 2026 — classified as 6 injuries, 8 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Fluid/Blood Leak4
- Fracture3
- Crack2
- Disconnection2
- Complete Blockage1
- Device Fell1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Cerebrospinal Fluid Leakage5
- Bacterial Infection1
- Headache1
- Insufficient Information1
- Pneumocephalus1
- Unspecified Infection1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
InjuryAug 6, 2026Crack; Fluid/Blood Leak
A facility reported that the Accudrain device (catlog# unknown) was used on a patient to drain CSF, and the port closest to the head with the yellow cap was leaking; when they attempted to remove the cap, it cracked off. It was reported that there was no injury to the patient, but infection was noted. Additional information regarding the type of infection and any required intervention has been requested.
Patient problems: Unspecified Infection
MAUDE report 50161280 ↗MalfunctionApr 24, 2026Fracture; Device Fell
MW5184806 WAS RECEIVED WITH THE FOLLOWING INFORMATION: "PLASTIC PART OF EXTERNAL VENTRICULAR DRAIN BROKE AND DEVICE FELL TO GROUND." ADDITIONAL INFORMATION WAS SUBSEQUENTLY RECEIVED AS FOLLOWS: 1. TYPE OF EVENT SELECTED IN THE MEDWATCH IS "SERIOUS INJURY". PROVIDE DETAILS OF THE SERIOUS INJURY. ANSWER: THIS IS INCORRECT. THERE WAS NO INJURY. I AM SURE THE EVENT TYPE CHECKED WAS POTENTIALLY HARMFUL EVENT. THERE WAS NO HARM TO THE PATIENT. 2. PLEASE ADVISE THE TYPE OF PROCEDURE FOR WHICH ACCUDRAIN WAS USED. ANSWER: CSF DRAINAGE AFTER SUBDURAL BLEED. 3. WAS THERE CSF LEAKAGE DURING THE INCIDENT? ANSWER: NO 4. IF CSF LEAKED, HOW MUCH WAS IT (IF AN EXACT AMOUNT CANNOT BE PROVIDED, PLEASE PROVIDE AN APPROXIMATE: SMALL AMOUNT, MODERATE, ETC.) ANSWER: N/A 5. AFTER ACCUDRAIN BROKE AND FELL, WAS IT REPLACED WITH ANOTHER DEVICE? ANSWER: YES, ANOTHER ACCUDRAIN WAS ATTACHED. 6. WAS ANY MEDICAL INTERVENTION PERFORMED AS A RESULT OF THIS INCIDENT? ANSWER: NO 7. WAS THERE SIGNIFICANT DELAY TO TREATMENT? IF YES, HOW LONG WAS THE DELAY IN MINUTES? NO ANSWER GIVEN 8. DID THE PATIENT PRESENT WITH CLINICAL SIGNS AND SYMPTOMS, AND IF SO, WHAT WERE THEY? ANSWER: FROM THIS EVENT- NONE. FROM INITIAL INJURY
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24984907 ↗InjuryMar 6, 2026Disconnection
MalfunctionNov 26, 2025Fluid/Blood Leak; Material Split, Cut or Torn; Fluid/Blood Leak; Material Split, Cut or Torn
MalfunctionNov 20, 2025Fracture; Fracture
InjuryApr 3, 2025Disconnection; Disconnection
MalfunctionNov 27, 2024Complete Blockage; Complete Blockage
MalfunctionNov 22, 2024Fluid/Blood Leak; Fluid/Blood Leak
THIS IS 2 OF 2 REPORTS LINKED TO MFG REPORT NUMBER 2648988-2024-00041 FOR THE ACCUDRAIN INS8401 BECAUSE THE CUSTOMER IS NOT SURE WHICH DEVICE WAS IN USE: A FACILITY REPORTED THAT EXTERNAL VENTRICULAR DRAIN (ACCUDRAIN INS8401) WAS PLACED AT THE BEDSIDE FOR IVH AND OBSTRUCTIVE HYDROCEPHALUS, AND SHORTLY THEREAFTER, BLOOD WAS NOTED TO BE LEAKING FROM THE DEVICE JUST AHEAD OF THE BURETTE. NURSING AND NEUROSURGERY ASSESSED THE DEVICE AND NOTED WHAT APPEARED TO BE A CRACK AT THE CONNECTION SITE. THE SYSTEM WAS REPLACED AND RECONNECTED TO THE PATIENT¿S EVD WITHOUT FURTHER LEAKING. THE PATIENT WAS STABLE AT THE TIME BUT SUBSEQUENTLY DIED. THE CUSTOMER REPORTED THAT THE PATIENT'S DEMISE IS UNRELATED TO THE DRAIN ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20758992 ↗MalfunctionOct 15, 2024Fracture; Fracture
MalfunctionSep 12, 2024Break; Break
A FACILITY REPORTED THAT THE ACCUDRAIN WITH ANTI-REFLUX VALVE (INS8401) BROKE PROXIMAL TO THE PATIENT NEAR THE STOPCOCK. THE DRAIN DISCONNECTED AND FELL APART AT THE PATIENT CONNECTION. THERE WAS NO PATIENT INJURY OR DELAY IN TREATMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20204848 ↗InjuryMar 22, 2024Break; Break
A FACILITY REPORTED THAT A PEDIATRIC PATIENT HAD AN ACCUDRAIN (INS8401) SINCE FEBRUARY 8, 2024, AND HAD BEEN ON THE MEDTRONIC DUET PRIOR TO THAT. THE LUER LOCK NEAR THE BAG FELL OFF DURING BAG REPLACEMENT YESTERDAY, AND A NEW ACCUDRAIN WAS PLACED. NO PATIENT INJURY HAS BEEN REPORTED; HOWEVER, THE PATIENT HAD A POSITIVE CEREBROSPINAL FLUID (CSF) CULTURE. THE DRAIN WAS PLACED BEDSIDE IN THE PICU AND IS NOT AVAILABLE AS IT WAS DISCARDED. ADDITIONAL INFORMATION RECEIVED FROM INTEGRA REPRESENTATIVE INDICATES THE FOLLOWING: -THE DRAIN WAS PURCHASED IN FEBRUARY AND THEN PLACED ON THE PATIENT AS SOON AS IN-SERVICE WAS DONE ON 2/8/2024. -THE DRAIN BROKE, AND THE REPORTING INTEGRA REPRESENTATIVE WAS INFORMED ON 2/20/2024. IT WAS ON THE PATIENT FOR QUITE SOME TIME. -THE STAFF IS USED TO A MEDTRONIC DRAIN WHICH HAS A LUER LOCK THAT TIGHTENS THE OPPOSITE DIRECTION FROM OUR ACCUDRAIN (IN RELATION TO THE DRAINAGE BAG REPLACEMENT). IT IS POSSIBLE IT WEAKENED FROM MULTIPLE ATTEMPTS TO LOOSEN IT (BUT IN THE WRONG DIRECTION). -THE BAG WAS REPLACED APPROXIMATELY EVERY OTHER DAY.
Patient problems: Bacterial Infection
MAUDE report 18960111 ↗MalfunctionApr 13, 2023Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
N/A.
Patient problems: Cerebrospinal Fluid Leakage; No Clinical Signs, Symptoms or Conditions
MAUDE report 16738187 ↗InjuryMar 23, 2023Break; Break
A FACILITY REPORTED THAT PATIENT WAS FOUND WITH A DISCONNECTED LUMBAR DRAIN AND A SIGNIFICANT AMOUNT OF CEREBROSPINAL FLUID (CSF) HAD DRAINED INTO THE BED. THE LUMBAR DRAIN DISCONNECTED AT A LOCATION THAT WAS NOT INTENDED TO DISCONNECT (NOT A STOPCOCK/CONNECTION). IT WAS LATER REPORTED THAT IT "LOOKS LIKE THE LUMBAR DRAIN TUBING WAS WELL ADHERED TO THE BACK OF THE PATIENT, SO NO DISLODGEMENT OCCURRED. THE LUMBAR TUBING BROKE AT THE PLACE WHERE THE SMALLER TUBING CONNECTS INSIDE THE LARGER BORE TUBING."
Patient problems: Cerebrospinal Fluid Leakage
MAUDE report 16603421 ↗InjuryJun 29, 2022Break; Break
A FACILITY REPORTED THAT THE ACCUDRAIN WITHOUT THE ANTI-REFLUX VALVE (INS8400) WAS PLACED AT BEDSIDE AND USED FOR MULTIPLE DAY. SUBSEQUENTLY, THE COLLECTION BAG BROKE OFF AT THE BOTTOM OF THE COLLECTION CHAMBER AND EXPOSED THE PATIENT TO DIRECT AIR THROUGH THE BROKEN BOTTOM. IT ALSO SPILLED THE PATIENTS CEREBROSPINAL FLUID (CSF) ALL OVER THE FLOOR AND BED. THE PATIENT HAD TO HAVE A SECOND ACCUDRAIN EVD PLACED AT BEDSIDE. NO FURTHER ADVERSE CONSEQUENCE TO THE PATIENT OCCURRED.
Patient problems: Cerebrospinal Fluid Leakage
MAUDE report 14861471 ↗
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page