Skip to content

OSV II® 909711Recalls & reported problems

Integra LifeSciences Switzerland Sàrl · OSV II® Lumbar Valve System · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10381780034025), received between Dec 6, 2024 and Apr 3, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Component Missing” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Component Missing1
  • Physical Resistance/Sticking1
  • Restricted Flow rate1

Patient problems

  • Headache1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 3, 2026Physical Resistance/Sticking

    A FACILITY REPORTED: "A 29-YEAR-OLD FEMALE PATIENT WAS HOSPITALIZED FOR REMOVAL OF AN INFECTED VENTRICULO-PERITONEAL SHUNT, VENTRICULO-CYSTERNOSTOMY, AND PLACEMENT OF AN EXTERNAL VENTRICULAR SHUNT. DURING INSERTION OF THE INTRA-ABDOMINAL CATHETER WITH AN INTERNAL GUIDEWIRE, THE GUIDEWIRE COULD NOT BE REMOVED. ONCE THE CATHETER WAS INSERTED BEYOND THE NEEDLE, IT BECAME VERY DIFFICULT TO REMOVE THE NEEDLE WITHOUT ALSO REMOVING THE CATHETER. THE CATHETER REMAINED BLOCKED IN THE LUMBAR PUNCTURE NEEDLE, AND THE GUIDEWIRE DAMAGED THE CATHETER." THE EVENT LED TO LESS THAN 30 MINUTES SURGICAL DELAY.

    Patient problems: Insufficient Information

    MAUDE report 24783348 ↗
  • InjuryDec 12, 2025Restricted Flow rate; Restricted Flow rate

    A FACILITY REPORTED THAT THE PATIENT WAS EXPERIENCING HEADACHES AND THE SURGEON THOUGHT THAT IT MAY BE DUE TO UNDER DRAINING OF THE OSV II LUMBAR VALVE (909711). THE VALVE WAS REMOVED AND ANOTHER IMPLANTED. THE INITIAL IMPLANTATION DATE HAS NOT BEEN REPORTED, AND IT IS UNKNOWN WHICH VALVE WAS IMPLANTED DURING REVISION SURGERY. PATIENT IS MALE WITH COMPLICATED INPH. ADDITIONAL INFORMATION RECEIVED AS FOLLOWS: "HIS LAST OSV REVISION WAS (B)(6). WELL UNTIL AUGUST. THEN GRADUALLY INCREASING LOW PRESSURE SYMPTOMS, PARTIALLY MANAGED BY GOING UP ON VP SHUNT SETTING & ASD UNTIL REACHING MAX SETTINGS THERE (20MM MAIN VALVE & 40MM ASD). THE PATIENT ALSO HAS A VP SHUNT IN (SO VP AND LP WITH OUR OSV BEING THE LP SHUNT) AND HAS QUITE COMPLEX MANAGEMENT NEEDS."

    Patient problems: Headache

    MAUDE report 23796597 ↗
  • MalfunctionDec 6, 2024Component Missing; Component Missing

    A FACILITY REPORTED THAT WHEN THE OSV II LUMBAR KIT (909711) WAS OPENED, IT WAS DISCOVERED THAT THE GUIDEWIRE WAS MISSING FROM THE PACK. THE PRODUCT WAS SPECIFICALLY ORDERED FOR THE PATIENT AND THE FACILITY DID NOT HAVE ANOTHER OSVII KIT AVAILABLE TO USE. THE SURGEON WAS ABLE TO INSERT THE CATHETER OK (RARE OCCURRENCE, USUALLY NEED GUIDEWIRE), AND THE SHUNT WAS IMPLANTED. NO OTHER ISSUES OCCURRED DURING THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20860859 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page