OSV II® 909712Recalls & reported problems
Integra LifeSciences Switzerland Sàrl · OSV II® Valve System WITH ANTECHAMBER AND SEPARATE 15CM VENTRICULAR CATHETER · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10381780034032), received between Nov 19, 2021 and May 1, 2026 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Complete Blockage” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Complete Blockage1
- Crack1
- Fluid/Blood Leak1
- Infusion or Flow Problem1
- Material Split, Cut or Torn1
- Material Too Soft/Flexible1
Patient problems
- Cerebrospinal Fluid Leakage1
- No Clinical Signs, Symptoms or Conditions1
- Obstruction/Occlusion1
- Visual Disturbances1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 1, 2026Material Too Soft/Flexible; Material Split, Cut or Torn
A FACILITY REPORTED THAT DURING VENTRICULOPERITONEAL (VP) SHUNT IMPLANTATION FOR HYDROCEPHALUS PROCEDURE, TUNNELING FOR THE PLACEMENT OF THE DISTAL (PERITONEAL) CATHETER WAS PERFORMED WITHOUT COMPLICATIONS. AFTER IMPLANTING THE VENTRICULAR CATHETER INTO THE CEREBRAL VENTRICLE, AT THE TIME OF CONNECTING THE VENTRICULAR CATHETER TO THE VALVE, THE SURGICAL TEAM OBSERVED THAT THE CATHETER PRESENTED A PERFORATION/TEAR AT THE CONNECTION SITE. THE DAMAGED SEGMENT WAS CUT, AND THE CONNECTION ATTEMPT WAS REPEATED; HOWEVER, THE SAME ISSUE OCCURRED THREE CONSECUTIVE TIMES. PROPER COUPLING WAS ONLY ACHIEVED AFTER EXTREMELY DELICATE HANDLING. IN THE NEXT STEP, WHEN POSITIONING THE VALVE IN THE LATERAL POCKET OF THE PATIENT¿S HEAD, THE DEVICE PRESENTED BILATERAL OPENING, WITH THE ONSET OF CEREBROSPINAL FLUID LEAKAGE. ACCORDING TO THE TEAM¿S REPORT, THE MATERIAL APPEARED MORE FRAGILE THAN USUAL, RESEMBLING A DRIED OR BRITTLE CONDITION, BECOMING DAMAGED EVEN UNDER CAREFUL HANDLING AND WITHIN STANDARD TECHNIQUE. THE ISSUE WITH THE PRODUCT DELAYED THE PROCEDURE BY APPROXIMATELY 30 MINUTES; HOWEVER, THERE WAS NO INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25061728 ↗InjuryAug 29, 2024Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
InjuryJun 13, 2022Complete Blockage; Complete Blockage
A FACILITY REPORTED THAT THE OSV II VALVE (909712) WAS NOT DRAINING AND THE PATIENT WAS STARTING TO GO BLIND FROM BOTH EYES. FURTHER INFORMATION RECEIVED INDICATES THAT THE VALVE WAS EXPLANTED DUE TO IT BEING BLOCKED, AND THE INCREASED ICP (INTRACRANIAL PRESSURE) LED TO WEAKENED OPTIC NERVES AND IMPACTED VISION. NO ADDITIONAL INFORMATION HAS BEEN MADE AVAILABLE.
Patient problems: Visual Disturbances
MAUDE report 14674988 ↗InjuryNov 19, 2021Infusion or Flow Problem; Infusion or Flow Problem
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026