MoniTorr™ 10110Recalls & reported problems
Integra Lifesciences Corporation · MoniTorr™ ICP External CSF Drainage and Monitoring System · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10381780070849), received between Oct 2, 2023 and Jul 8, 2024 — classified as 2 injuries.
The most frequently reported specific device problem is “Loose or Intermittent Connection” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K022554, K062599, K191684, K920156); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Loose or Intermittent Connection1
Patient problems
- Cerebrospinal Fluid Leakage1
- Pneumocephalus1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 8, 2024Loose or Intermittent Connection; Loose or Intermittent Connection
A FACILITY REPORTED THAT THE PROXIMAL STOPCOCK ON THE CSF DRAINAGE SYSTEM (INS1100) BECAME LOOSE. THERE WAS NO APPARENT CRACKS IN THE TUBING AND COPIOUS AMOUNT OF CEREBROSPINAL FLUID (CSF) WAS OBSERVED ON THE PILLOW. ADDITIONAL INFORMATION RECEIVED: 1. WHAT IS THE LOT/SERIAL NUMBERS OF THE REPORTED DEVICE? UNKNOWN. 2. WHEN WAS PRODUCT PROBLEM DISCOVERED (DURING INSPECTION, BEFORE/DURING/AFTER A PROCEDURE)? WHILE IN USE WITH THE PATIENT. 3. WAS THERE A DELAY IN SURGERY DUE TO PRODUCT PROBLEM? IF YES, HOW LONG? NO, BUT THE PATIENT REQUIRED ANOTHER PROCEDURE TO REPLACE AN EVD AND A MONITORR. 4. FOR WHAT TYPE OF PROCEDURE WAS PRODUCT OR DEVICE USED? SUBARACHNOID HEMORRHAGE. 5. PATIENT FAMILY HAS CHOSEN TO DE-ESCALATE CARE, PATIENT REMAINS ALIVE AT THIS TIME.
Patient problems: Cerebrospinal Fluid Leakage
MAUDE report 19697676 ↗InjuryOct 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING WAS INITIALLY REPORTED BY A FACILITY: "AIR FORFEITURE IN THE PATIENT LINE WITH GREEN STRIPE. AT CT CHECK ENCEPHALUS SHOWED THE PRESENCE OF PNEUMOVENTRICLE NOT VISIBLE AT PREVIOUS CHECK AND THEREFORE UNRELATED TO THE RECENT PLACEMENT OF THE SHUNT BUT PROBABLY TO THE DRAINAGE DM." ADDITIONAL INFORMATION WAS LATER RECEIVED INDICATING THE FOLLOWING: "NEUROSURGICAL PATIENT WAS ADMITTED ON (B)(6) 2023 FOR HEMORRHAGE SUBARACHNOID FROM RUPTURED ANEURYSM FOR WHICH SHE WAS PLACED IN ROOM. IN OPERATING ROOM, AN EXTERNAL VENTRICULAR SHUNT (DVE) REF.10110. ARRIVED FROM THE OPERATING ROOM TO THE INTENSIVE CARE UNIT. THE DVE WITH THE DRAINAGE AND MONITORING SYSTEM AND PROCEEDED TO DELIQUATION. THE EXTERNAL DRAINAGE AND MONITORING SYSTEM IS A CLOSED DESIGNED TO DRAIN THE CAFELORACHIDIAN FLUID (LCR) AND POSSIBLE BLOOD ALLOWING MONITORING OF INTRACRANIAL PRESSURE."
Patient problems: Pneumocephalus
MAUDE report 17851128 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K022554, K062599, K191684, K920156. One submission often covers many models, so check the product and lot details in each record.
- Z-1311-2019 · Apr 8, 2019 · TerminatedMoniTorr INS1100 INS-1100 CSF Drainage System used with Pole Mount System Product Usage: The MoniTorr ICP External CSF Drainage and Monitoring Systems provide a simple to use,
- Z-1310-2019 · Apr 8, 2019 · TerminatedMoniTorr 10100 10-100 CSF Drainage System w/Patient Line One Way Valve MoniTorr 10102 10-102 CSF Drainage System w/Patient Line One Way Valve w MoniTorr 10110 10-110 CSF Drainage
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page