Skip to content

MicroFrance® MCL206Recalls & reported problems

Integra Microfrance · MOUTH GAG MCL206 MORINIERE · Class 1

All models of MicroFrance (22)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10381780105220), received between Mar 3, 2025 and Mar 3, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Mechanical Jam” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam1

Patient problems

  • Insufficient Information1
  • Swelling/ Edema1
  • Tooth Fracture1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 3, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE RIGHT DISTAL SCREW OF THE MORINIERE MOUTH GAG (MCL206) "JAMMED DURING SURGERY"; THE USER WAS UNABLE TO ADVANCE THE RACK. THE DEVICE HAD BEEN LUBRICATED, BUT "OPERATED JERKILY." THE FOLLOWING INFORMATION WAS SUBSEQUENTLY REPORTED: "COMPLICATED EXPOSURE, ASSISTANT TOO FAR AWAY, NO LIGHT IN THE MOUTH, ¿SEIZED¿ MOLINIÈRE ROBOT MOUTH OPENER, VERY DIFFICULT TO HANDLE, SCREW, UNSCREW. REQUESTS A NASOTRACHEAL INTUBATION; NEW NASOTRACHEAL INTUBATION WITH EDEMA AND BLEEDING. CLAMPING OVER PREVIOUS CLAMPING. THE MOLINIÈRE MOUTH OPENER REMAINS STUCK IN THE OPEN POSITION IN THE PATIENT, VERY DIFFICULT TO REMOVE IT. RESULT OF THE EXTEMPO: CANCER CELLS ON THE TONSILLAR PILLAR. NEW ATTEMPT AT EXPOSURE BY LUBRICATING THE MOUTH OPENER WITH THE ANESTHETISTS¿ SILICONE SPRAY. EDEMA, BLEEDING, TOOTH BREAKAGE ON THE UPPER INCISOR (21) AND GUM WOUND, EXPOSURE MADE DANGEROUS BY THE DEFECTIVE MOLINIÈRE MOUTH OPENER. NO ROBOT-ASSISTED INTERVENTION, EXCISION OF THE RIGHT TONSIL ONLY. CONSEQUENCES: TUMOR EXCISION NOT PERFORMED: CHANGE IN THERAPEUTIC MANAGEMENT (RADIOTHERAPY)." SIMPLE TONSILLECTOMY AND RADIOTHERAPY MANAGEMENT WAS PERFORMED TO COMPLETE THE PROCEDURE AS EXCISION WAS NOT P

    Patient problems: Insufficient Information; Swelling/ Edema; Tooth Fracture

    MAUDE report 21504668 ↗

Other GAG, MOUTH devices

All GAG, MOUTH devices →

Sources and updates