Integra® INS030Recalls & reported problems
Integra Pain Management · Integra® Hand Drill - Sterile
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10381780263890), received between Aug 21, 2024 and Aug 21, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 21, 2024Device-Device Incompatibility; Device-Device Incompatibility
A FACILITY REPORTED THAT DURING NEUROSURGERY, AT THE END OF PROCEDURE, IT WAS NOTICED THAT ONE SCREW WAS COMING OUT FROM THE "HAND DRILL: STERILE (INS030)", DESPITE A PROPER FIXATION. THE PATIENT DID NOT HAVE ANY CLINICAL CONSEQUENCES (DESPITE THE RISK OF LOSING CONTROL OF THE INSTRUMENT THAT COULD HAVE LED TO CRANIAL BLOOD LOSS). ADDITIONAL INFORMATION RECEIVED AS FOLLOWS: DESCRIPTION: THE DRILL BIT COMES OUT OF THE HAND DRILL BECAUSE IT WAS NOT HELD PROPERLY DESPITE PROPER PLACEMENT. WHICH TYPE OF DRILL BIT WAS USED WITH INS030? DRILL BIT CONTAINED IN THE SOPHYSA KIT (PRESSIO, NOT INTEGRA PRODUCT) REFERENCE AND LOT NUMBER IF KNOWN? PSO-PB (PRESSIO® ICP MONITORING KIT, PARENCHYMAL WITH BOLT), LOT UNKNOWN. NOT INTEGRA PRODUCT. DRILL BIT DIAMETER? 2.7MM. WHICH PROCEDURE: IMPLANT OF INTRACRANIAL PRESSURE SENSOR. DID IT LEAD TO ANY INCREASE OF SURGERY TIME? ABOUT 3 MINUTES FOR A PROCEDURE WHICH USUALLY TAKES 15 MIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20013563 ↗
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page