Skip to content

Omni-Tract® 3335Recalls & reported problems

Integra Lifesciences Corporation · Omni-Tract® Mayo Swivel Retractor, 2-3/4in x 2in (7.0cm x 5.1cm) · Class 1

All models of Omni-Tract (447)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Jun 5, 2026Class IIOpen, Classified

    The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.

    Z-2714-2026 · Root cause: Labeling design · Recall details · FDA record ↗

FDA clearance and classification

The FDA regulates product code FFO (retractor, self-retaining) under 21 CFR 876.4730 as a Class I device. 21 CFR 876.4730 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Omni-Tract® 3335

Has the Omni-Tract® 3335 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2714-2026 (Class II), began on Jun 5, 2026: The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the Omni-Tract® 3335 FDA approved?

Devices of this type (product code FFO) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Integra Lifesciences Corporation.

Who makes the Omni-Tract® 3335, and is it still sold?

The FDA's device database lists Integra Lifesciences Corporation as the company responsible for it (brand Omni-Tract®, model 3335). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other RETRACTOR, SELF-RETAINING devices

All RETRACTOR, SELF-RETAINING devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page