Skip to content

CODMAN® HOLTER® 821654Recalls & reported problems

Integra Lifesciences Production Corporation · CODMAN® HOLTER® Ventricular Catheter 4cm · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10381780517900), received between Feb 21, 2025 and Feb 21, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Material Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation1
  • Migration1

Patient problems

  • Fatigue1
  • Headache1
  • Vomiting1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 21, 2025Material Separation; Migration; Material Separation; Migration

    A FACILITY REPORTED A VENTRICULAR CATHETER (ID 821654) WAS IMPLANTED ON (B)(6) 2024. THE PATIENT PRESENTED TO THE EMERGENCY ROOM WITH SYMPTOMS OF INTRACRANIAL HYPERTENSION (VOMITING, FATIGUE AND HEADACHES). A BRAIN SCAN WAS PERFORMED URGENTLY, SHOWING THE DISCONNECTION OF THE PROXIMAL CATHETER FROM THE VALVE AND ITS MIGRATION INTO THE INTRAPARENCHYMAL TOWARDS THE BRAINSTEM. THEREFORE, THE PROXIMAL CATHETER WAS REMOVED AND REPLACED. ALL PIECES WERE RECOVERED, AND INTRAVENTRICULAR BLEEDING WAS REPORTED.

    Patient problems: Fatigue; Headache; Vomiting

    MAUDE report 21432799 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page