Skip to content

CODMAN® 823041Recalls & reported problems

Integra Lifesciences Production Corporation · CODMAN® Ventricular Catheter Straight · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10381780518815), received between Aug 22, 2023 and Oct 17, 2025 — classified as 5 injuries.

The most frequently reported specific device problem is “Obstruction of Flow” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Obstruction of Flow3
  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • Cerebral Ventriculomegaly2
  • Insufficient Information2
  • Cerebral Edema1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 17, 2025Obstruction of Flow; Obstruction of Flow

    N/A.

    Patient problems: Insufficient Information

    MAUDE report 23322777 ↗
  • InjurySep 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS IS 1 OF 2 REPORTS LINKED TO MFG REPORT NUMBER: 3013886523-2024-00267 A PHYSICIAN REPORTED A VENTRICULAR CATHETER (ID 823041) AND A PERITONEAL CATHETER (ID 823045) WERE IMPLANTED VIA VENTRICULOPERITONEAL (VP) SHUNT ON UNKNOWN DATE WITH UNKNOWN SETTING. THE CATHETERS WERE USED TOGETHER WITH A CERTAS PLUS VALVE (ID 828804). ALL THREE DEVICES WERE REMOVED ON AN UNKNOWN DATE DUE TO THE VALVE NOT FUNCTIONING PROPERLY AFTER IMPLANTATION. BASED ON INFORMATION PROVIDED, IT IS UNKNOWN IF THERE WAS AN ISSUE WITH THE CATHETERS, HOWEVER, THEY WERE EXPLANTED. IT IS ALSO UNKNOWN IF THE PATIENT EXPERIENCED ANY SIGNS OR SYMPTOMS DUE TO VALVE FAILURE. THE PATIENT'S CONDITION IS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 20150751 ↗
  • InjuryAug 23, 2024Obstruction of Flow; Obstruction of Flow

    A PHYSICIAN REPORTED A VENTRICULAR CATHETER (ID 823041) AND A CERTAS VALVE WERE IMPLANTED WITH UNKNOWN SETTING DUE TO HYDROCEPHALUS. SUBCUTANEOUS CSF ACCUMULATION WAS OBSERVED AROUND THE DEVICE. THE VALVE'S PATENCY WAS CONFIRMED THROUGH RADIO ISOTOPES. THEREFORE, THE VALVE AND THE CATHETER WERE REMOVED AND REPLACED. WHEN IMPLANTING THE NEW VENTRICULAR CATHETER, THE INSERTION LOCATION WAS CHANGED FROM FRAZIER¿S POINT TO KOCCHER¿S POINT. IT WAS STATED THAT, THE SURGEON COULD NOT THINK OF THE CAUSATION OF SUBCUTANEOUS RETENTION OTHER THAN THE VALVE BREAKAGE. IT WAS CONFIRMED THAT THE VALVE WAS NOT BROKEN, HOWEVER, THERE WAS A SUSPICION OF VENTRICULAR CATHETER BLOCKAGE. THE PATIENT IS CURRENTLY IN FOLLOW UP. ACCORDING TO INFORMATION PROVIDED, IT IS UNKNOWN IF THE PATIENT EXPERIENCED ANY SIGNS AND SYMPTOMS DUE TO SUSPECTED CATHETER BLOCKAGE.

    Patient problems: Cerebral Edema

    MAUDE report 20051374 ↗
  • InjuryFeb 27, 2024Obstruction of Flow; Obstruction of Flow

    N/A.

    Patient problems: Cerebral Ventriculomegaly

    MAUDE report 18787810 ↗
  • InjuryAug 22, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS IS 2 OF 2 REPORTS LINKED TO MFG REPORT NUMBERS: 3013886523-2023-00280. A PHYSICIAN REPORTED A CERTAS VALVE (ID 828814) AND HAKIM VENTRICULAR CATHETER (ID 823041) WERE IMPLANTED VIA VENTRICULAR PERITONEAL SHUNT ON (B)(6), 2023 WITH SETTING 3. THE PATIENT VISITED THE HOSPITAL ON (B)(6) WITH RECURRENCE OF SYMPTOMS. ON (B)(6) 2023, VENTRICULAR ENLARGEMENT WAS CONFIRMED BY X-RAY EXAMINATION. THE CHAMBER COULD NOT BE PUSHED, AND THE FLOW HAD STOPPED. DUE TO SUSPICION OF OBSTRUCTION, THE VALVE AND CATHETER WERE REMOVED AND REPLACED ON (B)(6), 2023.

    Patient problems: Cerebral Ventriculomegaly

    MAUDE report 17596608 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page