Skip to content

CODMAN® HAKIM® 823115Recalls & reported problems

Integra Lifesciences Production Corporation · CODMAN® HAKIM® Programmable Valve Programmable Valve Cylindrical Valve Programmable in steps of 10mm H2O (98 Pa): 30mm H2O to 200mm H2O (294 Pa - 1960 Pa) · Class 2 · Implant

All models of CODMAN HAKIM (36)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10381780519188), received between Apr 5, 2023 and Mar 30, 2026 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak1
  • Improper Flow or Infusion1
  • Obstruction of Flow1

Patient problems

  • Cerebrospinal Fluid Leakage1
  • Discomfort1
  • Insufficient Information1
  • Vomiting1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 30, 2026Fluid/Blood Leak

    A FACILITY REPORTED A HAKIM VALVE (ID 823115) HAD CEREBROSPINAL FLUID LEAKAGE DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED WITH A REPLACEMENT PRODUCT AVAILABLE. NO PATIENT INJURY REPORTED AND THE EVENT DID NOT LED TO SURGICAL DELAY.

    Patient problems: Cerebrospinal Fluid Leakage

    MAUDE report 24712230 ↗
  • InjurySep 5, 2024Improper Flow or Infusion; Improper Flow or Infusion

    A FACILITY REPORTED A HAKIM VALVE (ID 823115) WAS IMPLANTED ON (B)(6) 2023. APPROXIMATELY 6 MONTHS AFTER IMPLANTATION, THE PATIENT SHOWED REPEATED PHASES OF VOMITING, DISCOMFORT AND DROWSINESS. THE IMAGING SHOWED NARROW VENTRICLES, SO MEDICAL STAFF SUSPECTED OVERDRAINAGE. THE VALVE OPENING PRESSURE WAS ADJUSTED, AND THE PATIENT FELT BETTER, WITH A DROP IN THE "AZ" IN THE DAYS THAT FOLLOWED. THEREFORE, THE VALVE WAS REMOVED AND REPLACED WITH A HAKIM VALVE WITH SIPHONGUARD ON (B)(6) 2024. THE PATIENT RECOVERED.

    Patient problems: Discomfort; Vomiting

    MAUDE report 20150907 ↗
  • InjuryApr 5, 2023Obstruction of Flow; Obstruction of Flow

    N/A.

    Patient problems: Insufficient Information

    MAUDE report 16684206 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →

Sources and updates