The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10381780519188), received between Apr 5, 2023 and Mar 30, 2026 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak1
- Improper Flow or Infusion1
- Obstruction of Flow1
Patient problems
- Cerebrospinal Fluid Leakage1
- Discomfort1
- Insufficient Information1
- Vomiting1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2026Fluid/Blood Leak
A FACILITY REPORTED A HAKIM VALVE (ID 823115) HAD CEREBROSPINAL FLUID LEAKAGE DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED WITH A REPLACEMENT PRODUCT AVAILABLE. NO PATIENT INJURY REPORTED AND THE EVENT DID NOT LED TO SURGICAL DELAY.
Patient problems: Cerebrospinal Fluid Leakage
MAUDE report 24712230 ↗InjurySep 5, 2024Improper Flow or Infusion; Improper Flow or Infusion
A FACILITY REPORTED A HAKIM VALVE (ID 823115) WAS IMPLANTED ON (B)(6) 2023. APPROXIMATELY 6 MONTHS AFTER IMPLANTATION, THE PATIENT SHOWED REPEATED PHASES OF VOMITING, DISCOMFORT AND DROWSINESS. THE IMAGING SHOWED NARROW VENTRICLES, SO MEDICAL STAFF SUSPECTED OVERDRAINAGE. THE VALVE OPENING PRESSURE WAS ADJUSTED, AND THE PATIENT FELT BETTER, WITH A DROP IN THE "AZ" IN THE DAYS THAT FOLLOWED. THEREFORE, THE VALVE WAS REMOVED AND REPLACED WITH A HAKIM VALVE WITH SIPHONGUARD ON (B)(6) 2024. THE PATIENT RECOVERED.
Patient problems: Discomfort; Vomiting
MAUDE report 20150907 ↗InjuryApr 5, 2023Obstruction of Flow; Obstruction of Flow
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page