CODMAN® HAKIM® 823164 (REF 82-3164)Recalls & reported problems

Integra Lifesciences Production Corporation · CODMAN® HAKIM® Programmable Valve In-Line Valve Programmable in steps of 10mm H2O (98 Pa): 30mm H2O to 200mm H2O (294 Pa - 1960 Pa) · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10381780519249), received between Jun 14, 2022 and Jan 8, 2026 — classified as 8 injuries, 1 malfunction.

The most frequently reported specific device problem is “Obstruction of Flow” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Obstruction of Flow4
  • Insufficient Device Problem Information2
  • Failure to Align1
  • Infusion or Flow Problem1
  • Output Problem1

Patient problems

  • Insufficient Information4
  • Convulsion/Seizure1
  • Decreased Appetite1
  • Dizziness1
  • Failure of Implant1
  • Fever1
  • High Blood Pressure/ Hypertension1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 8, 2026Failure to Align

    N/A.

    Patient problems: Insufficient Information

    MAUDE report 24011067 ↗
  • InjuryOct 17, 2025Obstruction of Flow; Obstruction of Flow

    N/A.

    Patient problems: Insufficient Information

    MAUDE report 23322778 ↗
  • InjuryAug 21, 2025Output Problem; Output Problem

    N/A.

    Patient problems: Insufficient Information

    MAUDE report 22864244 ↗
  • InjuryJun 13, 2025Infusion or Flow Problem; Infusion or Flow Problem

    N/A.

    Patient problems: Sleepiness/Drowsiness/Somnolence

    MAUDE report 22201635 ↗
  • InjuryJun 6, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    N/A.

    Patient problems: Convulsion/Seizure; Fever; High Blood Pressure/ Hypertension

    MAUDE report 22159957 ↗
  • InjuryNov 26, 2024Obstruction of Flow; Obstruction of Flow

    A PHYSICIAN REPORTED A MALFUNCTION DURING THE USE OF CODMAN PROGRAMMABLE INLINE VALVE. "ON (B)(6) 2024, A PHYSICIAN IMPLANTED CODMAN PROGRAMABLE INLINE VALVE (ID 823164). THE LOT/BATCH NUMBER AND EXPIRY DATE WERE NOT REPORTED. THE DEVICE WORKED FOR 24 HOURS ONLY THEN AUTOMATICALLY STOPPED/BLOCKED (MALFUNCTION OF SHUT CHAMBER). WHILE WORKING THROUGH DIGITALLY, IT WAS FUNCTIONING BUT NOT SPONTANEOUSLY, EVEN AFTER THE PRESSURE WAS REDUCED TO 70 MM/ H²O BY THE PHYSICIAN. IT WAS REPORTED THAT, THERE WAS NO CSF DRAINAGE. PHYSICIAN HAD TO REMOVE THE SHUNT AND CONVERT IT INTO AN EVD." REMARKS: "THE CAUSALITY FOR THE REPORTED EVENT IS CONSIDERED AS NOT ASSESSABLE DUE TO LACK OF INFORMATION REGARDING CIRCUMSTANCES LEADING TO THE EVENT, CO-SUSPECTS, CONCOMITANT MEDICATIONS, INDICATION, CONCURRENT CONDITIONS, RELEVANT MEDICAL HISTORY AND PAST DRUG HISTORY FOR COMPLETE ASSESSMENT."

    Patient problems: Insufficient Information

    MAUDE report 20783733 ↗
  • MalfunctionNov 15, 2024Obstruction of Flow; Obstruction of Flow

    A FACILITY REPORTED THAT AFTER IMPLANTATION OF A HAKIM VALVE (ID 823164) DURING PROCEDURE, CEREBROSPINAL FLUID (CSF) DID NOT FLOW OUT. THE VALVE WAS REPLACED, AND THE FLOW OCCURRED WITHOUT ANY ISSUES AFTER THE REPLACEMENT. THE EVENT HAPPENED BEFORE IMPLANTATION SITE CLOSURE, NO PATIENT INJURY REPORTED; HOWEVER, THE EVENT LED TO 30 MINUTES SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20694709 ↗
  • InjuryJul 5, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    N/A.

    Patient problems: Decreased Appetite; Dizziness; Vomiting

    MAUDE report 19680242 ↗
  • InjuryJun 14, 2022Obstruction of Flow; Obstruction of Flow

    A PHYSICIAN REPORTED A HAKIM VALVE (ID 823164) WAS IMPLANTED VIA V-P SHUNT ON UNKNOWN DATE WITH UNKNOWN SETTINGS TO TREAT A MIDLINE FISSURE CYST. A CATHETER COMING FROM THE EPIDURAL AND A CATHETER COMING FROM THE VENTRICLE WERE CONNECTED WITH A Y-SHAPED CONNECTOR AND CONNECTED TO THE VALVE. AN OBSTRUCTION WAS SUSPECTED THEREFORE, THE VALVE WAS REMOVED AND REPLACED ON (B)(6) 2022. BASED ON INFORMATION PROVIDED, IT IS UNKNOWN IF THE PATIENT EXPERIENCED ANY SIGNS AND SYMPTOMS.

    Patient problems: Failure of Implant

    MAUDE report 14687893 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →

Sources and updates