CODMAN CERTAS® Plus 828800PLRecalls & reported problems

Integra LifeSciences Switzerland Sàrl · CODMAN CERTAS® Plus Programable Valve Inline Valve Only Includes: Priming Adapter and Information Manual · Class 2 · Implant

Adverse event reports
14
Deaths reported
0
Injuries reported
11
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 10381780521013), received between Apr 5, 2023 and Dec 5, 2025 — classified as 11 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Device Difficult to Program or Calibrate” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Difficult to Program or Calibrate4
  • Insufficient Device Problem Information2
  • No Flow2
  • Contamination1
  • Imprecise or Erratic Results1
  • Improper Flow or Infusion1
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Infusion or Flow Problem1

Patient problems

  • Insufficient Information7
  • No Clinical Signs, Symptoms or Conditions4
  • Coma1
  • Confusion/ Disorientation1
  • Dizziness1
  • Headache1
  • Nausea1
  • Vomiting1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 5, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    N/A.

    Patient problems: Confusion/ Disorientation; Headache; Vomiting

    MAUDE report 23728865 ↗
  • InjuryApr 4, 2025Improper Flow or Infusion; Improper Flow or Infusion

    A PHYSICIAN REPORTED A PATIENT HAS A CERTAS VALVE (ID 828804PL) SET AT 7. THE VALVE OVER DRAINS WHEN THE PATIENT IS SITTING UP AND NORMALIZES WHEN LYING DOWN. THE VALVE WAS TESTED WITH A MANOMETER AND APPEARS TO BE DRAINING COMPLETELY. WHEN THE VALVE IS SET TO 8, IT APPEARS TO WORK AND HOLD STEADY. THE VALVE WAS REMOVED AND REPLACED ON (B)(6) 2025. ACCORDING TO THE INFORMATION RECEIVED, IT IS UNKNOWN IF THE PATIENT HAD ANY SIGNS OR SYMPTOMS. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 21773370 ↗
  • InjuryJan 29, 2025Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    SALES REPRESENTATIVE NARRATIVE: "THE PHYSICIAN REMOVED AND REVISED SHUNT ON (B)(6) 2025. HE NOTIFIED ME ON (B)(6) 2025, THAT THE VALVE THAT HE REMOVED WOULD NOT REPROGRAM. HE REPLACED THE VALVE WITH A NEW CERTAS VALVE. A DIFFERENT NEUROSURGEON IMPLANTED THE VALVE A FEW YEARS AGO".

    Patient problems: Insufficient Information

    MAUDE report 21264699 ↗
  • InjurySep 20, 2024Infusion or Flow Problem; Infusion or Flow Problem

    A FACILITY REPORTED A CERTAS VALVE (ID 828804PL) WAS NOT DRAINING CEREBROSPINAL FLUID (CSF), ( THE VENTRICLES WERE NOT DRAINING). THEREFORE THE VALVE WAS EXPLANTED AND REPLACED 3 DAYS AFTER IMPLANTATION. NO SURGICAL DELAY AND THE PATIENT IS DOING WELL.

    Patient problems: Insufficient Information

    MAUDE report 20274620 ↗
  • MalfunctionSep 20, 2024No Flow; No Flow

    A FACILITY REPORTED: "SURGEON CONNECTED CERTAS PLUS IN-LINE WITH MEDTRONIC CSF RESERVOIR. UPON CONNECTING THE SYSTEM WITH VENTRICULAR AND PERITONEAL CATHETERS, THE SURGEON DID NOT EXPERIENCE FLOW. THE CERTAS PLUS WAS SET TO SETTING 4. MULTIPLE ATTEMPTS WERE MADE TO VERIFY FLOW, BUT NONE OF THEM WORKED INCLUDING USING THE OCCLUDE-PUMP-RELEASE TECHNIQUE AND INJECTING LIQUID INTO THE CERTAS PLUS RESERVOIR WHILE THE SHUNT INLET WAS OCCLUDED. EVENTUALLY A NEW CERTAS PLUS VALVE WAS USED AND IT DID FLOW AS EXPECTED. BOTH TIMES THE CERTAS PLUS WAS SET TO SETTING 4. THE VALVE ISSUE WAS DETECTED DURING IMPLANTATION PROCEDURE, AND THE EVENT LED TO APPROXIMATELY 5 MINUTES SURGICAL DELAY. SURGEON STATED THAT HE WAS NOT SURE THAT ISSUE WAS CAUSED BY THE CERTAS PLUS SHUNT AS THIS WAS A RELATIVELY COMPLEX SHUNT CASE. THE CERTAS PLUS VALVE WAS USED WITH: VENTRICULAR CATHETER (MEDTRONIC 44104), CSF RESERVOIR (MEDTRONIC 44104), AND PERITONEAL CATHETER FROM MEDTRONIC.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20274618 ↗
  • InjuryMay 2, 2024No Flow; No Flow

    A FACILITY REPORTED A CERTAS VALVE (ID 828800PL) WAS NOT FLOWING. PHYSICIAN TRIED TO SET THE PRESSURE TO 5 AND ENDED IN 4. THE VALVE WAS EXPLANTED AND REPLACED.

    Patient problems: Insufficient Information

    MAUDE report 19234557 ↗
  • MalfunctionApr 11, 2024Pressure Problem; Pressure Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19091177 ↗
  • InjuryFeb 20, 2024Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    A FACILITY REPORTED A CERTAS VALVE (ID 828800PL) WAS UNABLE TO CHANGE SETTING AFTER IMPLANTATION PROCEDURE. THE PHYSICIAN ATTEMPTED TO PROGRAM THE VALVE TO A DIFFERENT SETTING, BUT IT REMAINED UNCHANGEABLE. AS A RESULT, THE SURGEON HAD TO REMOVE THE VALVE FROM THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 18740863 ↗
  • InjuryFeb 7, 2024Imprecise or Erratic Results; Imprecise or Erratic Results

    N/A.

    Patient problems: Dizziness; Nausea

    MAUDE report 18664029 ↗
  • InjuryDec 18, 2023Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    A FACILITY REPORTED THAT A CERTAS VALVE (ID 828800PL) SEEMS TO BE STUCK AFTER BEING IMPLANTED VIA VENTRICULOPERITONEAL SHUNT AND WAS UNABLE TO REPROGRAM WITH MANUAL NOR ELECTRONIC TOOL KIT. THE PATIENT UNDERWENT SHUNT REVISION.

    Patient problems: Insufficient Information

    MAUDE report 18346058 ↗
  • MalfunctionAug 22, 2023Contamination; Contamination

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17596155 ↗
  • InjuryJul 27, 2023Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    A FACILITY REPORTED A CERTAS VALVE (ID (B)(4)) WAS IMPLANTED ON (B)(6) 2020 AND THE SURGEON HAD AN ISSUE IN WHICH THE CERTAS VALVE WAS UNABLE TO BE CHANGED FROM SETTING NUMBER 5 DESPITE MULTIPLE ATTEMPTS WITH DIFFERENT PROGRAMMERS. THE VALVE WAS EXPLANTED ON (B)(6) 2023.

    Patient problems: Insufficient Information

    MAUDE report 17408938 ↗
  • InjuryJun 21, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A FACILITY REPORTED A CERTAS VALVE (ID 828800PL) WAS IMPLANTED IN (B)(6) 2023, COMPLICATED BY A HYGROMA AND THEN A SUBDURAL HEMATOMA. THE VALVE CHANGED SETTINGS WITHOUT INTERVENTION. THE SHUNT WAS TURNED TO 8, AND WHEN VENTRICLES BLEW UP, THE HEMATOMA WAS EVACUATED. BEFORE DISCHARGE, THE PATIENT HAD CT SCAN THAT SHOWED VERY SMALL VENTRICLES. X-RAY OF THE VALVE LOOKED LIKE 8 BUT A LITTLE OFF. IT WAS ADJUSTED AGAIN TO 8, HOWEVER, THEY THINK THEY DIDN'T DO ANYTHING. THE PATIENT WAS DISCHARGED AND WAS BACK THE NEXT DAY WITH LARGE VENTRICLES AGAIN. ON (B)(6) 2023, MEDICAL STAFF CHANGE THE VALVE TO 7/8 AND CONFIRMED ON XRAY. THE PATIENT HAS HAD A DIFFICULT COURSE; THEREFORE, IT WAS TAKING IT DOWN TO 6/8, HOWEVER, ON (B)(6) 2023 X-RAY SHOWED THE VALVE WAS AT 8/8. IT WAS ADJUSTED DOWN TO 6/8, HOWEVER, X-RAY LATER THAT DAY SHOWED IT IS AT 5/8. THE VALVE WAS REMOVED AND REPLACED WITH CERTAS PLUS SIPHONGUARD ON (B)(6) 2023. ACCORDING TO INFORMATION PROVIDED: MEDICAL STAFF BELIEVE THE VALVE WAS ACCEPTABLE WHEN INITIALLY IMPLANTED, BUT FAILED AFTER 6 WEEKS OF BEING IMPLANTED. IT IS POTENTIALLY HARD TO DETERMINE IF THE VALVE MALFUNCTIONED AS THERE ARE MANY FACTORS SUCH AS HUMAN ERROR WITH PROGRA

    Patient problems: Coma

    MAUDE report 17171865 ↗
  • InjuryApr 5, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    A FACILITY REPORTED A CERTAS VALVE (ID 828800PL) WAS IMPLANTED AND THE PATIENT WAS BROUGHT BACK TO SURGERY HOURS LATER TO REPLACE THE VALVE. ACCORDING TO REPORTER: "VALVE WAS EITHER NOT FLOWING PROPERLY OR THERE WAS A PINCH IN THE CATHETER LEADING TO VENTRICLE. I BELIEVE IT WAS PINCHED BUT TO BE CERTAIN WE SWAPPED THE VALVE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16686325 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.