Integra LifeSciences Switzerland Sàrl · CODMAN® CERTAS® Plus Programmable Valve Inline Small Valve with Accessories · Class 2 · Implant

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10381780535904), received between Jun 6, 2025 and Jun 6, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Output Problem” (1 report).

1 other recall cover devices cleared under the same FDA submission (K223330); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 6, 2025Output Problem; Output Problem

    A PHYSICIAN REPORTED A CERTAS VALVE (ID 828811PL) WAS IMPLANTED ON (B)(6) 2016, WITH A SETTING 4. ON (B)(6) 2017, THE PATIENT SETTING WAS SET TO 5. THE PATIENT HAD MAGNETIC RESONANCE IMAGING (MRI) ON (B)(6) 2023 AND WAS SET AT 5. ON (B)(6) 2025, THE PATIENT HAD A CRANIAL MRI FOR A 2-YEAR FOLLOW-UP. HOWEVER, THE SHUNT WAS MOVED AND SET AT 8. THEY WERE UNABLE TO SET THE VALVE TO 5. THEREFORE, A REPROGRAMMING WAS DONE UNDER FLUOROSCOPY, AND THE SHUNT WAS STILL SET TO 8. DUE TO THE ISSUE, THE VALVE WAS REMOVED AND REPLACED ON (B)(6) 2025. THE PATIENT DID NOT EXPERIENCE ANY SIGNS OR SYMPTOMS DUE TO THE PRODUCT ISSUE. THE PATIENT CONDITION IS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22159991 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K223330. One submission often covers many models, so check the product and lot details in each record.

  • Z-0978-2025 · Dec 16, 2024 · Open, Classified
    CODMAN CERTAS Plus Programmable Valve Inline Small Valve Only. Model Number: 828810PL. Implantable device that provides constant intraventricular pressure and drainage of CSF for

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.