BD BBL™ Columbia Agar with 5% Sheep Blood 221263Recalls & reported problems

Becton, Dickinson And Company · BD BBL™ Columbia Agar with 5% Sheep Blood, Carton of 100 plates · Class 1

All models of BD BBL Columbia Agar with 5% Sheep Blood (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This BD BBL Columbia Agar blood plate is used in laboratories to grow bacteria. Most of its 210 reports describe incorrect or unclear results, with Staphylococcus and Streptococcus not showing the expected reaction on the plate. The reports include no deaths or injuries and no patient impact. No recalls name this device. There were no reports in the last 12 months, compared with 199 in the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
210
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 210 adverse event reports that list this exact device (DI 10382902212635), received between Feb 17, 2021 and Aug 20, 2025 — classified as 210 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (205 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 199 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports202420252026199
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings205
  • Device Markings/Labelling Problem5
  • Contamination1

Patient problems

  • No Clinical Signs, Symptoms or Conditions210
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 210 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 20, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22846905 ↗
  • MalfunctionAug 8, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22746201 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22742389 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22742345 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22742328 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22740433 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22740422 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22740418 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22740397 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22740374 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22740366 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22740365 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22740361 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22740359 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22740352 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22740340 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22739271 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22739269 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22739247 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22739246 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22739245 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22739243 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22739241 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22739240 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL COLUMBIA AGAR WITH 5% SHEEP BLOOD PLATE 100PK THAT STAPHYLOCOCCUS AND STREPTOCOCCUS WERE NOT EXHIBITING THE EXPECTED HEMOLYSIS ON THE BLOOD AGAR PLATE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22739239 ↗

Other CULTURE MEDIA, ENRICHED devices

All CULTURE MEDIA, ENRICHED devices →