Omnipod DASH PDM PDM-USA1-D001-MG-USA1Recalls & reported problems
Insulet Corporation · Omnipod DASH Personal Diabetes Manager (PDM) mg/dL · Class 2
All models of Omnipod DASH (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Omnipod DASH controller is the handheld device used with insulin pods. Of its 1,130 reports, 10 describe a patient death, 343 describe injuries and 776 describe malfunctions. Many reports describe battery problems, controllers that overheat, and devices that fail to power up. Reports fell to 42 in the last 12 months from 78 in the previous 12 months. 1 closed recall, from October 2022, concerns batteries that swell, leak and overheat, posing a fire risk.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,129
- Deaths reported
- 10
- Injuries reported
- 343
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1,129 adverse event reports that list this exact device (DI 10385082000009), received between Dec 11, 2018 and Sep 30, 2026 — classified as 10 deaths, 343 injuries, 776 malfunctions.
The most frequently reported specific device problem is “Battery Problem” (466 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 17, 2022Class ITerminated
The firm has become aware of PDM battery issues, including battery swelling, fluid leaking from the battery, and extreme overheating, which may pose a fire hazard. The firm has received a complaint alleging fire as a result of thermal energy from a DASH PDM.
Z-0154-2023 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Battery Problem466
- Overheating of Device346
- Insufficient Device Problem Information189
- Thermal Decomposition of Device149
- Failure to Power Up33
- Premature Discharge of Battery29
- Data Problem22
- Unintended System Motion21
Patient problems
- No Clinical Signs, Symptoms or Conditions684
- Hyperglycemia259
- Elevated Ketones/ Diabetic Ketoacidosis111
- Hypoglycemia94
- Vomiting52
- Nausea50
- Loss of consciousness36
- No Information25
Recent adverse event reports
The 25 most recent of 1,129 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Overheating of Device; Battery Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50955211 ↗MalfunctionSep 28, 2026Overheating of Device; Battery Problem
It was reported by the patient that the Omnipod Dash Controller/Personal Diabetes Manager (PDM) that their newly replaced PDM is not holding a charge. The patient also reported that the PDM charger is overheating. No harm to the patient was reported and no medical assistance was required. A replacement PDM and charger was issued to the patient.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50915315 ↗MalfunctionAug 27, 2026Fire
The patient stated that the controller charging cord showed brown discoloration at the ends, which the patient alleged to be indicative of signs of burning. Additionally, the patient reported that the device became hot during operation. No medical event was reported. The patient will be issued a new device.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50479932 ↗MalfunctionAug 25, 2026Stretched; Explosion
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50435429 ↗InjuryJul 7, 2026No Display/Image
THE PATIENT DECLINED TO PROVIDE INCIDENT DETAILS BUT STATED THAT THE EVENT WAS SIMILAR TO AN OCCURRENCE FROM (B)(6) 2026. BASED ON INFORMATION FROM THE (B)(6) 2026 INCIDENT, THE PATIENT WENT TO THE EMERGENCY ROOM WITH HYPERGLYCEMIA. NO BLOOD GLUCOSE READINGS WERE REPORTED. THE PATIENT REPORTED THAT THE POD CONTROLLER HAD A UNRESPONSIVE, BLACK SCREEN, AND WAS MAKING CONTINUOUS BEEPING AND VIBRATING NOISES. THE PATIENT WAS TREATED WITH INSULIN VIA INTRAVENOUS THERAPY AND FLUIDS.
Patient problems: Hyperglycemia
MAUDE report 25736545 ↗MalfunctionJun 29, 2026Overheating of Device; Battery Problem
IT WAS REPORTED THAT THE OMNIPOD DASH CONTROLLER/PERSONAL DIABETES MANAGER (PDM) HAD A POOR BATTERY LIFE AND WAS HOT TO TOUCH. NO HARM TO THE PATIENT WAS REPORTED AND NO MEDICAL ASSISTANCE WAS REQUIRED. A REPLACEMENT PDM WAS ISSUED TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25649705 ↗MalfunctionJun 23, 2026Unintended Power Up; Overheating of Device; Battery Problem
THE PATIENT REPORTED THAT HER OMNIPOD DASH PDM BECAME UNRESPONSIVE WHILE ATTEMPTING TO FILL A POD. THE DEVICE REMAINED STUCK ON THE ¿OMNIPOD DASH¿ LOADING SCREEN FOR OVER AN HOUR AND BECAME EXCESSIVELY HOT. SHE WAS UNABLE TO POWER IT OFF, EVEN AFTER REMOVING AND REINSERTING THE BATTERY. THE PDM WAS APPROXIMATELY TWO WEEKS OLD AND HAD SHOWN ISSUES SINCE INITIAL USE, INCLUDING UNEXPECTED DATE/TIME CHANGES, A SIX-HOUR DELAY BEFORE ALLOWING INSULIN DELIVERY, AND POOR BATTERY PERFORMANCE (HOLDING A CHARGE FOR ONLY 24¿30 HOURS). A REPLACEMENT OF THE PDM WAS PROCESSED FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25594218 ↗InjuryJun 11, 2026No Display/Image; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT WENT TO THE EMERGENCY ROOM (ER) WITH HYPERGLYCEMIA. THE PATIENT'S BLOOD GLUCOSE LEVELS EXCEEDED 600 MG/DL DUE TO LACK OF INSULIN DELIVERY. THE PATIENT REPORTED THAT THE POD CONTROLLER HAD A UNRESPONSIVE, BLACK SCREEN, AND WAS MAKING CONTINUOUS BEEPING AND VIBRATING NOISES. SYMPTOMS REPORTED INCLUDE THIRSTY, NAUSEA, AND DIZZINESS. THE PATIENT WAS TREATED WITH INSULIN VIA INTRAVENOUS (IV) THERAPY AND FLUIDS. THE PATIENT WAS DISCHARGED ONCE BLOOD SUGAR LEVELS DECREASED ON THE SAME DAY.
Patient problems: Dizziness; Hyperglycemia; Nausea; Polydipsia
MAUDE report 25480242 ↗MalfunctionJun 11, 2026Charging Problem; Excessive Heating
THE PATIENT REPORTED THAT WHEN CHARGING THE OMNIPOD PERSONAL DIABETES MANAGER (PDM), THE BOTTOM OF THE DEVICE OVERHEATS AND THE PDM CHARGES SLOWLY. ADDITIONALLY, THE PATIENT REPORTED THAT THE DEVICE CHARGES TO 100% BUT DRAINS COMPLETELY OVERNIGHT. THE PATIENT'S BLOOD GLUCOSE LEVELS WERE CONSISTENTLY ELEVATED (CURRENT READING AT 320 MG/DL) OVER A 3¿4-DAY PERIOD. THE PATIENT STATED TO BE USING A GREEN CHARGER (DEXCOM-ISSUED).
Patient problems: Hyperglycemia
MAUDE report 25470658 ↗MalfunctionJun 5, 2026Application Program Problem: Medication Error
IT WAS REPORTED BY THE LEGAL GUARDIAN (LG) THAT THE BASAL PRESETS OF THE PATIENT'S POD CONTROLLER WAS INACTIVE. THE LG NOTICED THIS WHEN THE PATIENT WOKE UP WITH ELEVATED BLOOD GLUCOSE (BG) LEVELS. NO IMPACT TO THE PATIENT'S GLUCOSE LEVELS WERE REPORTED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25417477 ↗MalfunctionMay 22, 2026Leak/Splash
IT WAS REPORTED THAT THE PATIENT'S CONTROLLER WAS MELTING AND THAT THE BATTERY WAS LEAKING. NO IMPACT TO BLOOD GLUCOSE LEVELS WAS REPORTED. MEDICAL TREATMENT WAS NOT REQUIRED. THE PATIENT WAS PROVIDED A REPLACEMENT CONTROLLER. THE PATIENT REPORTED THEY HAVE INSULIN INJECTION PENS TO USE, UNTIL THEIR NEW CONTROLLER ARRIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25254247 ↗MalfunctionMay 21, 2026Program or Algorithm Execution Failure
THE PATIENT REPORTED ISSUES WITH THEIR BASAL INSULIN PROGRAM. THE PATIENT STATED THAT THEIR BASAL INSULIN PRESET PROGRAMS WERE INACTIVE AND AS A RESULT THE PATIENT DID NOT RECEIVE INSULIN OVERNIGHT. THE PATIENT REPORTED THAT THEIR BLOOD GLUCOSE LEVELS WENT HIGH BUT DID NOT PROVIDE SPECIFIC VALUES. MEDICAL TREATMENT WAS NOT REQUIRED. PATIENT WAS PROVIDED A REPLACEMENT CONTROLLER.
Patient problems: Hyperglycemia
MAUDE report 25243648 ↗MalfunctionMay 15, 2026Overheating of Device
PATIENT REPORTED THAT THE PDM GETS HOT AFTER BEING ON THE CHARGER FOR ABOUT 20 MINUTES. THE PATIENT WAS USING AN INSULET PROVIDED CHARGING CABLE. NO PATIENT INJURY WAS REPORTED, AND MEDICAL INTERVENTION WAS NOT REQUIRED. THE PATIENT WAS PROVIDED A REPLACEMENT CONTROLLER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25182544 ↗MalfunctionMay 14, 2026Stretched
PATIENT REPORTED THAT THE BATTERY OF THE CONTROLLER HAS FULLY EXPANDED FROM THE BACK SIDE, AND HE IS AFRAID OF PLUGGING THE DEVICE INTO A CHARGER, AS THEY FEEL THE BATTERY MIGHT EXPLODE. NO IMPACT TO BLOOD GLUCOSE LEVELS WERE REPORTED. MEDICAL TREATMENT WAS NOT REQUIRED. THE PATIENT WAS PROVIDED A REPLACEMENT CONTROLLER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25171420 ↗MalfunctionMay 13, 2026Overheating of Device
IT WAS REPORTED THAT THE CHARGING CABLE GETS HOT WHENEVER THE CONTROLLER IS CHARGING. NO IMPACT TO THE PATIENT'S BLOOD GLUCOSE LEVELS WERE REPORTED. MEDICAL TREATMENT WAS NOT REQUIRED. THE PATIENT WAS PROVIDED A REPLACEMENT CONTROLLER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25162974 ↗MalfunctionMay 11, 2026Overheating of Device
IT WAS REPORTED THAT WHILE CHARGING, THE CONTROLLER AND CHARGER WERE OVERHEATING AND IT WOULD NOT TURN ON. NO IMPACT TO BLOOD GLUCOSE LEVELS WERE REPORTED. MEDICAL TREATMENT WAS NOT REQUIRED. THE PATIENT WAS SENT A REPLACEMENT CONTROLLER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25137594 ↗InjuryMay 1, 2026Operating System Becomes Nonfunctional
THE PATIENT REPORTED THAT HER GLUCOSE LEVELS INCREASED TO ABOVE 500 MG/DL AFTER THE DASH CONTROLLER DISPLAYED A BOOT/RECOVERY MODE SCREEN THAT THE PATIENT COULD NOT NAVIGATE AND SUBSEQUENTLY TURNED BLACK. MULTIPLE ATTEMPTS TO REBOOT THE CONTROLLER WERE UNSUCCESSFUL, INCLUDING BATTERY REMOVAL/REINSERTION AND CHARGING. THIS RESULTED IN OVERNIGHT HYPERGLYCEMIA. THE PATIENT REPORTED ADMINISTERING A MANUAL INSULIN INJECTION, WHICH RESULTED IN SEVERE HYPOGLYCEMIA, WITH BLOOD GLUCOSE DECREASING BELOW 2 MG/DL. SHE SUBSEQUENTLY LOST CONSCIOUSNESS AND WAS TRANSPORTED BY AMBULANCE TO THE EMERGENCY ROOM ON (B)(6) 2026, WHERE SHE REMAINED FOR APPROXIMATELY SEVEN HOURS. REPORTED DIAGNOSES INCLUDED HYPOGLYCEMIA, LOW BLOOD PRESSURE (BELOW 70), ATRIAL FIBRILLATION, AND HIGH RISK FOR STROKE OR HEART ATTACK; HOWEVER, NO STROKE OR HEART ATTACK WAS CONFIRMED. THE PATIENT WAS TREATED WITH INTRAVENOUS GLUCOSE AND INTRAVENOUS FLUIDS AND RETURNED HOME THE SAME DAY.
Patient problems: Hyperglycemia; Hypoglycemia; Loss of consciousness
MAUDE report 25058378 ↗MalfunctionApr 30, 2026Overheating of Device
THE PATIENT REPORTED THAT THEIR CONTROLLER WAS OVERHEATING, WHILE CHARGING. PATIENT WAS USING AN INSULET SUPPLIED CHARGING CABLE. NO IMPACT TO BLOOD GLUCOSE LEVELS WAS REPORTED. MEDICAL TREATMENT WAS NOT REQUIRED. A REPLACMENT CONTROLLER WAS SENT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25044302 ↗MalfunctionApr 15, 2026Stretched
IT WAS REPORTED THAT THE PERSONAL DIABETES MANAGER (PDM) BATTERY WAS EXPANDING. THE PATIENT REPORTED THE PDM HAD BEEN STILL WORKING AS EXPECTED AND THERE WAS NO IMPACT REPORTED TO THEIR BLOOD GLUCOSE LEVELS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24888142 ↗MalfunctionApr 9, 2026Failure to Deliver
IT WAS REPORTED THAT A BOLUS COULD NOT BE DELIVERED ON THE PATIENT'S DASH CONTROLLER/PERSONAL DIABETES MANAGER (PDM). A MESSAGE APPEARED ON THE PDM THAT THE PATIENT COULD NOT RECALL, HOWEVER, THERE WERE NO AUDIBLE ALARMS GENERATED. NO HARM TO THE PATIENT WAS REPORTED AND NO MEDICAL ASSISTANCE WAS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24829954 ↗MalfunctionApr 6, 2026Thermal Decomposition of Device; Display Difficult to Read; Overheating of Device
IT WAS REPORTED THAT THE CONTROLLER OVERHEATED, MELTED AND THE SCREEN CRACKED AND THE COLOR OF THE SCREEN CHANGED, WHILE THE CONTROLLER WAS CHARGING. THE USER WAS SENT A REPLACEMENT DEVICE. PATIENT'S BLOOD GLUCOSE LEVELS WERE REPORTED TO BE GREATER THAN 250 MG/DL. NO PATIENT INJURY WAS REPORTED, AND MEDICAL INTERVENTION WAS NOT REQUIRED.
Patient problems: Hyperglycemia
MAUDE report 24789324 ↗MalfunctionMar 19, 2026Smoking
IT WAS REPORTED BY THE PRESTATAIRE THAT THE PATIENT SAW SMOKE APPEARING AROUND THE CHARGING PORT AND CABLE WHILE THE OMNIPOD 5 CONTROLLER/PERSONAL DIABETES MANAGER WAS CHARGING. NO HARM TO THE PATIENT WAS REPORTED AND NO MEDICAL ASSISTANCE WAS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24636744 ↗MalfunctionMar 18, 2026Excess Flow or Over-Infusion
IT WAS REPORTED THAT ON (B)(6) 2026 THE PATIENT'S BLOOD GLUCOSE DECREASED TO 40 MG/DL WHILE WEARING THE POD BETWEEN 4 AND 24 HOURS. THE PATIENT THOUGHT THE POD KEPT DELIVERING INSULIN EVEN WHEN SUSPENDED. WHEN THEY REMOVED THE POD, IT STILL HAD INSULIN SLOWLY COMING OUT OF THE CANNULA.
Patient problems: Hypoglycemia
MAUDE report 24629913 ↗MalfunctionMar 17, 2026Failure to Deliver
IT WAS REPORTED THE PATIENT'S OMNIPOD DASH CONTROLLER/PERSONAL DIABETES MANAGER (PDM) DID NOT DELIVER A BOLUS. NO MESSAGES APPEARED ON THE PDM DISPLAY. THE PATIENT REMOVED THE POD. NO HARM TO THE PATIENT WAS REPORTED AND NO MEDICAL ASSISTANCE WAS REQUIRED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24615864 ↗MalfunctionMar 13, 2026Failure to Deliver
IT WAS REPORTED BY THE PATIENT THAT THEIR DASH SYSTEM WAS NOT DELIVERING BOLUSES THROUGHOUT THE NIGHT AND INTO THE NEXT DAY WITH NO ERROR MESSAGE PRODUCED. NO MEDICAL ASSISTANCE WAS REQUIRED AND THEY REPORTED THEIR BLOOD GLUCOSE VALUES REMAINED STABLE. THE PATIENT APPLIED A NEW POD AND REPORTED NO FURTHER ISSUES, CONFIRMING THAT BOLUSES WERE BEING DELIVERED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24589096 ↗
Other System, Test, Blood Glucose, Over The Counter devices
- OneTouch Verio Flex 023-271LifeScan Europe GmbH · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER3,703 reports
- FreeStyle Libre 2 71953Abbott Diabetes Care Inc · System, test, blood glucose, over the counter2 recalls1 death3,286 reports
- OneTouch Ultra2 021-105LifeScan U.S. LLC · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER1,731 reports
- Omnipod UST400Insulet Corporation · System, Test, Blood Glucose, Over The Counter12 deaths816 reports
- TRUE METRIX TRUE METRIX SELF MONITORING BLOOD GLUCOSE METER - NOT FOR RETAIL SALETrividia Health, Inc. · System, Test, Blood Glucose, Over The Counter418 reports
- FreeStyle Freedom Lite 70923-70Abbott Diabetes Care Inc · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER378 reports
- FreeStyle Lite 70829-70Abbott Diabetes Care Inc · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER378 reports
- FreeStyle Libre 2 71952Abbott Diabetes Care Inc · System, test, blood glucose, over the counter1 recall324 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page