TRI-LOCK GRIPTION 1012-04-070Recalls & reported problems

Depuy Orthopaedics, Inc. · TRI-LOCK BPS FEMORAL STEM 12/14 TAPER TRI-LOCK BPS W/GRIPTION SIZE 7 STD 109mm · Class 3 · Implant

All models of TRI-LOCK GRIPTION (26)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Orthopaedics Tri-Lock Gription hip stem is a metal-on-metal hip replacement used in total hip arthroplasty. Its 31 reports mostly describe pain, metal debris and fluid from the metal parts (metallosis), loosening or sinking of the stem in the bone, and broken stems or fractures around the implant, several needing revision surgery;29 reports describe injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
31
Deaths reported
0
Injuries reported
29
Recalls naming it
0

Summary

The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 10603295000785), received between Feb 23, 2018 and Nov 18, 2025 — classified as 29 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem15
  • Fracture3
  • Osseointegration Problem3
  • Appropriate Device Problem Term/Code Not Available2
  • Biocompatibility2
  • Degraded2
  • Insufficient Device Problem Information2
  • Migration2

Patient problems

  • No Code Available17
  • Pain15
  • Test Result9
  • Foreign Body Reaction7
  • Bone Fracture(s)5
  • Hypersensitivity/Allergic reaction5
  • Ambulation or Postural Difficulties4
  • Discomfort4

Recent adverse event reports

The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 18, 2025Naturally Worn; Naturally Worn

    PATIENT UNDERWENT REVISION SURGERY FOR METAL ON METAL PINNACLE / TRILOCK FRETTING CAUSING PSEUDO TUMOR. REMOVAL OF METAL LINER AND METAL HEAD WAS DONE WITH REPLACEMENT TO POLY LINER AND TS CERAMIC HEAD. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Hypersensitivity/Allergic reaction

    MAUDE report 23586752 ↗
  • InjuryAug 13, 2024Osseointegration Problem; Osseointegration Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO PERIPROSTHETIC FRACTURE. THERE WAS A FEMORAL STEM LOOSENING. DOI: UNKNOWN. DOR:(B)(6) 2024. AFFECTED SIDE: LEFT HIP.

    Patient problems: Inadequate Osseointegration; Limb Fracture

    MAUDE report 19974143 ↗
  • MalfunctionJul 18, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE SIZE 7 STEM WAS SITTING UP FROM THE SIZE 7 BROACH SO HAD TO OPEN A NEW STEM. SURGERY WAS DELAYED 2 MINUTES DUE TO THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19770094 ↗
  • InjuryJul 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE STEM AND HEAD WERE REVISED DUE TO PERIPROSTHETIC FRACTURE AFTER A FALL. DOI: UNKNOWN. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT HIP.

    Patient problems: Fall; Limb Fracture

    MAUDE report 19726183 ↗
  • InjuryJul 27, 2022Fracture; Fracture

    THE PATIENT, WHILE STANDING, HAS HAD A SUDDEN STRUCTURAL FAILURE. AT THE EMERGENCY ROOM IT WAS DIAGNOSED A RUPTURE OF THE STEM AT NECK LEVEL. A REVISION SURGERY HAS BEEN PLANNED FOR ON (B)(6) 2022.

    Patient problems: Insufficient Information

    MAUDE report 15112636 ↗
  • InjuryMar 29, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO METALLOSIS. ULTAMET METAL LINER/HEAD. THE HIP JOINT WAS BLACK AND GREEN WITH METAL DEBRIS. ADDITIONAL INFORMATION EVENT IN THE DER REPORTED OSTEOLYSIS. PRODUCT ISSUES WAS SUSPECTED AND CONTRIBUTED TO THE EVENT. DOI: UNKNOWN: DOR: (B)(6) 2021: LEFT HIP.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Discomfort; Distress; Fatigue; Hypersensitivity/Allergic reaction; Inflammation; Loss of Range of Motion; Metal Related Pathology; Necrosis; Pain; Scar Tissue; Swelling/ Edema; Unspecified Tissue Injury

    MAUDE report 13937600 ↗
  • InjuryMay 25, 2021Osseointegration Problem; Migration; Osseointegration Problem; Migration

    IT WAS REPORTED THAT TRILOCK STEM APPEARS TO HAVE SUBSIDED ON X-RAY FILMS. SURGEON STARTED STEM APPEARS TO BE UNDERSIZED. IN THE ADDITIONAL EVENT INFORMATION REPORTED LOOSENING OF THE STEM TO BONE-IMPLANT INTERFACE . DOI: (B)(6) 2019; DOR: (B)(6) 2021: HIP UNKNOWN.

    Patient problems: Inadequate Osseointegration

    MAUDE report 11880894 ↗
  • InjuryJan 13, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LITIGATION RECORD RECEIVED. PATIENT ALLEGES PAIN, EXCESSIVE COBALT AND CHROMIUM, ABRASION, AND EMOTIONAL DISTRESS FOR ECONOMIC LOSS AS WELL AS PUNITIVE DAMAGES.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Cyst(s); Foreign Body Reaction; Pain

    MAUDE report 11165017 ↗
  • InjuryOct 14, 2020Degraded; Naturally Worn; Degraded; Naturally Worn

    OFFICE NOTES(B)(6)2021 INDICATED THAT THE PATIENT EXPERIENCED A DISLOCATION POST THE (B)(6)2020 REVISION THAT WAS TREATED WITH CLOSED REDUCTION. IT WAS NOTED THERE WAS PAIN WITH THE DISLOCATION, BUT IT IS IMPROVING SINCE THE CLOSED REDUCTION. IT IS ALSO NOTED THAT HIS LEFT HIP IS SQUEAKY AND SORE. OFFICE NOTES (B)(6) 2021 INDICATE THE PATIENT IS CONTINUING TO EXPERIENCE PAIN, STIFFNESS, HIP SQUEAKING, AND DIFFICULTY AMBULATING. OBSERVATION AND CONTINUED STRENGTH TRAINING ARE THE ONLY INTERVENTIONS AT THIS TIME.

    Patient problems: Abscess; Ambulation or Postural Difficulties; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Foreign Body Reaction; Hypersensitivity/Allergic reaction; Inflammation; Limb Fracture; Muscular Rigidity; No Code Available; Pain; Physical Asymmetry; Swelling/ Edema

    MAUDE report 10680160 ↗
  • InjuryApr 1, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    UNF AND MEDICAL RECORDS RECEIVED. AFTER REVIEW OF MEDICAL RECORDS, IT INDICATED THAT THE PATIENT HAS A METAL ON METAL HIP IMPLANT. NO ALLEGATIONS WERE REPORTED. DOI: (B)(6) 2009. DOR: NONE REPORTED, (RIGHT HIP).

    Patient problems: No Code Available; Test Result

    MAUDE report 9911162 ↗
  • InjuryApr 1, 2020Degraded; Degraded

    UNF AND MEDICAL RECORDS RECEIVED. AFTER REVIEW OF MEDICAL RECORDS, PATIENT WAS REVISED TO ADDRESS FAILED LEFT TOTAL HIP REPLACEMENT DUE TO METALLOSIS, ARTICULAR SURFACE WEAR, AND CORROSION. IT WAS INDICATED THAT THE PATIENT HAS ELEVATED METAL LINES AND SIGNS OF CORROSION AND WEAR OF ARTICULAR SURFACE OF PREVIOUSLY PLACED METAL-ON-METAL HIP REPLACEMENT. OPERATIVE FINDINGS STATED FAILED LEFT TOTAL HIP REPLACEMENT METAL-ON-METAL CONSTRUCT DUE TO BOTH TRUNNIONOSIS AND METALLOSIS. THERE WAS NO SIGNIFICANT OSTEOLYSIS IN THE ACETABULAR SHELL AND FEMORAL STEM WAS WELL FIXATED WITH NO MECHANICAL COMPROMISE OF THE LOCKING MECHANISM OR TRUNNION. THIS WAS DESPITE SOME ELEMENT OF TRUNNIONOSIS AND SOME BACK SITE CORROSION AND METALLOSIS. THERE WAS NO PSEUDOTUMOR: THE PROCEDURE INCORPORATED SYNOVECTOMY OF THE JOINT, REMOVAL OF THE FEMORAL AND ACETABULAR BEARING SURFACE COMPONENT, AND REVISION OF BOTH THESE COMPONENTS. DOI: (B)(6) 2009; DOR: (B)(6) 2019 (LEFT HIP).

    Patient problems: Foreign Body Reaction; No Code Available; Test Result

    MAUDE report 9911108 ↗
  • InjuryMar 17, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PINNACLE CLAIM SUBMISSION FORM ALLEGES PAIN, OSTEOLYSIS, METALLOSIS AND ELEVATED METAL IONS. AFTER REVIEW OF MEDICAL RECORDS PATIENT WAS REVISED TO ADDRESS METALOSIS LEFT HIP. REVISION NOTES REPORTED THAT THERE WAS EVIDENCE OF METALOSIS. THERE WAS LOBULAR SYNOVIUM WHICH WAS GRAY IN COLOR. DOI: (B)(6) 2010; DOR: (B)(6) 2018 (LEFT HIP). THE PATIENT HAS BILATERAL HIP IMPLANTS, PLEASE SEE (B)(4) FOR THE RIGHT HIP.

    Patient problems: Foreign Body Reaction; No Code Available; Osteolysis; Pain; Test Result

    MAUDE report 9843240 ↗
  • InjuryJan 14, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING INSERTION OF FEMORAL IMPLANT, THE FEMUR CRACKED, AND SO THE SURGEON APPLIED AN ORTHOPEDIC CABLE. THERE WAS A 10-MINUTE SURGICAL DELAY ATTRIBUTED TO THE CASE. DOE: (B)(6) 2019, (UNKNOWN HIP).

    Patient problems: Bone Fracture(s); No Code Available; Not Applicable

    MAUDE report 9587182 ↗
  • InjuryJan 3, 2020Migration or Expulsion of Device; Migration or Expulsion of Device

    IT WAS REPORTED THAT THE PATIENT WAS DOING FINE AFTER SURGERY THEN HAD PAIN. X-RAY SHOWED SHE CRACKED HER FEMUR AND STEM SUBSIDED. FEMUR WAS CABLED AND A REVISION CORAIL STEM WAS IMPLANTED.

    Patient problems: Bone Fracture(s); No Code Available; Pain

    MAUDE report 9545382 ↗
  • MalfunctionDec 4, 2019Patient-Device Incompatibility; Patient-Device Incompatibility

    IT WAS REPORTED THAT THE SURGEON WAS USING TRILOCK FOR A BIPOLAR CASE. BROACHED ACCORDINGLY TO TECHNIQUE GUIDE UP TO A SIZE 7. WHEN IMPLANTING A SIZE 7 IMPLANT , THE IMPLANT WOULD NOT ADVANCE INTO FEMUR ALL THE WAY PHYSICIAN RE-BROACHED AND STILL WOULD NOT ADVANCE. WE HAD TO WASTE THE 7 AND IMPLANT SIZE 6. THOUGHT THE SIZE 7 IMPLANT WAS LARGE THAN ACTUAL SIZE. DOE: UNKNOWN. UNKNOWN AFFECTED SIDE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9412129 ↗
  • InjuryOct 18, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PF ALLEGES DIFFICULTY WALKING AND LIMITED MOBILITY. AFTER REVIEW OF MEDICAL RECORDS, PATIENT WAS REVISED TO ADDRESS PAINFUL RIGHT HIP METAL ON METAL TOTAL HIP REPLACEMENT AND PARTIAL GLUTEUS MEDIUS TENDON AVULSION. REVISION NOTE REPORTED ADHESIONS, GLUTEUS MEDIUS TENDON WAS NOT ATTACHED TO THE TROCHANTER AND ACETABULAR COMPONENTS WERE CONFIRMED TO BE WELL FIXED INTO 45 DEGREES OF ABDUCTION AND 20 DEGREES OF ANTEVERSION; AND IT WAS ALSO STATED THAT THERE WAS NO EVIDENT TRUNIONOSIS.

    Patient problems: Adhesion(s); Ambulation or Postural Difficulties; Foreign Body Reaction; Hypersensitivity/Allergic reaction; Limited Mobility Of The Implanted Joint; No Code Available; Pain; Test Result

    MAUDE report 9207666 ↗
  • InjuryOct 16, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PPF ALLEGES DIFFICULTY WALKING AND LIMITED MOBILITY.

    Patient problems: Ambulation or Postural Difficulties; Discomfort; Foreign Body Reaction; Hypersensitivity/Allergic reaction; Limited Mobility Of The Implanted Joint; Pain; Test Result

    MAUDE report 9198303 ↗
  • InjuryMay 15, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    TRILOCK STEM REVISED DUE TO PERIPROSTHETIC FRACTURE. REASON FOR REVISION SURGERY. PERIPROSTHETIC FRACTURE,

    Patient problems: Bone Fracture(s); No Code Available; Pain

    MAUDE report 8611496 ↗
  • InjuryMay 8, 2019Biocompatibility; Adverse Event Without Identified Device or Use Problem; Biocompatibility; Adverse Event Without Identified Device or Use Problem

    LITIGATION ALLEGES THAT DUE TO METALLOSIS AND FAILURE OF OSSEOINTEGRATION OF THE PINNACLE DEVICE, PATIENT UNDERWENT REVISION ON BOTH SIDES. UPDATE: IN ADDITION TO WHAT WERE PREVIOUSLY ALLEGED, PPF ALLEGES COMPONENT FRACTURE, METAL WEAR, METALLOSIS, AND ELEVATED METAL IONS. DOI: (B)(6) 2009; DOR: (B)(6) 2011; (RIGHT HIP).

    Patient problems: No Code Available; Test Result; Tissue Damage

    MAUDE report 8590767 ↗
  • InjuryFeb 15, 2019Corroded; Corroded

    COMPLAINT DESCRIPTION: LITIGATION ALLEGES PATIENT SUFFERS FROM PAIN, DISCOMFORT, SWELLING IN HER LEG, AND DIFFICULTY SLEEPING. UPDATE: PATIENT WAS REVISED DUE TO PAIN AND A PSEUDOTUMOR. UPDATE: IN ADDITION TO WHAT WERE PREVIOUSLY ALLEGED, PPF ALLEGES PSEUDOTUMOR, METAL WEAR, METALLOSIS, AND ELEVATED METAL IONS. AFTER REVIEW OF MEDICAL RECORDS, PATIENT WAS REVISED TO ADDRESS PAIN DUE TO INTERNAL ORTHOPAEDIC PROSTHETIC DEVICE, INITIAL ENCOUNTER. REVISION NOTES STATED OF SCAR TISSUE AND REMNANTS OF PSEUDOTUMOR CAPSULE. THERE WAS SOME MILD CORROSION ON THE ARTICULATION WHICH WAS REMOVED WITH A BOVIE SCRATCH PAD. DOI: (B)(6) 2009; DOR: (B)(6) 2016; (RIGHT HIP).

    Patient problems: Discomfort; No Code Available; Pain; Scar Tissue; Swelling; Test Result; Tissue Damage

    MAUDE report 8343744 ↗
  • InjuryFeb 7, 2019Fracture; Osseointegration Problem; Migration; Fracture; Osseointegration Problem; Migration

    IT WAS REPORTED THAT ONE MONTH AGO, PATIENT WAS OUT FROM HIS TOTAL HIP REPLACEMENT. HE WAS LEAVING A RESTAURANT WALKING UNASSISTED AND FELL RESULTING IN A PERIPROSTHETIC FRACTURE OF HIS FEMUR. THE STEM AND HEAD WERE REMOVED AND REPLACED WITH NEW COMPONENTS.

    Patient problems: Fall; Hip Fracture; Inadequate Osseointegration; No Code Available

    MAUDE report 8315733 ↗
  • InjuryJan 30, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO PAIN.

    Patient problems: No Code Available; Pain

    MAUDE report 8288687 ↗
  • InjuryJan 24, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT REVISED DUE TO INFECTION.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 8273780 ↗
  • InjuryJan 4, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT FELL AND BROKE HER FEMUR. DURING THE OPEN REDUCTION AND INTERNAL FIXATION (ORIF), THE SURGEON NOTICED A HEMATOMA AND SOME FLUID IN THE JOINT. THE SURGEON DECIDED TO DO A HEAD AND LINER EXCHANGE. DOI: (B)(6) 2018; DOR: (B)(6) 2018; UNKNOWN HIP.

    Patient problems: Bone Fracture(s); Hematoma; No Code Available

    MAUDE report 8217721 ↗
  • InjuryDec 19, 2018Biocompatibility; Biocompatibility

    LITIGATION ALLEGES PATIENT EXPERIENCED SEVERE PAIN, DISCOMFORT, AND SWELLING. UPDATE 29 JUN 2018 RECEIPT OF PPF AND STICKER SHEETS. IN ADDITION TO WHAT PREVIOUSLY ALLEGED, PPF ALLEGED ELEVATED METAL IONS.

    Patient problems: Discomfort; No Code Available; Pain; Swelling; Test Result

    MAUDE report 8180981 ↗

Other Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented devices →