Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10603295004899), received between Mar 14, 2019 and Jul 22, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1
  • Device-Device Incompatibility1
  • Difficult to Insert1
  • Naturally Worn1
  • Off-Label Use1

Patient problems

  • Insufficient Information1
  • Joint Dislocation1
  • No Code Available1
  • Not Applicable1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 22, 2025Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE DOCTOR SAID PATIENT HAD HIP DONE OVER 35 YEARS AGO. THE CUP WAS IDENTIFIED AS A DEPUY ACS CUP, TO WHICH SURGEON PERFORMED POLY EXCHANGE. THE OLD POLY HAD SEVERE WEAR, SO A NEW 20 DEGREE 28MM ID GROUP 3 LINER WAS IMPLANTED. THE HEAD THAT WAS REMOVED WAS CONFIRMED TO BE A LARGE TAPER, BUT ALL NUMBERS INDICATING NECK LENGTH HAD BEEN WORN AWAY AND WAS UNABLE TO BE READ. TRIAL REDUCTIONS WERE PERFORMED WITH 28MM LARGE TAPER TRIALS AND THE +8 WAS THE PREFERRED OPTION. THE REAL HEAD WAS THEN OPENED AND IMPLANTED. STABILITY OF THE HIP WAS THEN CONFIRMED AND THE CLOSING PROCESS BEGAN. NO ADVERSE EVENTS OR DELAYS TOOK PLACE. DOI: 35 YEARS AGO. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT SIDE.

    Patient problems: Insufficient Information

    MAUDE report 22569348 ↗
  • MalfunctionJun 3, 2020Difficult to Insert; Off-Label Use; Device-Device Incompatibility; Difficult to Insert; Off-Label Use; Device-Device Incompatibility

    THE PATIENT UNDERGO REVISION SURGERY DUE TO POLY WEAR THAT WAS REPORTED IN (B)(4). THE LINER WOULD NOT LOCK INTO THE CUP. A SECOND LINER WAS THEN OPENED WITH THE SAME RESULT. THE DOCTOR DETERMINED THAT THE LOCKING MECHANISM HAD BEEN COMPROMISED SO HE ELECTED TO CEMENT A LINER IN PLACE. DOE: (B)(6) 2020 RIGHT SIDE.

    Patient problems: Not Applicable

    MAUDE report 10116939 ↗
  • InjuryMar 14, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    SUBJECT ID: (B)(6). CLINICAL NOTIFICATION RECEIVED FOR REVISION DUE TO DISLOCATION. DOI CUP: (B)(6) 1996; DOI HEAD AND LINER: (B)(6) 2017; DOR: (B)(6) 2019; (LEFT HIP). HEAD, LINER AND CUP REVISED.

    Patient problems: Joint Dislocation; No Code Available

    MAUDE report 8421211 ↗

Other PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices →