PINNACLE 1217-11-048Recalls & reported problems

Depuy Orthopaedics, Inc. · PINNACLE HIP SOLUTIONS DUOFIX HA ACETABULAR SHELL 100 SERIES 48mm OD · Class 2 · Implant

All models of PINNACLE (182)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10603295009047), received between Jan 8, 2019 and Sep 8, 2026 — classified as 5 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated4
  • Noise, Audible2
  • Separation Problem1
  • Use of Device Problem1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Foreign Body Reaction1
  • Insufficient Information1
  • No Code Available1
  • Pain1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjurySep 8, 2026Device Dislodged or Dislocated

    It was reported the patient underwent a revision THR. The Poly Liner disengaged from the Shell and broke into a few pieces. The Cup and Implant Sector Shell were extracted and a new Poly Insert and Screw were added. The Stem was well fixed; the Taper was not damaged and a revision Delta TS Head was used.

    Patient problems: Insufficient Information

    MAUDE report 50670778 ↗
  • InjuryJul 11, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ADDITIONAL INFORMATION RECEIVED: THERE WAS NO SURGICAL DELAY.

    Patient problems: Foreign Body Reaction

    MAUDE report 19719862 ↗
  • InjuryOct 18, 2023Use of Device Problem; Device Dislodged or Dislocated; Noise, Audible; Use of Device Problem; Device Dislodged or Dislocated; Noise, Audible

    CUP REVISION FOR PAIN AND SQUEAKING OF IMPLANT. ON X RAY, CUP APPEARS OPEN AND TOO ANTEVERTED. FOR OPEN ASSESSMENT OF CUP AND LINER. INTRAOPERATIVELY THE LINER WAS FOUND TO HAVE COMPLETELY MALFORMED AND CERAMIC HEAD HAS BEEN RUBBING AGAINST METAL SHELL. SHELL LEFT IN SITU DUE TO PATIENTS AGE AND BMI AND BONE LOSS - +4 10 DEGREE FACE CHANGING LINER INSERTED WITH 32 +1 REVISION HEAD. DOI: (B)(6) 2017. DOE: (B)(6) 2023. AFFECTED SIDE.

    Patient problems: Pain

    MAUDE report 17957760 ↗
  • InjuryJul 23, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated

    REVISION HIP ON (B)(6) 2021, WHERE A PINNACLE ACETABULAR POLY LINER WAS COMPLETELY WORN AND DELAMINATED. PRIMARY SURGERY TOOK PLACE IN 2014 AT ANOTHER (B)(6) HOSPITAL, THE PINNACLE SHELL AND LINER WAS REVISED. THE PRIMARY FEMORAL STEM WAS A CORAIL.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12218003 ↗
  • InjuryJan 8, 2019Noise, Audible; Separation Problem; Noise, Audible; Separation Problem

    IT WAS REPORTED THAT PATIENT WAS PREVIOUSLY A REPEAT DISLOCATOR. AFTER CLOSE TO A YEAR OF NO REPEAT PROBLEMS, PATIENT NOTICED A SQUEAKING NOISE AND WAS X-RAYED. LOOKED AS THOUGH SHE HAD DISLOCATED AGAIN AND HEAD WAS RESTING ON LIP OF 10 "DEGRESS". LINER: DURING SURGERY IT WAS FOUND THAT THE LINER HAD FRACTURED IN TWO PIECES AND POSSIBLY HAD DISASSOCIATED FROM THE CUP WHILE IN PATIENT. NOTICED DAMAGE ON VARIOUS PARTS OF LINER, MARKS ON A HEAD FROM METAL ON CERAMIC ARTICULATION. CUP WAS WELL FIXED. PATIENT WAS REVISED TO LARGER CUP AND LARGER HEAD.

    Patient problems: No Code Available

    MAUDE report 8228976 ↗

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices

All PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices →