Skip to content

PINNACLE 1217-11-054Recalls & reported problems

Depuy Orthopaedics, Inc. · PINNACLE HIP SOLUTIONS DUOFIX HA ACETABULAR SHELL 100 SERIES 54mm OD · Class 2 · Implant

All models of PINNACLE (182)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10603295009078), received between Mar 13, 2020 and Feb 3, 2026 — classified as 4 injuries.

The most frequently reported specific device problem is “Malposition of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Malposition of Device1
  • Migration1
  • Use of Device Problem1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Foreign Body Reaction1
  • Insufficient Information1
  • Joint Disorder1
  • No Code Available1
  • Pain1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 3, 2026Migration

    IT WAS REPORTED THAT THE PATIENT UNDERWENT THE HIP REVISION PROCEDURE DUE TO MIGRATION OF THE PINNACLE ACETABULAR CUP. THE STEM CAME OUT WHEN THE HEAD WAS REMOVED, THEREFORE THE STEM WAS REPLACED AS WELL. IMPLANTS WERE REPLACED WITH PINNACLE MULTIHOLE SZ 56, +4/10 LINER, 36 +12 CERAMIC HEAD AND C-STEM SZ 2 HIGH OFFSET.

    Patient problems: Insufficient Information

    MAUDE report 24247785 ↗
  • InjuryJun 18, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION CASE DUE TO INFECTION ON (B)(6) 2021. THE CASE HAD OCCURRED AND ALL IMPLANTS REMOVED DUE TO INFECTION BEING PRESENT. COMPETITOR IMPLANTS CEMENTED INSITU FOR 1ST STAGE REVISION.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12025680 ↗
  • InjuryDec 23, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE ACETABULAR CUP, CERAMIC INSERT AND CERAMIC FEMORAL HEAD WERE REMOVED, AND REPLACED WITH COMPETITOR PRODUCTS. THE SURGEON STATED THAT PREVIOUS SURGEON, USAGE AND SURGERY DATE WERE UNKNOWN AS HE HAD INHERITED THE PATIENT.

    Patient problems: Pain

    MAUDE report 11065841 ↗
  • InjuryMar 13, 2020Use of Device Problem; Malposition of Device; Use of Device Problem; Malposition of Device

    THE CASE WAS A HEAD AND LINER EXCHANGE FOR IMPINGEMENT. DETAILS ¿ IMPLANTS INSITU. WERE CERAMIC ON CERAMIC, CUP WAS MAL-POSITIONED CAUSING IMPINGEMENT AND SQUEAKING. UPON DISSECTION NOTICED METALLOSIS FROM STEM NECK IMPINGING ON RIM OF CUP. REMOVED CERAMIC LINER AND HEAD AND REPLACED WITH A +4 NEUTRAL POLY LINER AND TS CERAMIC HEAD. IMPLANT DETAILS FOR EXTRACTION AND INSERTION ATTACHED.

    Patient problems: Foreign Body Reaction; Joint Disorder; No Code Available

    MAUDE report 9831859 ↗

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices

All PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page