PINNACLE POROCOAT 1217-20-050Recalls & reported problems
Depuy Orthopaedics, Inc. · PINNACLE POROCOAT ACETABULAR SHELL MULTIHOLE II 50mm OD · Class 3 · Implant
All models of PINNACLE POROCOAT (38)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10603295009610), received between Aug 14, 2018 and Sep 20, 2024 — classified as 10 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Naturally Worn” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem5
- Naturally Worn4
- Malposition of Device3
- Device Dislodged or Dislocated1
- Loose or Intermittent Connection1
- Loss of Osseointegration1
- Osseointegration Problem1
Patient problems
- Pain6
- Inflammation4
- No Code Available4
- Discomfort3
- Inadequate Osseointegration3
- Unspecified Infection3
- Not Applicable2
- Bone Fracture(s)1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 20, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2011, THE PRIMARY SURGERY WAS PERFORMED. AFTER THE SURGERY, THE DISLOCATION WAS CONFIRMED. THE EVENT DATE IS UNKNOWN. THE REVISION SURGERY WAS PERFORMED ON (B)(6) 2016, TO REPLACE THE LINER, THE HEAD AND THE STEM. AFTER THE REVISION SURGERY, THE LINER IN QUESTION WAS BROKEN. THE EVENT DATE IS UNKNOWN. THE REVISION SURGERY IS SCHEDULED ON (B)(6) 2023, TO REPLACE THE CUP, THE HEAD AND THE STEM IN QUESTION. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 20273027 ↗InjuryJul 26, 2023Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE PATIENT RETURN FOR AN X-RAY AS PART OF FOLLOW UP AND IT WAS IDENTIFIED THAT THE IMPLANT WAS FRACTURED. A REVISION SURGERY WAS ARRANGED AND PERFORMED ON (B)(6) 2023. UPON REMOVAL, THE IMPLANT WAS FOUND TO BE FRACTURED AND 2 OTHER DEVICES WERE FOUND TO BE SCRATCHED. PINNACLE POROCOAT MULTIHOLE COATING SCRATCHED OFF SURFACE. ALTRX LINER CRACKED. CERAMIC HEAD SURFACE SCRATCHED. DOI: UNKNOWN. DOE: (B)(6) 2023. AFFECTED SIDE: UNKNOWN.
Patient problems: Inflammation
MAUDE report 17397416 ↗MalfunctionMar 8, 2022Naturally Worn; Adverse Event Without Identified Device or Use Problem; Naturally Worn; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2012, THE PATIENT UNDERWENT THE PRIMARY THA SURGERY WITH THE CUP, THE STEM AND THE HEAD IMPLANTS IN QUESTION. AFTER THE SURGERY, IT WAS CONFIRMED THAT OSTEOLYSIS SUSPECTED TO BE ARMD AT THE BACK OF THE CUP ON AN UNKNOWN DATE. A REVISION SURGERY IS PLANNED ON (B)(6) 2022. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Foreign Body Reaction; Osteolysis
MAUDE report 13693649 ↗InjuryOct 21, 2019Malposition of Device; Device Dislodged or Dislocated; Naturally Worn; Malposition of Device; Device Dislodged or Dislocated; Naturally Worn
IT WAS REPORTED THAT THE SURGEON WAS SCHEDULED TO DO AN ABDUCTER TENOTOMY REPAIR. WHEN HE OPENED THE PATIENT HE FOUND METAL DEBRIS FROM THE STEM IMPINGING ON THE POSTERIOR PART OF THE CUP AND LINER. HE DECIDED TO REMOVE THE CUP, LINER, SCREW; AND HEAD, TO REPOSITION IN THE PROPER PLACE TO GIVE ADEQUATE STABILITY AND STOP THE IMPINGEMENT ON THE STEM. THE SUMMIT STEM WAS RETAINED. WE REAMED THE ACETABULUM UP TO A 56, PUT IN A 56 GRIPTION MULTI, AND IMPLANTED LINER AND HEAD TO FINISH THE CASE. THE SURGEON REQUESTED THE IMPLANTS BACK TO GIVE TO THE PATIENT. DOI: (B)(6) 2018; DOR: (B)(6) 2019; LEFT HIP.
Patient problems: Joint Disorder; No Code Available; Not Applicable
MAUDE report 9213439 ↗InjuryApr 16, 2019Malposition of Device; Malposition of Device
LITIGATION ALLEGES PAIN AND MISPOSITIONED CUP. DOI: (B)(6) 2006 - DOR: (B)(6) 2014 (LEFT HIP).
Patient problems: No Code Available; Pain; Scarring
MAUDE report 8518946 ↗InjuryApr 16, 2019Malposition of Device; Malposition of Device
LITIGATION ALLEGES PAIN AND MISPOSITIONED CUP. DOI: (B)(6) 2007 - DOR: (B)(6) 2014, (RIGHT HIP).
Patient problems: No Code Available; Pain
MAUDE report 8518936 ↗InjuryMar 19, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PINNACLE LITIGATION RECORDS RECEIVED. LITIGATION ALLEGED INCREASED BLOOD HEAVY METAL IONS DUE TO THE FRICTION AND WEAR BETWEEN THE HEAD AND LINER, SEVERE PAIN, DISCOMFORT AND INFLAMMATION. DOI: (B)(6) 2008; DOR: (B)(6) 2017; RIGHT HIP.
Patient problems: Discomfort; Fatigue; Inflammation; Injury; Limited Mobility Of The Implanted Joint; Necrosis; Pain; Tissue Damage; Unspecified Infection
MAUDE report 8431731 ↗InjuryFeb 5, 2019Loss of Osseointegration; Loss of Osseointegration
LITIGATION ALLEGES THAT THE PATIENT SUFFERS FROM PAIN. UPDATE 07 MAY 2018: (B)(4) HAS BEEN RE-OPENED UNDER (B)(4) DUE TO RECEIPT OF PPF AND STICKER SHEETS. IN ADDITION TO WHAT WERE PREVIOUSLY ALLEGED, PPF ALLEGES LOOSENING OF CUP. DOI: (B)(6) 2009 - DOR: (B)(6) 2010, (RIGHT HIP).
Patient problems: Inadequate Osseointegration; No Code Available; Pain
MAUDE report 8304525 ↗InjuryJan 15, 2019Osseointegration Problem; Osseointegration Problem
PPF ALLEGES LOOSENING OF CUP AND METALLOSIS. DOI: (B)(6) 2007, DOR: FEB 19, 2010, (RIGHT HIP).
Patient problems: Discomfort; Inadequate Osseointegration; Inflammation; Pain
MAUDE report 8247483 ↗InjuryNov 8, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
InjuryOct 23, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IMPLANTS WERE REMOVED DURING HIP REVISION SURGERY ON FRIDAY (B)(6) BY MR (B)(6). THE SURGEON HASN¿T REQUESTED ANY FURTHER INFORMATION. PELVIC DISCONTINUITY WAS THE REASON FOR REVISION. THE LOT NUMBERS WEREN¿T READABLE ON THE EXPLANTS.
Patient problems: Bone Fracture(s)
MAUDE report 7994474 ↗InjuryAug 14, 2018Loose or Intermittent Connection; Naturally Worn; Loose or Intermittent Connection; Naturally Worn
LITIGATION PAPERS ALLEGE: PATIENT EXPERIENCED SEVERE PAIN, DISCOMFORT, AND INFLAMMATION IN THE LEG, A POPPING AND CLICKING SOUND IN THE HIP; THE SENSATION THAT THE IMPLANT IS UNSTABLE AND WILL GIVE OUT; INABILITY TO WALK MODERATE OR LONG DISTANCES; INABILITY TO SLEEP ON PATIENT'S SIDE WITHOUT SEVERE PAIN; INABILITY TO SIT FOR MODERATE TO LONG PERIODS OF TIME; METALLOSIS; COBALT-CHROMIUM METAL TOXICITY; INFECTION; AND OTHER COMPLICATIONS. PPF ALLEGES LOOSENING OF CUP, METAL WEAR AND METALLOSIS. IT WAS ALSO INDICATED IN THE PPF THAT CUP, LINER, HEAD AND STEM WERE REVISED. DOI: (B)(6) 2005; DOR: (B)(6) 2005; (LEFT HIP).
Patient problems: Discomfort; Inadequate Osseointegration; Inflammation; Not Applicable; Pain; Unspecified Infection
MAUDE report 7780934 ↗
Other Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented devices
- BIOLOX DELTA 1365-36-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented839 reports
- BIOLOX DELTA 1365-36-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented643 reports
- PINNACLE POROCOAT 1217-22-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented335 reports
- PINNACLE POROCOAT 1217-22-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented263 reports
- BIOLOX DELTA 1365-36-330Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented255 reports
- PINNACLE POROCOAT 1217-22-050Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented253 reports
- PINNACLE POROCOAT 1217-22-056Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented220 reports
- SUMMIT POROCOAT 1570-11-120Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented1 death204 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page